99mtc -高techate SPECT-CT在甲状腺组织体积测量中的作用:幻影研究和临床病例系列。

IF 1.5 4区 医学 Q3 PHARMACOLOGY & PHARMACY
Enrico Calandri, Maria Teresa Giraudo, Roberta Sirovich, Antonella Ostan, Mirco Pultrone, Viviana Frantellizzi, Lucia Conversano, Paolo Bagnasacco, Sonya Gallina, Giuseppe De Vincentis
{"title":"99mtc -高techate SPECT-CT在甲状腺组织体积测量中的作用:幻影研究和临床病例系列。","authors":"Enrico Calandri,&nbsp;Maria Teresa Giraudo,&nbsp;Roberta Sirovich,&nbsp;Antonella Ostan,&nbsp;Mirco Pultrone,&nbsp;Viviana Frantellizzi,&nbsp;Lucia Conversano,&nbsp;Paolo Bagnasacco,&nbsp;Sonya Gallina,&nbsp;Giuseppe De Vincentis","doi":"10.2174/1874471015666220111145550","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>An accurate measurement of the target volume is of primary importance in theragnostics of hyperthyroidism.</p><p><strong>Objective: </strong>Our purpose was to evaluate the accuracy of a threshold-based isocontour extraction procedure for thyroid tissue volumetry from SPECT-CT.</p><p><strong>Methods: </strong>Cylindrical vials with a fixed volume of 99mTcO4 at different activities were inserted into a neck phantom in two different thickness settings. Images were acquired by orienting the phantom in different positions, i.e., 40 planar images and 40 SPECT-CT. The fixed values of the isocontouring threshold for SPECT and SPECT-CT were calculated by means of linear and spline regression models. Mean, Median, Standard Deviation, Standard Error, Mean Absolute Percentage Error and Root Mean-Square Error were computed. Any difference between the planar method, SPECT and SPECT-CT and the effective volume was evaluated by means of ANOVA and posthoc tests. Moreover, planar and SPECT-CT acquisitions were performed in 8 patients with hyperthyroidism, considering relevant percentage differences greater than > 20% from the CT gold standard.</p><p><strong>Results: </strong>Concerning phantom studies, the planar method shows higher values of each parameter than the other two methods. SPECT-CT shows lower variability. However, no significant differences were observed between SPECT and SPECT-CT measurements. In patients, relevant differences were found in 7 out of 9 lesions with the planar method, in 6 lesions with SPECT, but in only one with SPECT-CT.</p><p><strong>Conclusion: </strong>Our study confirms the superiority of SPECT in volume measurement if compared with the planar method. A more accurate measurement can be obtained from SPECT-CT.</p>","PeriodicalId":10991,"journal":{"name":"Current radiopharmaceuticals","volume":"15 3","pages":"205-217"},"PeriodicalIF":1.5000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Usefulness of 99mTc-Pertechnetate SPECT-CT in Thyroid Tissue Volumetry: Phantom Studies and a Clinical Case Series.\",\"authors\":\"Enrico Calandri,&nbsp;Maria Teresa Giraudo,&nbsp;Roberta Sirovich,&nbsp;Antonella Ostan,&nbsp;Mirco Pultrone,&nbsp;Viviana Frantellizzi,&nbsp;Lucia Conversano,&nbsp;Paolo Bagnasacco,&nbsp;Sonya Gallina,&nbsp;Giuseppe De Vincentis\",\"doi\":\"10.2174/1874471015666220111145550\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>An accurate measurement of the target volume is of primary importance in theragnostics of hyperthyroidism.</p><p><strong>Objective: </strong>Our purpose was to evaluate the accuracy of a threshold-based isocontour extraction procedure for thyroid tissue volumetry from SPECT-CT.</p><p><strong>Methods: </strong>Cylindrical vials with a fixed volume of 99mTcO4 at different activities were inserted into a neck phantom in two different thickness settings. Images were acquired by orienting the phantom in different positions, i.e., 40 planar images and 40 SPECT-CT. The fixed values of the isocontouring threshold for SPECT and SPECT-CT were calculated by means of linear and spline regression models. Mean, Median, Standard Deviation, Standard Error, Mean Absolute Percentage Error and Root Mean-Square Error were computed. Any difference between the planar method, SPECT and SPECT-CT and the effective volume was evaluated by means of ANOVA and posthoc tests. Moreover, planar and SPECT-CT acquisitions were performed in 8 patients with hyperthyroidism, considering relevant percentage differences greater than > 20% from the CT gold standard.</p><p><strong>Results: </strong>Concerning phantom studies, the planar method shows higher values of each parameter than the other two methods. SPECT-CT shows lower variability. However, no significant differences were observed between SPECT and SPECT-CT measurements. In patients, relevant differences were found in 7 out of 9 lesions with the planar method, in 6 lesions with SPECT, but in only one with SPECT-CT.</p><p><strong>Conclusion: </strong>Our study confirms the superiority of SPECT in volume measurement if compared with the planar method. A more accurate measurement can be obtained from SPECT-CT.</p>\",\"PeriodicalId\":10991,\"journal\":{\"name\":\"Current radiopharmaceuticals\",\"volume\":\"15 3\",\"pages\":\"205-217\"},\"PeriodicalIF\":1.5000,\"publicationDate\":\"2022-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Current radiopharmaceuticals\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.2174/1874471015666220111145550\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"PHARMACOLOGY & PHARMACY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Current radiopharmaceuticals","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2174/1874471015666220111145550","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0

