{"title":"技术报告:在第三方ClearCheck软件的帮助下,对Eclipse治疗计划系统的光子和电子剂量计算算法进行有效的恒定性评估。","authors":"Eric Lobb MS","doi":"10.1016/j.meddos.2023.08.009","DOIUrl":null,"url":null,"abstract":"<div><p>Our institutional approach to performing periodic constancy checks on dose calculations for the Varian Eclipse treatment planning system<span> is described, including details of how the Radformation ClearCheck software is leveraged for highly efficient results analysis. Photon- and electron-specific synthetic phantoms are created which allow all modality-specific fields to be spatially separated and calculated in a single plan, with field-specific structures encompassing individual irradiated volumes facilitating field-specific dose distribution<span> evaluations. Plan comparison templates in ClearCheck allow for the near-instantaneous evaluation of 116 and 165 individual calculation metrics for photon and electron fields, respectively. The methodology is further applied to the Radformation ClearCalc finite-size pencil beam and electron algorithms for secondary algorithm validation. The total time required for constancy evaluations of 6 primary algorithms and 4 secondary algorithms is less than one hour, with much of that time used for passive dose calculation.</span></span></p></div>","PeriodicalId":49837,"journal":{"name":"Medical Dosimetry","volume":"49 2","pages":"Pages 81-84"},"PeriodicalIF":1.1000,"publicationDate":"2023-09-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Technical Report: Efficient constancy evaluation of the Eclipse treatment planning system's photon and electron dose calculation algorithms aided by the third-party ClearCheck software\",\"authors\":\"Eric Lobb MS\",\"doi\":\"10.1016/j.meddos.2023.08.009\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Our institutional approach to performing periodic constancy checks on dose calculations for the Varian Eclipse treatment planning system<span> is described, including details of how the Radformation ClearCheck software is leveraged for highly efficient results analysis. Photon- and electron-specific synthetic phantoms are created which allow all modality-specific fields to be spatially separated and calculated in a single plan, with field-specific structures encompassing individual irradiated volumes facilitating field-specific dose distribution<span> evaluations. Plan comparison templates in ClearCheck allow for the near-instantaneous evaluation of 116 and 165 individual calculation metrics for photon and electron fields, respectively. The methodology is further applied to the Radformation ClearCalc finite-size pencil beam and electron algorithms for secondary algorithm validation. The total time required for constancy evaluations of 6 primary algorithms and 4 secondary algorithms is less than one hour, with much of that time used for passive dose calculation.</span></span></p></div>\",\"PeriodicalId\":49837,\"journal\":{\"name\":\"Medical Dosimetry\",\"volume\":\"49 2\",\"pages\":\"Pages 81-84\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2023-09-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical Dosimetry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0958394723000742\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical Dosimetry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0958394723000742","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
Technical Report: Efficient constancy evaluation of the Eclipse treatment planning system's photon and electron dose calculation algorithms aided by the third-party ClearCheck software
Our institutional approach to performing periodic constancy checks on dose calculations for the Varian Eclipse treatment planning system is described, including details of how the Radformation ClearCheck software is leveraged for highly efficient results analysis. Photon- and electron-specific synthetic phantoms are created which allow all modality-specific fields to be spatially separated and calculated in a single plan, with field-specific structures encompassing individual irradiated volumes facilitating field-specific dose distribution evaluations. Plan comparison templates in ClearCheck allow for the near-instantaneous evaluation of 116 and 165 individual calculation metrics for photon and electron fields, respectively. The methodology is further applied to the Radformation ClearCalc finite-size pencil beam and electron algorithms for secondary algorithm validation. The total time required for constancy evaluations of 6 primary algorithms and 4 secondary algorithms is less than one hour, with much of that time used for passive dose calculation.
期刊介绍:
Medical Dosimetry, the official journal of the American Association of Medical Dosimetrists, is the key source of information on new developments for the medical dosimetrist. Practical and comprehensive in coverage, the journal features original contributions and review articles by medical dosimetrists, oncologists, physicists, and radiation therapy technologists on clinical applications and techniques of external beam, interstitial, intracavitary and intraluminal irradiation in cancer management. Articles dealing primarily with physics will be reviewed by a specially appointed team of experts in the field.