{"title":"成人头颅计算机断层扫描辐射剂量的优化:辐射剂量对图像质量的评估","authors":"N. Okoro, V. Changizi, E. Jazayeri, F. Pak","doi":"10.22038/IJMP.2019.40477.1565","DOIUrl":null,"url":null,"abstract":"Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. \nMaterial and Methods: A poly(methyl methacrylate) phantom was used for quality control test on CT scanners. Data of 214 scan parameters, dose indicators; volume CT dose index (CTDIvol) and dose-length product (DLP) of patients who underwent cranial CT scans were collected. The data were grouped into three, with respect to the slice numbers of 24, 28, and 32. Tube voltage (kVp) and slice thickness were constant; (110 kVp and 4.8 mm, respectively), at variable tube currents (mAs). A one-sample t-test was used to compare the dose indicator values of the hospital protocol with a recommended protocol. Scan parameters were optimized for radiation dose against image quality. \nResults: Increased mAs resulted in increased CTDIvol and DLP at constant kVp and slice thickness. Moreover, dose indicators recorded the lowest and highest values at the slice numbers of 24 and 32, respectively. An increase in slice numbers affected dose indicators. Dose indicators recorded significant reduction (P<0.001) in comparison to the recommended protocol. \nConclusion: Optimization of CT protocol considers radiation dose and image quality. Radiologists adopted protocols acquired with lower scan parameters and dose indicators lower than the recommended achievable dose limit of 58 mGy.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"322-330"},"PeriodicalIF":0.0000,"publicationDate":"2020-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality\",\"authors\":\"N. Okoro, V. Changizi, E. Jazayeri, F. Pak\",\"doi\":\"10.22038/IJMP.2019.40477.1565\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. \\nMaterial and Methods: A poly(methyl methacrylate) phantom was used for quality control test on CT scanners. Data of 214 scan parameters, dose indicators; volume CT dose index (CTDIvol) and dose-length product (DLP) of patients who underwent cranial CT scans were collected. The data were grouped into three, with respect to the slice numbers of 24, 28, and 32. Tube voltage (kVp) and slice thickness were constant; (110 kVp and 4.8 mm, respectively), at variable tube currents (mAs). A one-sample t-test was used to compare the dose indicator values of the hospital protocol with a recommended protocol. Scan parameters were optimized for radiation dose against image quality. \\nResults: Increased mAs resulted in increased CTDIvol and DLP at constant kVp and slice thickness. Moreover, dose indicators recorded the lowest and highest values at the slice numbers of 24 and 32, respectively. An increase in slice numbers affected dose indicators. Dose indicators recorded significant reduction (P<0.001) in comparison to the recommended protocol. \\nConclusion: Optimization of CT protocol considers radiation dose and image quality. Radiologists adopted protocols acquired with lower scan parameters and dose indicators lower than the recommended achievable dose limit of 58 mGy.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\"17 1\",\"pages\":\"322-330\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2019.40477.1565\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.40477.1565","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

摘要

引言:由于患者剂量的增加,计算机断层扫描(CT)的快速使用备受关注。CT方案的优化是减少剂量的一个重要问题。本研究旨在优化颅骨CT的辐射剂量,并评估在辐射剂量减少的情况下图像质量的变化。材料和方法:采用聚甲基丙烯酸甲酯体模对CT扫描仪进行质量控制测试。214个扫描参数、剂量指标的数据;收集接受颅骨CT扫描的患者的体积CT剂量指数(CTDIvol)和剂量-长度乘积(DLP)。根据切片编号24、28和32,将数据分为三组。管电压(kVp)和切片厚度是恒定的;(分别为110 kVp和4.8 mm)。单样本t检验用于比较医院方案和推荐方案的剂量指标值。扫描参数针对辐射剂量和图像质量进行了优化。结果:在恒定的kVp和切片厚度下,mAs的增加导致CTDIvol和DLP的增加。此外,剂量指示器分别在切片编号24和32处记录了最低值和最高值。切片数量的增加影响了剂量指标。与推荐方案相比,剂量指标显著降低(P<0.001)。结论:CT方案的优化考虑了辐射剂量和图像质量。放射科医生采用的方案具有较低的扫描参数和剂量指标,低于58 mGy的推荐可实现剂量限制。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality
Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(methyl methacrylate) phantom was used for quality control test on CT scanners. Data of 214 scan parameters, dose indicators; volume CT dose index (CTDIvol) and dose-length product (DLP) of patients who underwent cranial CT scans were collected. The data were grouped into three, with respect to the slice numbers of 24, 28, and 32. Tube voltage (kVp) and slice thickness were constant; (110 kVp and 4.8 mm, respectively), at variable tube currents (mAs). A one-sample t-test was used to compare the dose indicator values of the hospital protocol with a recommended protocol. Scan parameters were optimized for radiation dose against image quality. Results: Increased mAs resulted in increased CTDIvol and DLP at constant kVp and slice thickness. Moreover, dose indicators recorded the lowest and highest values at the slice numbers of 24 and 32, respectively. An increase in slice numbers affected dose indicators. Dose indicators recorded significant reduction (P<0.001) in comparison to the recommended protocol. Conclusion: Optimization of CT protocol considers radiation dose and image quality. Radiologists adopted protocols acquired with lower scan parameters and dose indicators lower than the recommended achievable dose limit of 58 mGy.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信