轫致辐射Y90 SPECT-CT成像的能量窗优化:一项幻象研究

Q3 Health Professions
D. Curto, F. Bonutti, Youssef Bouzekraoui, Hicham Asmi, F. Bentayeb
{"title":"轫致辐射Y90 SPECT-CT成像的能量窗优化:一项幻象研究","authors":"D. Curto, F. Bonutti, Youssef Bouzekraoui, Hicham Asmi, F. Bentayeb","doi":"10.22038/IJMP.2021.51256.1837","DOIUrl":null,"url":null,"abstract":"Introduction: In Yttrium-90 SPECT imaging, the energy window and collimator used during acquisition can have a major effect on image quality. In this work, we used a new and independent method to verify the prevoius results suggest different energy ranges, but all centered at 130 keV. Materials and method: We used Siemens Symbia SPECT-CT system fitted with High Energy General Purpose (HEGP), Medium Energy General Purpose (MEGP) and Low Energy High Resolution (LEHR) to acquire data from NEMA IEC PET Body Phantom with its 6 different spheres of 3.7, 2.8, 2.2, 1.7, 1.3, 1.0 cm diameter. Results:HEGP collimator is the most suitable for acquisitions of 90Y bremsstrahlung radiation from the point of view of the correct volume reproduction. For the bigger sphere’s study, the optimum ISO-counting curves is related to the energy range centred in 130 keV.Conclusion: The results obtained are consistent with previous studies. The Iso Counting Curves method can help to improve image quality.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-02-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Energy window optimization for Bremsstrahlung Y90 SPECT-CT imaging : a phantom study\",\"authors\":\"D. Curto, F. Bonutti, Youssef Bouzekraoui, Hicham Asmi, F. Bentayeb\",\"doi\":\"10.22038/IJMP.2021.51256.1837\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: In Yttrium-90 SPECT imaging, the energy window and collimator used during acquisition can have a major effect on image quality. In this work, we used a new and independent method to verify the prevoius results suggest different energy ranges, but all centered at 130 keV. Materials and method: We used Siemens Symbia SPECT-CT system fitted with High Energy General Purpose (HEGP), Medium Energy General Purpose (MEGP) and Low Energy High Resolution (LEHR) to acquire data from NEMA IEC PET Body Phantom with its 6 different spheres of 3.7, 2.8, 2.2, 1.7, 1.3, 1.0 cm diameter. Results:HEGP collimator is the most suitable for acquisitions of 90Y bremsstrahlung radiation from the point of view of the correct volume reproduction. For the bigger sphere’s study, the optimum ISO-counting curves is related to the energy range centred in 130 keV.Conclusion: The results obtained are consistent with previous studies. The Iso Counting Curves method can help to improve image quality.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-02-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2021.51256.1837\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2021.51256.1837","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

在钇-90 SPECT成像中,采集过程中使用的能量窗口和准直器会对图像质量产生重大影响。在这项工作中,我们使用了一种新的独立方法来验证先前的结果,这些结果提出了不同的能量范围,但都集中在130 keV。材料和方法:我们使用西门子Symbia SPECT-CT系统,配备高能通用(HEGP),中能通用(MEGP)和低能高分辨率(LEHR),获取NEMA IEC PET体体幻影的数据,其直径为3.7,2.8,2.2,1.7,1.3,1.0 cm的6种不同球体。结果:从正确的体积再现角度看,HEGP准直仪最适合采集90Y轫致辐射。对于较大球体的研究,最佳的iso计数曲线与以130 keV为中心的能量范围有关。结论:所得结果与前人研究结果一致。Iso计数曲线方法可以帮助提高图像质量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Energy window optimization for Bremsstrahlung Y90 SPECT-CT imaging : a phantom study
Introduction: In Yttrium-90 SPECT imaging, the energy window and collimator used during acquisition can have a major effect on image quality. In this work, we used a new and independent method to verify the prevoius results suggest different energy ranges, but all centered at 130 keV. Materials and method: We used Siemens Symbia SPECT-CT system fitted with High Energy General Purpose (HEGP), Medium Energy General Purpose (MEGP) and Low Energy High Resolution (LEHR) to acquire data from NEMA IEC PET Body Phantom with its 6 different spheres of 3.7, 2.8, 2.2, 1.7, 1.3, 1.0 cm diameter. Results:HEGP collimator is the most suitable for acquisitions of 90Y bremsstrahlung radiation from the point of view of the correct volume reproduction. For the bigger sphere’s study, the optimum ISO-counting curves is related to the energy range centred in 130 keV.Conclusion: The results obtained are consistent with previous studies. The Iso Counting Curves method can help to improve image quality.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信