气缸式下肢康复执行器的瞬态响应研究

Duc minh Dao, P. Pham, Tuyetlien Tran, T. Le
{"title":"气缸式下肢康复执行器的瞬态响应研究","authors":"Duc minh Dao, P. Pham, Tuyetlien Tran, T. Le","doi":"10.14203/j.mev.2018.v9.65-72","DOIUrl":null,"url":null,"abstract":"A lower limb rehabilitation device was designed using the compressed air cylinder in order to answer the particular request in Vietnam. This paper is presenting the results of a study of the device response. Dynamic equation of the actuator and equations of the proportional valve have been established. The relationship between the input signal and the output signal of the actuator was derived. Inventor® software was used to design the mechanical structure of the device. Matlab® software was used to calculate the parameters values of the PID controller by simulating the response of the actuator. The results show that the response time of both knee drive and hip drive mechanisms are 8 seconds while the overshoot of both knee drive and hip drive mechanisms are 1%. Moreover, the starting torque of the knee drive mechanism is 17 Nm, and the starting torque of the hip drive mechanism is 35 Nm. The simulation results show that the PID controller gives a fast response time and a low overshoot.","PeriodicalId":30530,"journal":{"name":"Journal of Mechatronics Electrical Power and Vehicular Technology","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2018-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":"{\"title\":\"Study on the transient response of lower limb rehabilitation actuator using the pneumatic cylinder\",\"authors\":\"Duc minh Dao, P. Pham, Tuyetlien Tran, T. Le\",\"doi\":\"10.14203/j.mev.2018.v9.65-72\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A lower limb rehabilitation device was designed using the compressed air cylinder in order to answer the particular request in Vietnam. This paper is presenting the results of a study of the device response. Dynamic equation of the actuator and equations of the proportional valve have been established. The relationship between the input signal and the output signal of the actuator was derived. Inventor® software was used to design the mechanical structure of the device. Matlab® software was used to calculate the parameters values of the PID controller by simulating the response of the actuator. The results show that the response time of both knee drive and hip drive mechanisms are 8 seconds while the overshoot of both knee drive and hip drive mechanisms are 1%. Moreover, the starting torque of the knee drive mechanism is 17 Nm, and the starting torque of the hip drive mechanism is 35 Nm. The simulation results show that the PID controller gives a fast response time and a low overshoot.\",\"PeriodicalId\":30530,\"journal\":{\"name\":\"Journal of Mechatronics Electrical Power and Vehicular Technology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-12-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"5\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Mechatronics Electrical Power and Vehicular Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.14203/j.mev.2018.v9.65-72\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Mechatronics Electrical Power and Vehicular Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14203/j.mev.2018.v9.65-72","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

摘要

针对越南的特殊要求,设计了一种使用压缩气缸的下肢康复装置。本文介绍了器件响应的研究结果。建立了执行机构的动力学方程和比例阀的动力学方程。推导了执行器输入信号与输出信号之间的关系。采用Inventor®软件设计装置的机械结构。通过仿真作动器的响应,利用Matlab®软件计算PID控制器的参数值。结果表明,膝关节驱动机构和髋关节驱动机构的响应时间均为8 s,膝关节驱动机构和髋关节驱动机构的超调量均为1%。其中,膝关节驱动机构的启动力矩为17 Nm,髋关节驱动机构的启动力矩为35 Nm。仿真结果表明,该PID控制器具有较快的响应时间和较低的超调量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Study on the transient response of lower limb rehabilitation actuator using the pneumatic cylinder
A lower limb rehabilitation device was designed using the compressed air cylinder in order to answer the particular request in Vietnam. This paper is presenting the results of a study of the device response. Dynamic equation of the actuator and equations of the proportional valve have been established. The relationship between the input signal and the output signal of the actuator was derived. Inventor® software was used to design the mechanical structure of the device. Matlab® software was used to calculate the parameters values of the PID controller by simulating the response of the actuator. The results show that the response time of both knee drive and hip drive mechanisms are 8 seconds while the overshoot of both knee drive and hip drive mechanisms are 1%. Moreover, the starting torque of the knee drive mechanism is 17 Nm, and the starting torque of the hip drive mechanism is 35 Nm. The simulation results show that the PID controller gives a fast response time and a low overshoot.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.70
自引率
0.00%
发文量
10
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信