生物材料碳纤维增强聚醚酮(CFR-PEEK)牙种植体结构行为的双线性数值分析:有限元分析

IF 0.6 Q4 DENTISTRY, ORAL SURGERY & MEDICINE
Miguel Martinez-Mondragon, G. Urriolagoitia-Sosa, B. Romero-Ángeles, Juan Pérez-Partida, Itzel Cruz-Olivares, G. Urriolagoitia-Calderón
{"title":"生物材料碳纤维增强聚醚酮(CFR-PEEK)牙种植体结构行为的双线性数值分析:有限元分析","authors":"Miguel Martinez-Mondragon, G. Urriolagoitia-Sosa, B. Romero-Ángeles, Juan Pérez-Partida, Itzel Cruz-Olivares, G. Urriolagoitia-Calderón","doi":"10.4103/denthyp.denthyp_42_23","DOIUrl":null,"url":null,"abstract":"Introduction: This research aims to compare the distribution of stresses and general displacements between different dental implant abutments against a healthy tooth, as well as the mechanical behavior of the carbon fiber-reinforced polyether-ether-ketone (CFR-PEEK) material under load conditions. Methods: A biomodel of a healthy tooth was implemented by computed tomography (CT), considering three essential parts of the tooth (enamel, dentin, and pulp). Three different dental abutments were produced using the SolidWorks computer program. Applying the finite element method (FEM), a numerical evaluation was performed by introducing a critical load of 550 N and then unloading it to 0 N taking into consideration the behavior of the material (titanium and CFR-PEEK) as bilinear, isotropic, and homogeneous. Results: The difference in stress and total displacement between the dental implant (titanium and CFR-PEEK) and the healthy tooth was significant, going from critical stress values of 1087 to 324 MPa. Conclusion: When removing the load from the CFR-PEEK material, it presented residual stresses because the material passed its elastic limit despite this, demonstrating a better mechanical behavior than titanium.","PeriodicalId":43354,"journal":{"name":"Dental Hypotheses","volume":"14 1","pages":"45 - 48"},"PeriodicalIF":0.6000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"Bilinear Numerical Analysis of the Structural Behavior of a Dental Implant Applied as a Biomaterial Carbon Fiber Reinforced Polyether-Ether-Ketone (CFR-PEEK): A Finite Element Analysis\",\"authors\":\"Miguel Martinez-Mondragon, G. Urriolagoitia-Sosa, B. Romero-Ángeles, Juan Pérez-Partida, Itzel Cruz-Olivares, G. Urriolagoitia-Calderón\",\"doi\":\"10.4103/denthyp.denthyp_42_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: This research aims to compare the distribution of stresses and general displacements between different dental implant abutments against a healthy tooth, as well as the mechanical behavior of the carbon fiber-reinforced polyether-ether-ketone (CFR-PEEK) material under load conditions. Methods: A biomodel of a healthy tooth was implemented by computed tomography (CT), considering three essential parts of the tooth (enamel, dentin, and pulp). Three different dental abutments were produced using the SolidWorks computer program. Applying the finite element method (FEM), a numerical evaluation was performed by introducing a critical load of 550 N and then unloading it to 0 N taking into consideration the behavior of the material (titanium and CFR-PEEK) as bilinear, isotropic, and homogeneous. Results: The difference in stress and total displacement between the dental implant (titanium and CFR-PEEK) and the healthy tooth was significant, going from critical stress values of 1087 to 324 MPa. Conclusion: When removing the load from the CFR-PEEK material, it presented residual stresses because the material passed its elastic limit despite this, demonstrating a better mechanical behavior than titanium.\",\"PeriodicalId\":43354,\"journal\":{\"name\":\"Dental Hypotheses\",\"volume\":\"14 1\",\"pages\":\"45 - 48\"},\"PeriodicalIF\":0.6000,\"publicationDate\":\"2023-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Dental Hypotheses\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/denthyp.denthyp_42_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dental Hypotheses","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/denthyp.denthyp_42_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 2

摘要

前言:本研究旨在比较不同种植体基台对健康牙齿的应力分布和一般位移,以及碳纤维增强聚醚醚酮(CFR-PEEK)材料在载荷条件下的力学行为。方法:通过计算机断层扫描(CT)建立健康牙齿的生物模型,考虑牙齿的三个基本部分(牙釉质、牙本质和牙髓)。使用SolidWorks计算机程序制作了三种不同的牙基。考虑材料(钛和CFR-PEEK)的双线性、各向同性和均匀性,采用有限元方法,引入临界载荷550 N,然后卸载至0 N,进行了数值评估。结果:种植体(钛和CFR-PEEK)与健康牙齿的应力和总位移差异显著,临界应力值为1087 ~ 324 MPa。结论:CFR-PEEK材料在卸除载荷时,由于材料超过了弹性极限,仍然存在残余应力,表现出比钛更好的力学行为。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Bilinear Numerical Analysis of the Structural Behavior of a Dental Implant Applied as a Biomaterial Carbon Fiber Reinforced Polyether-Ether-Ketone (CFR-PEEK): A Finite Element Analysis
Introduction: This research aims to compare the distribution of stresses and general displacements between different dental implant abutments against a healthy tooth, as well as the mechanical behavior of the carbon fiber-reinforced polyether-ether-ketone (CFR-PEEK) material under load conditions. Methods: A biomodel of a healthy tooth was implemented by computed tomography (CT), considering three essential parts of the tooth (enamel, dentin, and pulp). Three different dental abutments were produced using the SolidWorks computer program. Applying the finite element method (FEM), a numerical evaluation was performed by introducing a critical load of 550 N and then unloading it to 0 N taking into consideration the behavior of the material (titanium and CFR-PEEK) as bilinear, isotropic, and homogeneous. Results: The difference in stress and total displacement between the dental implant (titanium and CFR-PEEK) and the healthy tooth was significant, going from critical stress values of 1087 to 324 MPa. Conclusion: When removing the load from the CFR-PEEK material, it presented residual stresses because the material passed its elastic limit despite this, demonstrating a better mechanical behavior than titanium.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Dental Hypotheses
Dental Hypotheses DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
1.00
自引率
0.00%
发文量
16
审稿时长
21 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信