{"title":"IDREAM迭代重建算法与计算机层析成像中的滤波反投影算法的比较","authors":"Eslam Maamoun","doi":"10.22038/IJMP.2019.41872.1613","DOIUrl":null,"url":null,"abstract":"Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm. \nMaterial and Methods: Two CT protocols (i.e., A: 120 KV /150 mA, FBP; B: 120KV/ (20-150) mAs, IDREAM) to scan water and acrylic phantoms. A number of 40 patients were assigned to two CT protocols (C: n=20, 120KV/160 ±10 mAs, FBP and D: n=20, 120 KV/ (30-150 mAs, IDREAM), the two groups (C and D) were then referred to abdomen and pelvis CT scan (Sinovision, insitum 16) with contrast. Image quality parameters, dose calculations were measured for all groups (i.e., A, B, C, and D). \nResults: Group B had a highly significant SNR with less significant noise (P 0.05) in acrylic phantom, as compared to group A. CTDIvol (A: 13.94 mGy ; B: 6.91 mGy , P<0.05 ) and DLP (A:501.76 mGy.cm ; B :248.88 mGy.cm). Noise and SNR were significantly different (P<0.05) in group D against C. CTDIvol (C: 30.3±5.2 mGy ; D : 15.4 ±2.7 mGy, P<0.05 ) , DLP (C:544±100 mGy.cm; D :272.3±50.3 mGy.cm ,P<0.05) and the effective dose (C:8.1±1.5 mSv; D :4.08±0.75 mSv,P<0.05) \nConclusion: The results of the present study were indicative of the feasibility of IDREAMas an iterative reconstruction algorithm.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"170-174"},"PeriodicalIF":0.0000,"publicationDate":"2020-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparing IDREAM as an Iterative Reconstruction Algorithm against In Filtered Back Projection in Computed Tomography\",\"authors\":\"Eslam Maamoun\",\"doi\":\"10.22038/IJMP.2019.41872.1613\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm. \\nMaterial and Methods: Two CT protocols (i.e., A: 120 KV /150 mA, FBP; B: 120KV/ (20-150) mAs, IDREAM) to scan water and acrylic phantoms. A number of 40 patients were assigned to two CT protocols (C: n=20, 120KV/160 ±10 mAs, FBP and D: n=20, 120 KV/ (30-150 mAs, IDREAM), the two groups (C and D) were then referred to abdomen and pelvis CT scan (Sinovision, insitum 16) with contrast. Image quality parameters, dose calculations were measured for all groups (i.e., A, B, C, and D). \\nResults: Group B had a highly significant SNR with less significant noise (P 0.05) in acrylic phantom, as compared to group A. CTDIvol (A: 13.94 mGy ; B: 6.91 mGy , P<0.05 ) and DLP (A:501.76 mGy.cm ; B :248.88 mGy.cm). Noise and SNR were significantly different (P<0.05) in group D against C. CTDIvol (C: 30.3±5.2 mGy ; D : 15.4 ±2.7 mGy, P<0.05 ) , DLP (C:544±100 mGy.cm; D :272.3±50.3 mGy.cm ,P<0.05) and the effective dose (C:8.1±1.5 mSv; D :4.08±0.75 mSv,P<0.05) \\nConclusion: The results of the present study were indicative of the feasibility of IDREAMas an iterative reconstruction algorithm.\",\"PeriodicalId\":14613,\"journal\":{\"name\":\"Iranian Journal of Medical Physics\",\"volume\":\"17 1\",\"pages\":\"170-174\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22038/IJMP.2019.41872.1613\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Health Professions\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.41872.1613","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
Comparing IDREAM as an Iterative Reconstruction Algorithm against In Filtered Back Projection in Computed Tomography
Introduction: Recent studies of Computed Tomography (CT) conducted on patient dose reduction have recommended using an iterative reconstruction algorithm and mA (mili-Ampere) dose modulation. The current study aimed to evaluate Iterative Dose Reduction Algorithm (IDREAM) as an iterative reconstruction algorithm.
Material and Methods: Two CT protocols (i.e., A: 120 KV /150 mA, FBP; B: 120KV/ (20-150) mAs, IDREAM) to scan water and acrylic phantoms. A number of 40 patients were assigned to two CT protocols (C: n=20, 120KV/160 ±10 mAs, FBP and D: n=20, 120 KV/ (30-150 mAs, IDREAM), the two groups (C and D) were then referred to abdomen and pelvis CT scan (Sinovision, insitum 16) with contrast. Image quality parameters, dose calculations were measured for all groups (i.e., A, B, C, and D).
Results: Group B had a highly significant SNR with less significant noise (P 0.05) in acrylic phantom, as compared to group A. CTDIvol (A: 13.94 mGy ; B: 6.91 mGy , P<0.05 ) and DLP (A:501.76 mGy.cm ; B :248.88 mGy.cm). Noise and SNR were significantly different (P<0.05) in group D against C. CTDIvol (C: 30.3±5.2 mGy ; D : 15.4 ±2.7 mGy, P<0.05 ) , DLP (C:544±100 mGy.cm; D :272.3±50.3 mGy.cm ,P<0.05) and the effective dose (C:8.1±1.5 mSv; D :4.08±0.75 mSv,P<0.05)
Conclusion: The results of the present study were indicative of the feasibility of IDREAMas an iterative reconstruction algorithm.
期刊介绍:
Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.