{"title":"带有锥形连接的钛锆合金植入体在植入成形术后的强度。","authors":"","doi":"10.1016/j.prosdent.2022.08.015","DOIUrl":null,"url":null,"abstract":"<div><h3>Statement of problem</h3><p>Peri-implantitis occurs around dental implants, and implantoplasty has been used to address this ongoing disease; however, the changes to the physical properties of an implant after implantoplasty have not been well documented.</p></div><div><h3>Purpose</h3><p>The purpose of this in vitro study was to determine the effect of implantoplasty on fracture strength and the load required for plastic deformation after cyclic fatigue on dental implants.</p></div><div><h3>Material and methods</h3><p>Twenty-six titanium/zirconium (TiZr) alloy implants (Roxolid Bone Level Implant; 4.1×10 mm) were embedded with 50% thread exposure and divided into 4 groups based on whether they had implantoplasty treatment by using different diamond rotary instruments and/or cyclic loading at 250 N for 2 million cycles: C0 (control, no cyclic loading), T0 (test, no cyclic loading), CM (control, cyclic loading), and TM (test, cyclic loading). After implantoplasty and/or cyclic loading, all implants underwent a load-to-failure test. The maximum fracture strength (FS) and load required for the onset of plastic deformation (PD) were recorded in Newtons. One-way ANOVA and nonparametric comparisons with control by using the Dunn and Wilcoxon method for joint ranking were used for statistical analysis.</p></div><div><h3>Results</h3><p>The mean ±standard deviation FS for C0, CM, T0, and TM was 1465.2 ±86.4 N, 1480.7 ±64.1 N, 1299.3 ±123.8 N, and 1252.1 ±85.7 N, respectively. The mean ±standard deviation load for onset of PD for C0, CM, T0, and TM was 860.2 ±88.1 N, 797.0 ±130.5 N, 776.5 ±181.8 N, and 631.3 ±84.5 N, respectively. The TM group had a significantly lower FS and PD than the C0, CM, and T0 groups (<em>P</em>&lt;.05)</p></div><div><h3>Conclusions</h3><p>Both fracture strength (FS) and the onset of plastic deformation (PD) were significantly reduced after a TiZr alloy implant received implantoplasty and cyclic loading.</p></div>","PeriodicalId":16866,"journal":{"name":"Journal of Prosthetic Dentistry","volume":null,"pages":null},"PeriodicalIF":4.3000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S002239132200511X/pdfft?md5=782b12180d6e5d33b07e358957cfd023&pid=1-s2.0-S002239132200511X-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Strength of titanium-zirconium alloy implants with a conical connection after implantoplasty\",\"authors\":\"\",\"doi\":\"10.1016/j.prosdent.2022.08.015\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Statement of problem</h3><p>Peri-implantitis occurs around dental implants, and implantoplasty has been used to address this ongoing disease; however, the changes to the physical properties of an implant after implantoplasty have not been well documented.</p></div><div><h3>Purpose</h3><p>The purpose of this in vitro study was to determine the effect of implantoplasty on fracture strength and the load required for plastic deformation after cyclic fatigue on dental implants.</p></div><div><h3>Material and methods</h3><p>Twenty-six titanium/zirconium (TiZr) alloy implants (Roxolid Bone Level Implant; 4.1×10 mm) were embedded with 50% thread exposure and divided into 4 groups based on whether they had implantoplasty treatment by using different diamond rotary instruments and/or cyclic loading at 250 N for 2 million cycles: C0 (control, no cyclic loading), T0 (test, no cyclic loading), CM (control, cyclic loading), and TM (test, cyclic loading). After implantoplasty and/or cyclic loading, all implants underwent a load-to-failure test. The maximum fracture strength (FS) and load required for the onset of plastic deformation (PD) were recorded in Newtons. One-way ANOVA and nonparametric comparisons with control by using the Dunn and Wilcoxon method for joint ranking were used for statistical analysis.</p></div><div><h3>Results</h3><p>The mean ±standard deviation FS for C0, CM, T0, and TM was 1465.2 ±86.4 N, 1480.7 ±64.1 N, 1299.3 ±123.8 N, and 1252.1 ±85.7 N, respectively. The mean ±standard deviation load for onset of PD for C0, CM, T0, and TM was 860.2 ±88.1 N, 797.0 ±130.5 N, 776.5 ±181.8 N, and 631.3 ±84.5 N, respectively. The TM group had a significantly lower FS and PD than the C0, CM, and T0 groups (<em>P</em>&lt;.05)</p></div><div><h3>Conclusions</h3><p>Both fracture strength (FS) and the onset of plastic deformation (PD) were significantly reduced after a TiZr alloy implant received implantoplasty and cyclic loading.</p></div>\",\"PeriodicalId\":16866,\"journal\":{\"name\":\"Journal of Prosthetic Dentistry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.3000,\"publicationDate\":\"2024-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S002239132200511X/pdfft?md5=782b12180d6e5d33b07e358957cfd023&pid=1-s2.0-S002239132200511X-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Prosthetic Dentistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S002239132200511X\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Prosthetic Dentistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S002239132200511X","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

