{"title":"基于cmm的人工全髋关节置换术中回收聚乙烯杯体积变化评估方法","authors":"Xijin Hua,&nbsp;Junyan Li","doi":"10.1049/bsbt.2019.0011","DOIUrl":null,"url":null,"abstract":"<div>\n <p>Wear occurring at the bearing surface and the consequent generation of wear debris has been identified as the primary cause of aseptic loosening in metal-on-polyethylene (MoP) hip joint replacements. The accurate estimation of volume change in polyethylene cups due to creep and wear is, therefore, an important step for identifying the cause of failure and improving the longevity of MoP prosthesis. The purposes of this study were to present and apply a co-ordinate measuring machine (CMM)-based method for assessing the volume change of retrieved components due to wear and creep by using a combination of CMM data and a bespoke computer programme. The method was firstly validated against the standard gravimetric technique, and then applied to four retrieved polyethylene cups for wear assessment and analysis. The results show that the volume changes calculated using the present method match well with those assessed through the gravimetric technique. The CMM-based method presented in the study is capable of effectively and reliably determining the volume change and characterising the wear patch of retrieved components from MoP hip joint replacements.</p>\n </div>","PeriodicalId":52235,"journal":{"name":"Biosurface and Biotribology","volume":"6 2","pages":"37-42"},"PeriodicalIF":1.6000,"publicationDate":"2020-04-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ietresearch.onlinelibrary.wiley.com/doi/epdf/10.1049/bsbt.2019.0011","citationCount":"0","resultStr":"{\"title\":\"CMM-based method for assessing the volume change of retrieved polyethylene cups in MoP total hip replacements\",\"authors\":\"Xijin Hua,&nbsp;Junyan Li\",\"doi\":\"10.1049/bsbt.2019.0011\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n <p>Wear occurring at the bearing surface and the consequent generation of wear debris has been identified as the primary cause of aseptic loosening in metal-on-polyethylene (MoP) hip joint replacements. The accurate estimation of volume change in polyethylene cups due to creep and wear is, therefore, an important step for identifying the cause of failure and improving the longevity of MoP prosthesis. The purposes of this study were to present and apply a co-ordinate measuring machine (CMM)-based method for assessing the volume change of retrieved components due to wear and creep by using a combination of CMM data and a bespoke computer programme. The method was firstly validated against the standard gravimetric technique, and then applied to four retrieved polyethylene cups for wear assessment and analysis. The results show that the volume changes calculated using the present method match well with those assessed through the gravimetric technique. The CMM-based method presented in the study is capable of effectively and reliably determining the volume change and characterising the wear patch of retrieved components from MoP hip joint replacements.</p>\\n </div>\",\"PeriodicalId\":52235,\"journal\":{\"name\":\"Biosurface and Biotribology\",\"volume\":\"6 2\",\"pages\":\"37-42\"},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2020-04-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ietresearch.onlinelibrary.wiley.com/doi/epdf/10.1049/bsbt.2019.0011\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biosurface and Biotribology\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1049/bsbt.2019.0011\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biosurface and Biotribology","FirstCategoryId":"1087","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1049/bsbt.2019.0011","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

在轴承表面发生的磨损以及随之而来的磨损碎片的产生已被确定为金属对聚乙烯(MoP)髋关节置换术中无菌性松动的主要原因。因此,准确估计聚乙烯杯因蠕变和磨损引起的体积变化是确定失效原因和提高MoP假体使用寿命的重要步骤。本研究的目的是提出并应用一种基于坐标测量机(CMM)的方法,通过使用CMM数据和定制计算机程序的组合来评估由于磨损和蠕变而产生的回收部件的体积变化。该方法首先与标准重量法进行了验证,然后应用于四个回收的聚乙烯杯进行磨损评估和分析。结果表明,用该方法计算的体积变化与用重量法计算的结果吻合较好。本研究提出的基于cmm的方法能够有效可靠地确定MoP髋关节置换术中回收部件的体积变化和表征磨损片。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

CMM-based method for assessing the volume change of retrieved polyethylene cups in MoP total hip replacements

CMM-based method for assessing the volume change of retrieved polyethylene cups in MoP total hip replacements

Wear occurring at the bearing surface and the consequent generation of wear debris has been identified as the primary cause of aseptic loosening in metal-on-polyethylene (MoP) hip joint replacements. The accurate estimation of volume change in polyethylene cups due to creep and wear is, therefore, an important step for identifying the cause of failure and improving the longevity of MoP prosthesis. The purposes of this study were to present and apply a co-ordinate measuring machine (CMM)-based method for assessing the volume change of retrieved components due to wear and creep by using a combination of CMM data and a bespoke computer programme. The method was firstly validated against the standard gravimetric technique, and then applied to four retrieved polyethylene cups for wear assessment and analysis. The results show that the volume changes calculated using the present method match well with those assessed through the gravimetric technique. The CMM-based method presented in the study is capable of effectively and reliably determining the volume change and characterising the wear patch of retrieved components from MoP hip joint replacements.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biosurface and Biotribology
Biosurface and Biotribology Engineering-Mechanical Engineering
CiteScore
1.70
自引率
0.00%
发文量
27
审稿时长
11 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信