{"title":"[68Ge/68Ga发电机长期质量控制试验]。","authors":"Sho Tomita, Kei Higashikawa, Satoshi Ueno, Yuki Mizuno, Shusaku Tazawa, Yuji Kuge","doi":"10.18893/kakuigaku.tr.2102","DOIUrl":null,"url":null,"abstract":"<p><p>Quality of the <sup>68</sup>Ga solution eluted from a <sup>68</sup>Ge/<sup>68</sup>Ga generator (Galli Eo<sup>®</sup>) was evaluated. Elution was performed 488 times from 21 to 484 days (ca. 15 months) after the calibration date. The eluted <sup>68</sup>Ga activity was 329 MBq to 1,148 MBq, which decreased with the decay of <sup>68</sup>Ge, but no significant change was observed in the elution yields (62.8±2.2%). The half-life of the eluted radioactivity was 67.8±0.1 min, the radionuclide purity of <sup>68</sup>Ga was ≥99.9%, the <sup>68</sup>Ge breakthrough was 0.000024 ±0.000004%, and the radiochemical purity of <sup>68</sup>Ga<sup>3+</sup> was 99.7±0.2%. Fe and Zn were detected as metal impurities in the eluent, but both were ≤10 μg/GBq. The endotoxin concentration of the eluate was ≤5 EU/mL, and the eluent passed the sterility test. These results show that the generator can stably provide <sup>68</sup>Ga solution over a 15-month period.</p>","PeriodicalId":17738,"journal":{"name":"Kaku igaku. The Japanese journal of nuclear medicine","volume":"58 1","pages":"47-58"},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Long-term Quality Control Test of a <sup>68</sup>Ge/<sup>68</sup>Ga Generator].\",\"authors\":\"Sho Tomita, Kei Higashikawa, Satoshi Ueno, Yuki Mizuno, Shusaku Tazawa, Yuji Kuge\",\"doi\":\"10.18893/kakuigaku.tr.2102\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Quality of the <sup>68</sup>Ga solution eluted from a <sup>68</sup>Ge/<sup>68</sup>Ga generator (Galli Eo<sup>®</sup>) was evaluated. Elution was performed 488 times from 21 to 484 days (ca. 15 months) after the calibration date. The eluted <sup>68</sup>Ga activity was 329 MBq to 1,148 MBq, which decreased with the decay of <sup>68</sup>Ge, but no significant change was observed in the elution yields (62.8±2.2%). The half-life of the eluted radioactivity was 67.8±0.1 min, the radionuclide purity of <sup>68</sup>Ga was ≥99.9%, the <sup>68</sup>Ge breakthrough was 0.000024 ±0.000004%, and the radiochemical purity of <sup>68</sup>Ga<sup>3+</sup> was 99.7±0.2%. Fe and Zn were detected as metal impurities in the eluent, but both were ≤10 μg/GBq. The endotoxin concentration of the eluate was ≤5 EU/mL, and the eluent passed the sterility test. These results show that the generator can stably provide <sup>68</sup>Ga solution over a 15-month period.</p>\",\"PeriodicalId\":17738,\"journal\":{\"name\":\"Kaku igaku. The Japanese journal of nuclear medicine\",\"volume\":\"58 1\",\"pages\":\"47-58\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Kaku igaku. The Japanese journal of nuclear medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18893/kakuigaku.tr.2102\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kaku igaku. The Japanese journal of nuclear medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18893/kakuigaku.tr.2102","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
[Long-term Quality Control Test of a 68Ge/68Ga Generator].
Quality of the 68Ga solution eluted from a 68Ge/68Ga generator (Galli Eo®) was evaluated. Elution was performed 488 times from 21 to 484 days (ca. 15 months) after the calibration date. The eluted 68Ga activity was 329 MBq to 1,148 MBq, which decreased with the decay of 68Ge, but no significant change was observed in the elution yields (62.8±2.2%). The half-life of the eluted radioactivity was 67.8±0.1 min, the radionuclide purity of 68Ga was ≥99.9%, the 68Ge breakthrough was 0.000024 ±0.000004%, and the radiochemical purity of 68Ga3+ was 99.7±0.2%. Fe and Zn were detected as metal impurities in the eluent, but both were ≤10 μg/GBq. The endotoxin concentration of the eluate was ≤5 EU/mL, and the eluent passed the sterility test. These results show that the generator can stably provide 68Ga solution over a 15-month period.
期刊介绍:
The main scope of JSNM is the appropriate development of the application of radioactive substances and stable nuclides in medicine. Towards this scope, JSNM promotes the exchange of idea and information, and researches in nuclear medicine. JSNM are interested in the international collaboration for the development of nuclear medicine. Through the activities, JSNM contributes to the medicine, as well as humanity, not only in Japan but also in any countries over the world.