摘要

背景:靶体积的准确测量在甲状腺机能亢进的诊断中是至关重要的。目的:我们的目的是评估基于阈值的等轮廓提取程序甲状腺组织体积从SPECT-CT的准确性。方法:将固定体积的99mTcO4圆柱形小瓶以不同的活度插入两种不同厚度的颈模中。将幻体定位在不同位置获取图像,即平面图像40张,SPECT-CT图像40张。利用线性和样条回归模型计算SPECT和SPECT- ct等轮廓阈值的固定值。计算平均值、中位数、标准差、标准误差、平均绝对百分比误差和均方根误差。平面法、SPECT和SPECT- ct与有效体积之间的差异通过方差分析和事后检验进行评价。此外,考虑到与CT金标准的相关百分比差异大于> 20%,对8例甲状腺功能亢进患者进行了平面和SPECT-CT采集。结果:在幻影研究中,平面法的各参数值均高于其他两种方法。SPECT-CT显示较低的变异性。然而,在SPECT和SPECT- ct测量之间没有观察到显著差异。在9例患者中,平面法有7例发现相关差异,SPECT有6例发现相关差异,SPECT- ct只有1例。结论:本研究证实了SPECT在体积测量方面的优越性。SPECT-CT可以获得更精确的测量结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Usefulness of 99mTc-Pertechnetate SPECT-CT in Thyroid Tissue Volumetry: Phantom Studies and a Clinical Case Series.

Background: An accurate measurement of the target volume is of primary importance in theragnostics of hyperthyroidism.

Objective: Our purpose was to evaluate the accuracy of a threshold-based isocontour extraction procedure for thyroid tissue volumetry from SPECT-CT.

Methods: Cylindrical vials with a fixed volume of 99mTcO4 at different activities were inserted into a neck phantom in two different thickness settings. Images were acquired by orienting the phantom in different positions, i.e., 40 planar images and 40 SPECT-CT. The fixed values of the isocontouring threshold for SPECT and SPECT-CT were calculated by means of linear and spline regression models. Mean, Median, Standard Deviation, Standard Error, Mean Absolute Percentage Error and Root Mean-Square Error were computed. Any difference between the planar method, SPECT and SPECT-CT and the effective volume was evaluated by means of ANOVA and posthoc tests. Moreover, planar and SPECT-CT acquisitions were performed in 8 patients with hyperthyroidism, considering relevant percentage differences greater than > 20% from the CT gold standard.

Results: Concerning phantom studies, the planar method shows higher values of each parameter than the other two methods. SPECT-CT shows lower variability. However, no significant differences were observed between SPECT and SPECT-CT measurements. In patients, relevant differences were found in 7 out of 9 lesions with the planar method, in 6 lesions with SPECT, but in only one with SPECT-CT.

Conclusion: Our study confirms the superiority of SPECT in volume measurement if compared with the planar method. A more accurate measurement can be obtained from SPECT-CT.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Current radiopharmaceuticals
Current radiopharmaceuticals PHARMACOLOGY & PHARMACY-
CiteScore
3.20
自引率
4.30%
发文量
43
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信