{"title":"降低注射放射性对乳腺分子成像断层合成成像质量的影响。","authors":"Olivia Sullivan,&nbsp;Zongyi Gong,&nbsp;Kelly Klanian,&nbsp;Tushita Patel,&nbsp;Mark B Williams","doi":"10.1007/978-3-642-31271-7_39","DOIUrl":null,"url":null,"abstract":"<p><p>This study's objective is to compare image quality in 3-D molecular breast imaging tomosynthesis (MBIT) with that in planar molecular breast imaging (MBI) over a range of breast radioactivity concentrations. Using gelatin and point source phantoms lesion contrast, lesion signal-to-noise ratio (SNR) and spatial resolution were compared for a range of lesion sizes and depths. For both MBI and MBIT, lesion contrast is essentially constant with changing activity while SNR decreases by a factor of 1.5 - 2 between 100% and 25% activity levels. For nearly all lesion sizes and locations contrast and SNR are significantly higher for MBIT than MBI, potentially permitting greater reductions in injected dose. Spatial resolution in MBI is dependent on lesion depth but independent of lesion location with MBIT. Reconstructed MBIT spatial resolution is substantially better than that in the projection images, suggesting future use of higher sensitivity collimators for even further reductions in injected activity.</p>","PeriodicalId":92005,"journal":{"name":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","volume":"7361 ","pages":"300-307"},"PeriodicalIF":0.0000,"publicationDate":"2012-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5565232/pdf/nihms424638.pdf","citationCount":"4","resultStr":"{\"title\":\"The Impact of Reduced Injected Radioactivity on Image Quality of Molecular Breast Imaging Tomosynthesis.\",\"authors\":\"Olivia Sullivan,&nbsp;Zongyi Gong,&nbsp;Kelly Klanian,&nbsp;Tushita Patel,&nbsp;Mark B Williams\",\"doi\":\"10.1007/978-3-642-31271-7_39\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This study's objective is to compare image quality in 3-D molecular breast imaging tomosynthesis (MBIT) with that in planar molecular breast imaging (MBI) over a range of breast radioactivity concentrations. Using gelatin and point source phantoms lesion contrast, lesion signal-to-noise ratio (SNR) and spatial resolution were compared for a range of lesion sizes and depths. For both MBI and MBIT, lesion contrast is essentially constant with changing activity while SNR decreases by a factor of 1.5 - 2 between 100% and 25% activity levels. For nearly all lesion sizes and locations contrast and SNR are significantly higher for MBIT than MBI, potentially permitting greater reductions in injected dose. Spatial resolution in MBI is dependent on lesion depth but independent of lesion location with MBIT. Reconstructed MBIT spatial resolution is substantially better than that in the projection images, suggesting future use of higher sensitivity collimators for even further reductions in injected activity.</p>\",\"PeriodicalId\":92005,\"journal\":{\"name\":\"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)\",\"volume\":\"7361 \",\"pages\":\"300-307\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2012-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5565232/pdf/nihms424638.pdf\",\"citationCount\":\"4\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/978-3-642-31271-7_39\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/978-3-642-31271-7_39","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

摘要

本研究的目的是比较三维分子乳房成像断层合成(MBIT)与平面分子乳房成像(MBI)在乳房放射性浓度范围内的图像质量。利用明胶和点源图像对比病变,比较不同大小和深度的病变信噪比(SNR)和空间分辨率。对于MBI和mbbi,病变对比度随着活动的变化基本不变,而在100%和25%的活动水平之间,信噪比下降了1.5 - 2倍。对于几乎所有的病变大小和部位,MBI的对比度和信噪比都明显高于MBI,这可能使注射剂量的减少幅度更大。MBI的空间分辨率依赖于病灶深度,而与病灶位置无关。重建的MBIT空间分辨率明显优于投影图像,这表明未来使用更高灵敏度的准直器来进一步降低注入活度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Impact of Reduced Injected Radioactivity on Image Quality of Molecular Breast Imaging Tomosynthesis.

This study's objective is to compare image quality in 3-D molecular breast imaging tomosynthesis (MBIT) with that in planar molecular breast imaging (MBI) over a range of breast radioactivity concentrations. Using gelatin and point source phantoms lesion contrast, lesion signal-to-noise ratio (SNR) and spatial resolution were compared for a range of lesion sizes and depths. For both MBI and MBIT, lesion contrast is essentially constant with changing activity while SNR decreases by a factor of 1.5 - 2 between 100% and 25% activity levels. For nearly all lesion sizes and locations contrast and SNR are significantly higher for MBIT than MBI, potentially permitting greater reductions in injected dose. Spatial resolution in MBI is dependent on lesion depth but independent of lesion location with MBIT. Reconstructed MBIT spatial resolution is substantially better than that in the projection images, suggesting future use of higher sensitivity collimators for even further reductions in injected activity.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信