问题简介:种植体周围炎发生在牙科种植体周围,种植体成形术已被用于解决这一持续存在的疾病;然而,种植体成形术后种植体物理性质的变化尚未得到很好的记录。目的:本体外研究的目的是确定种植体成形术对牙科种植体断裂强度和循环疲劳后塑性变形所需负荷的影响:将26颗钛/锆(TiZr)合金种植体(Roxolid骨水平种植体;4.1×10 mm)嵌入50%的螺纹暴露中,并根据是否使用不同的金刚石旋转器械进行种植整形治疗和/或以250 N的循环载荷进行200万次循环将其分为4组:C0组(对照组,无循环加载)、T0组(测试组,无循环加载)、CM组(对照组,循环加载)和TM组(测试组,循环加载)。在植入和/或循环加载后,所有植入体都要进行加载至破坏试验。以牛顿为单位记录了最大断裂强度(FS)和塑性变形开始所需的载荷(PD)。统计分析采用了单因素方差分析和非参数比较,并使用 Dunn 和 Wilcoxon 方法与对照组进行联合排序:C0、CM、T0 和 TM 的平均±标准偏差 FS 分别为 1465.2 ±86.4 N、1480.7 ±64.1 N、1299.3 ±123.8 N 和 1252.1 ±85.7 N。C0、CM、T0 和 TM 组 PD 开始时的平均±标准偏差负荷分别为 860.2 ±88.1 N、797.0 ±130.5 N、776.5 ±181.8 N 和 631.3 ±84.5 N。TM 组的 FS 和 PD 明显低于 C0、CM 和 T0 组(P<0.05)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Strength of titanium-zirconium alloy implants with a conical connection after implantoplasty

Statement of problem

Peri-implantitis occurs around dental implants, and implantoplasty has been used to address this ongoing disease; however, the changes to the physical properties of an implant after implantoplasty have not been well documented.

Purpose

The purpose of this in vitro study was to determine the effect of implantoplasty on fracture strength and the load required for plastic deformation after cyclic fatigue on dental implants.

Material and methods

Twenty-six titanium/zirconium (TiZr) alloy implants (Roxolid Bone Level Implant; 4.1×10 mm) were embedded with 50% thread exposure and divided into 4 groups based on whether they had implantoplasty treatment by using different diamond rotary instruments and/or cyclic loading at 250 N for 2 million cycles: C0 (control, no cyclic loading), T0 (test, no cyclic loading), CM (control, cyclic loading), and TM (test, cyclic loading). After implantoplasty and/or cyclic loading, all implants underwent a load-to-failure test. The maximum fracture strength (FS) and load required for the onset of plastic deformation (PD) were recorded in Newtons. One-way ANOVA and nonparametric comparisons with control by using the Dunn and Wilcoxon method for joint ranking were used for statistical analysis.

Results

The mean ±standard deviation FS for C0, CM, T0, and TM was 1465.2 ±86.4 N, 1480.7 ±64.1 N, 1299.3 ±123.8 N, and 1252.1 ±85.7 N, respectively. The mean ±standard deviation load for onset of PD for C0, CM, T0, and TM was 860.2 ±88.1 N, 797.0 ±130.5 N, 776.5 ±181.8 N, and 631.3 ±84.5 N, respectively. The TM group had a significantly lower FS and PD than the C0, CM, and T0 groups (P<.05)

Conclusions

Both fracture strength (FS) and the onset of plastic deformation (PD) were significantly reduced after a TiZr alloy implant received implantoplasty and cyclic loading.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Prosthetic Dentistry
Journal of Prosthetic Dentistry 医学-牙科与口腔外科
CiteScore
7.00
自引率
13.00%
发文量
599
审稿时长
69 days
期刊介绍: The Journal of Prosthetic Dentistry is the leading professional journal devoted exclusively to prosthetic and restorative dentistry. The Journal is the official publication for 24 leading U.S. international prosthodontic organizations. The monthly publication features timely, original peer-reviewed articles on the newest techniques, dental materials, and research findings. The Journal serves prosthodontists and dentists in advanced practice, and features color photos that illustrate many step-by-step procedures. The Journal of Prosthetic Dentistry is included in Index Medicus and CINAHL.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信