{"title":"依维莫司和紫杉醇洗脱支架治疗小冠状动脉疾病血管反应的差异:光学相干断层扫描分析","authors":"Kenya Nasu ,&nbsp;Yuji Oikawa ,&nbsp;Tadanori Aizawa ,&nbsp;Takahiko Suzuki ,&nbsp;on behalf of SACRA and PLUM registries investigators","doi":"10.1016/j.ijchv.2014.01.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Previous clinical trials have demonstrated the clinical and angiographic superiority of everolimus-eluting stents (EES) compared with paclitaxel-eluting stents (PES) in the small coronary vessel. However, the differences of vascular response including assessment of morphological neointimal tissue (NIT) characteristics using optical coherence tomography (OCT) have not been fully evaluated. The aim of this study is to evaluate the differences of chronic vascular response following small coronary stenting between EES and PES using OCT.</p></div><div><h3>Methods and results</h3><p>A prospective OCT examination at 9 month follow-up was performed for 50 small coronary artery diseases (50 patients) treated by a single 2.5 mm stent for each stent group. Cross-sectional area within stent segments were analyzed at an interval of 1 mm. NIT structure (homogeneous or heterogeneous) was evaluated for qualitative assessment. Homogeneous NIT was observed significantly higher and heterogeneous NIT was lower in EES compared with PES (93% vs. 89%; p = 0.003, 6.5% vs. 10.3%; p = 0.002, respectively). The frequencies of exposed and malapposed struts were lower in EES compared with PES (0.2% vs. 1.7%; p = 0.0001, 0.1% vs. 0.3%; p = 0.001, respectively). NIT eccentricity index and NIT area were lower in EES compared with PES (0.69 ± 0.08 vs. 0.76 ± 0.10; p = 0.001, 0.97 ± 0.42 mm<sup>2</sup> vs. 1.27 ± 0.67 mm<sup>2</sup>; p = 0.01, respectively).</p></div><div><h3>Conclusions</h3><p>A favorable vascular response was observed after EES implantation compared with PES for small coronary artery disease. In addition, the characteristics of NIT after EES implantation were more stable than PES at 9 month follow-up.</p></div>","PeriodicalId":90542,"journal":{"name":"International journal of cardiology. Heart & vessels","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2014-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.ijchv.2014.01.001","citationCount":"2","resultStr":"{\"title\":\"Difference of vascular response between everolimus- and paclitaxel-eluting stents for small coronary artery disease: Optical coherence tomography analysis\",\"authors\":\"Kenya Nasu ,&nbsp;Yuji Oikawa ,&nbsp;Tadanori Aizawa ,&nbsp;Takahiko Suzuki ,&nbsp;on behalf of SACRA and PLUM registries investigators\",\"doi\":\"10.1016/j.ijchv.2014.01.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Previous clinical trials have demonstrated the clinical and angiographic superiority of everolimus-eluting stents (EES) compared with paclitaxel-eluting stents (PES) in the small coronary vessel. However, the differences of vascular response including assessment of morphological neointimal tissue (NIT) characteristics using optical coherence tomography (OCT) have not been fully evaluated. The aim of this study is to evaluate the differences of chronic vascular response following small coronary stenting between EES and PES using OCT.</p></div><div><h3>Methods and results</h3><p>A prospective OCT examination at 9 month follow-up was performed for 50 small coronary artery diseases (50 patients) treated by a single 2.5 mm stent for each stent group. Cross-sectional area within stent segments were analyzed at an interval of 1 mm. NIT structure (homogeneous or heterogeneous) was evaluated for qualitative assessment. Homogeneous NIT was observed significantly higher and heterogeneous NIT was lower in EES compared with PES (93% vs. 89%; p = 0.003, 6.5% vs. 10.3%; p = 0.002, respectively). The frequencies of exposed and malapposed struts were lower in EES compared with PES (0.2% vs. 1.7%; p = 0.0001, 0.1% vs. 0.3%; p = 0.001, respectively). NIT eccentricity index and NIT area were lower in EES compared with PES (0.69 ± 0.08 vs. 0.76 ± 0.10; p = 0.001, 0.97 ± 0.42 mm<sup>2</sup> vs. 1.27 ± 0.67 mm<sup>2</sup>; p = 0.01, respectively).</p></div><div><h3>Conclusions</h3><p>A favorable vascular response was observed after EES implantation compared with PES for small coronary artery disease. In addition, the characteristics of NIT after EES implantation were more stable than PES at 9 month follow-up.</p></div>\",\"PeriodicalId\":90542,\"journal\":{\"name\":\"International journal of cardiology. Heart & vessels\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/j.ijchv.2014.01.001\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of cardiology. Heart & vessels\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2214763214000030\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of cardiology. Heart & vessels","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2214763214000030","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

