血液中醛固酮浓度不同检测方法的一致性

Q4 Medicine
Fengfan Zheng, Ying Song, Shumin Yang, Jinbo Hu, Wenwen He, Zhihong Wang, Q. Cheng
{"title":"血液中醛固酮浓度不同检测方法的一致性","authors":"Fengfan Zheng, Ying Song, Shumin Yang, Jinbo Hu, Wenwen He, Zhihong Wang, Q. Cheng","doi":"10.3760/CMA.J.ISSN.1000-6699.2019.11.003","DOIUrl":null,"url":null,"abstract":"Objective \nTo evaluate the consistency of different methods for detecting aldosterone concentration in blood and to establish a reference interval of serum aldosterone concentration by liquid chromatography-tandem mass spectrometry(LC-MS/MS). \n \n \nMethods \nConcentrations of blood aldosterone were measured by LC-MS/MS, chemiluminescent assays (Diasorin, Domestic A and B systems) and radioimmunoassay (RIA) in 138 healthy adults, 67 patients with essential hypertension and 23 patients with primary aldosteronism. \n \n \nResults \nAldosterone concentrations measured by various methods were quiet different(P 0.05) between CLIA(Domestic B) and LC-MS/MS, 0.338(P<0.01) as well as between RIA and LC-MS/MS. Bland-Altman analysis for the measurements showed poor consistency among methods for aldosterone concentrations. The reference interval was 15.2-222.2 pg/ml for serum aldosterone concentrations by LC-MS/MS. \n \n \nConclusions \nThere are significant differences of blood aldosterone concentrations determined by different methods. Clinically, a specified reference interval might be needed while using different methods. \n \n \nKey words: \nAldosterone; Liquid chromatography-tandem mass spectrometry; Chemiluminescent assays; Radioimmunoassay","PeriodicalId":10120,"journal":{"name":"中华内分泌代谢杂志","volume":"35 1","pages":"934-938"},"PeriodicalIF":0.0000,"publicationDate":"2019-11-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"On the consistency of different methods for detecting aldosterone concentration in blood\",\"authors\":\"Fengfan Zheng, Ying Song, Shumin Yang, Jinbo Hu, Wenwen He, Zhihong Wang, Q. Cheng\",\"doi\":\"10.3760/CMA.J.ISSN.1000-6699.2019.11.003\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo evaluate the consistency of different methods for detecting aldosterone concentration in blood and to establish a reference interval of serum aldosterone concentration by liquid chromatography-tandem mass spectrometry(LC-MS/MS). \\n \\n \\nMethods \\nConcentrations of blood aldosterone were measured by LC-MS/MS, chemiluminescent assays (Diasorin, Domestic A and B systems) and radioimmunoassay (RIA) in 138 healthy adults, 67 patients with essential hypertension and 23 patients with primary aldosteronism. \\n \\n \\nResults \\nAldosterone concentrations measured by various methods were quiet different(P 0.05) between CLIA(Domestic B) and LC-MS/MS, 0.338(P<0.01) as well as between RIA and LC-MS/MS. Bland-Altman analysis for the measurements showed poor consistency among methods for aldosterone concentrations. The reference interval was 15.2-222.2 pg/ml for serum aldosterone concentrations by LC-MS/MS. \\n \\n \\nConclusions \\nThere are significant differences of blood aldosterone concentrations determined by different methods. Clinically, a specified reference interval might be needed while using different methods. \\n \\n \\nKey words: \\nAldosterone; Liquid chromatography-tandem mass spectrometry; Chemiluminescent assays; Radioimmunoassay\",\"PeriodicalId\":10120,\"journal\":{\"name\":\"中华内分泌代谢杂志\",\"volume\":\"35 1\",\"pages\":\"934-938\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-11-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华内分泌代谢杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.1000-6699.2019.11.003\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华内分泌代谢杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.1000-6699.2019.11.003","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1

摘要

目的评价不同检测血中醛固酮浓度方法的一致性,建立液相色谱-串联质谱(LC-MS/MS)测定血清醛固酮浓度的参考区间。方法对138例健康成人、67例原发性高血压患者和23例原发性醛固酮增多症患者,采用LC-MS/MS、化学发光法(Diasorin、国产A、B系统)和放射免疫法(RIA)测定血醛固酮浓度。结果CLIA(Domestic B)法测定的醛固酮浓度与LC-MS/MS法测定的醛固酮浓度差异无统计学意义(P 0.05), RIA法测定的醛固酮浓度与LC-MS/MS法测定的醛固酮浓度差异无统计学意义(P<0.01);Bland-Altman分析表明,醛固酮浓度测定方法的一致性较差。LC-MS/MS测定血清醛固酮浓度的参考区间为15.2 ~ 222.2 pg/ml。结论不同方法测定的血醛固酮浓度有显著性差异。临床上,使用不同的方法可能需要一个特定的参考区间。关键词:醛固酮;液相色谱-串联质谱法;化学发光分析;放射免疫检定法
本文章由计算机程序翻译,如有差异,请以英文原文为准。
On the consistency of different methods for detecting aldosterone concentration in blood
Objective To evaluate the consistency of different methods for detecting aldosterone concentration in blood and to establish a reference interval of serum aldosterone concentration by liquid chromatography-tandem mass spectrometry(LC-MS/MS). Methods Concentrations of blood aldosterone were measured by LC-MS/MS, chemiluminescent assays (Diasorin, Domestic A and B systems) and radioimmunoassay (RIA) in 138 healthy adults, 67 patients with essential hypertension and 23 patients with primary aldosteronism. Results Aldosterone concentrations measured by various methods were quiet different(P 0.05) between CLIA(Domestic B) and LC-MS/MS, 0.338(P<0.01) as well as between RIA and LC-MS/MS. Bland-Altman analysis for the measurements showed poor consistency among methods for aldosterone concentrations. The reference interval was 15.2-222.2 pg/ml for serum aldosterone concentrations by LC-MS/MS. Conclusions There are significant differences of blood aldosterone concentrations determined by different methods. Clinically, a specified reference interval might be needed while using different methods. Key words: Aldosterone; Liquid chromatography-tandem mass spectrometry; Chemiluminescent assays; Radioimmunoassay
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中华内分泌代谢杂志
中华内分泌代谢杂志 Medicine-Endocrinology, Diabetes and Metabolism
CiteScore
0.60
自引率
0.00%
发文量
7243
期刊介绍: The Chinese Journal of Endocrinology and Metabolism was founded in July 1985. It is a senior academic journal in the field of endocrinology and metabolism sponsored by the Chinese Medical Association. The journal aims to be the "Chinese broadcaster of new knowledge on endocrinology and metabolism worldwide". It reports leading scientific research results and clinical diagnosis and treatment experience in endocrinology and metabolism and related fields, as well as basic theoretical research that has a guiding role in endocrinology and metabolism clinics and is closely integrated with clinics. The journal is a core journal of Chinese science and technology (a statistical source journal of Chinese science and technology papers), and is included in Chinese and foreign statistical source journal databases such as the Chinese Science and Technology Papers and Citation Database, Chemical Abstracts, and Scopus.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信