{"title":"在高动态范围和高灵敏度模式下用于乳腺断层合成的cmos - cmos x射线探测器的探测量子效率。","authors":"Tushita Patel,&nbsp;Kelly Klanian,&nbsp;Zongyi Gong,&nbsp;Mark B Williams","doi":"10.1007/978-3-642-31271-7_11","DOIUrl":null,"url":null,"abstract":"<p><p>The spatial frequency dependent detective quantum efficiency (DQE) of a CsI-CMOS x-ray detector was measured in two operating modes: a high dynamic range (HDR) mode and a high sensitivity (HS) mode. DQE calculations were performed using the IEC-62220-1-2 Standard. For detector entrance air kerma values between ~7 µGy and 60 µGy the DQE is similar in either HDR mode or HS mode, with a value of ~0.7 at low frequency and ~ 0.15 - 0.20 at the Nyquist frequency f<sub>N</sub> = 6.7 mm<sup>-1</sup>. In HDR mode the DQE remains virtually constant for operation with K<sub>a</sub> values between ~7 µGy and 119 µGy but decreases for K<sub>a</sub> levels below ~ 7 µGy. In HS mode the DQE is approximately constant over the full range of entrance air kerma tested between 1.7 µGy and 60 µGy but kerma values above ~75 µGy produce hard saturation. Quantum limited operation in HS mode for entrance kerma as small as 1.7 µGy makes it possible to use a large number of low dose views to improve angular sampling and decrease acquisition time.</p>","PeriodicalId":92005,"journal":{"name":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","volume":"7361 ","pages":"80-87"},"PeriodicalIF":0.0000,"publicationDate":"2012-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/978-3-642-31271-7_11","citationCount":"11","resultStr":"{\"title\":\"Detective Quantum Efficiency of a CsI-CMOS X-ray Detector for Breast Tomosynthesis Operating in High Dynamic Range and High Sensitivity Modes.\",\"authors\":\"Tushita Patel,&nbsp;Kelly Klanian,&nbsp;Zongyi Gong,&nbsp;Mark B Williams\",\"doi\":\"10.1007/978-3-642-31271-7_11\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The spatial frequency dependent detective quantum efficiency (DQE) of a CsI-CMOS x-ray detector was measured in two operating modes: a high dynamic range (HDR) mode and a high sensitivity (HS) mode. DQE calculations were performed using the IEC-62220-1-2 Standard. For detector entrance air kerma values between ~7 µGy and 60 µGy the DQE is similar in either HDR mode or HS mode, with a value of ~0.7 at low frequency and ~ 0.15 - 0.20 at the Nyquist frequency f<sub>N</sub> = 6.7 mm<sup>-1</sup>. In HDR mode the DQE remains virtually constant for operation with K<sub>a</sub> values between ~7 µGy and 119 µGy but decreases for K<sub>a</sub> levels below ~ 7 µGy. In HS mode the DQE is approximately constant over the full range of entrance air kerma tested between 1.7 µGy and 60 µGy but kerma values above ~75 µGy produce hard saturation. Quantum limited operation in HS mode for entrance kerma as small as 1.7 µGy makes it possible to use a large number of low dose views to improve angular sampling and decrease acquisition time.</p>\",\"PeriodicalId\":92005,\"journal\":{\"name\":\"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)\",\"volume\":\"7361 \",\"pages\":\"80-87\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2012-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1007/978-3-642-31271-7_11\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/978-3-642-31271-7_11\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Breast imaging : 11th International Workshop, IWDM 2012, Philadelphia, PA, USA, July 8-11, 2012 : proceedings. International Workshop on Breast Imaging (11th : 2012 : Philadelphia, Pa.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/978-3-642-31271-7_11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

摘要

在高动态范围(HDR)和高灵敏度(HS)两种工作模式下,测量了cmos x射线探测器的空间频率相关探测量子效率(DQE)。DQE计算使用IEC-62220-1-2标准执行。在HDR模式和HS模式下,探测器入口空气kerma值在~7 ~ 60µGy之间时,DQE值在低频为~0.7,在奈奎斯特频率fN = 6.7 mm-1时为~ 0.15 ~ 0.20。在HDR模式下,当Ka值介于~7µGy和119µGy之间时,DQE几乎保持不变,但当Ka水平低于~7µGy时,DQE会下降。在HS模式下,DQE在1.7µGy到60µGy之间的整个测试入口空气kerma范围内近似恒定,但kerma值高于~75µGy会产生硬饱和。在HS模式下,入口kerma小至1.7µGy的量子限制操作使得使用大量低剂量视图来改善角度采样并减少采集时间成为可能。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Detective Quantum Efficiency of a CsI-CMOS X-ray Detector for Breast Tomosynthesis Operating in High Dynamic Range and High Sensitivity Modes.

The spatial frequency dependent detective quantum efficiency (DQE) of a CsI-CMOS x-ray detector was measured in two operating modes: a high dynamic range (HDR) mode and a high sensitivity (HS) mode. DQE calculations were performed using the IEC-62220-1-2 Standard. For detector entrance air kerma values between ~7 µGy and 60 µGy the DQE is similar in either HDR mode or HS mode, with a value of ~0.7 at low frequency and ~ 0.15 - 0.20 at the Nyquist frequency fN = 6.7 mm-1. In HDR mode the DQE remains virtually constant for operation with Ka values between ~7 µGy and 119 µGy but decreases for Ka levels below ~ 7 µGy. In HS mode the DQE is approximately constant over the full range of entrance air kerma tested between 1.7 µGy and 60 µGy but kerma values above ~75 µGy produce hard saturation. Quantum limited operation in HS mode for entrance kerma as small as 1.7 µGy makes it possible to use a large number of low dose views to improve angular sampling and decrease acquisition time.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信