摘要

背景:先前的临床试验已经证明依维莫司洗脱支架(EES)与紫杉醇洗脱支架(PES)在冠状动脉小血管中的临床和血管造影优势。然而,血管反应的差异,包括使用光学相干断层扫描(OCT)评估形态学新生内膜组织(NIT)特征还没有得到充分的评估。本研究的目的是评估EES和PES在小冠状动脉支架置入术后慢性血管反应的差异。方法和结果对50例(50例)小冠状动脉疾病(每个支架组使用单个2.5 mm支架)进行了9个月的前瞻性OCT检查。每隔1mm分析支架段内的横截面积。对NIT结构(同质或异质)进行定性评价。与PES相比,EES的均匀性NIT明显较高,而非均匀性NIT较低(93% vs 89%;P = 0.003, 6.5% vs. 10.3%;P = 0.002)。与PES相比,EES中暴露和错位支撑的频率较低(0.2% vs. 1.7%;P = 0.0001, 0.1% vs. 0.3%;P = 0.001)。EES组NIT偏心率指数和NIT面积较PES组低(0.69±0.08∶0.76±0.10);P = 0.001, 0.97±0.42 mm2 vs. 1.27±0.67 mm2;P = 0.01)。结论EES植入术治疗小冠状动脉病变较PES有较好的血管反应。此外,在9个月的随访中,EES植入后NIT的特征比PES更稳定。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

Difference of vascular response between everolimus- and paclitaxel-eluting stents for small coronary artery disease: Optical coherence tomography analysis

Difference of vascular response between everolimus- and paclitaxel-eluting stents for small coronary artery disease: Optical coherence tomography analysis

Difference of vascular response between everolimus- and paclitaxel-eluting stents for small coronary artery disease: Optical coherence tomography analysis

Difference of vascular response between everolimus- and paclitaxel-eluting stents for small coronary artery disease: Optical coherence tomography analysis

Background

Previous clinical trials have demonstrated the clinical and angiographic superiority of everolimus-eluting stents (EES) compared with paclitaxel-eluting stents (PES) in the small coronary vessel. However, the differences of vascular response including assessment of morphological neointimal tissue (NIT) characteristics using optical coherence tomography (OCT) have not been fully evaluated. The aim of this study is to evaluate the differences of chronic vascular response following small coronary stenting between EES and PES using OCT.

Methods and results

A prospective OCT examination at 9 month follow-up was performed for 50 small coronary artery diseases (50 patients) treated by a single 2.5 mm stent for each stent group. Cross-sectional area within stent segments were analyzed at an interval of 1 mm. NIT structure (homogeneous or heterogeneous) was evaluated for qualitative assessment. Homogeneous NIT was observed significantly higher and heterogeneous NIT was lower in EES compared with PES (93% vs. 89%; p = 0.003, 6.5% vs. 10.3%; p = 0.002, respectively). The frequencies of exposed and malapposed struts were lower in EES compared with PES (0.2% vs. 1.7%; p = 0.0001, 0.1% vs. 0.3%; p = 0.001, respectively). NIT eccentricity index and NIT area were lower in EES compared with PES (0.69 ± 0.08 vs. 0.76 ± 0.10; p = 0.001, 0.97 ± 0.42 mm2 vs. 1.27 ± 0.67 mm2; p = 0.01, respectively).

Conclusions

A favorable vascular response was observed after EES implantation compared with PES for small coronary artery disease. In addition, the characteristics of NIT after EES implantation were more stable than PES at 9 month follow-up.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信