{"title":"定性尿促性腺激素测定的性能。","authors":"S Nickmans,&nbsp;P Vermeersch,&nbsp;J Van Eldere,&nbsp;J Billen","doi":"10.1179/0001551214Z.00000000069","DOIUrl":null,"url":null,"abstract":"<p><p>Significant differences in the sensitivity of eight frequently used qualitative urine human chorionic gonadotropin (hCG) tests in Belgium were observed in this study. Although most manufacturers claimed to detect hCG levels as low as 25 mIU/ml, only two out of six tests for home use and one out of two tests for professional use only, achieved the claimed detection limit. According to a survey, we performed among 20 acute care hospitals, 80% of the surveyed hospitals claimed to use these types of hCG analysis in a diagnostic setting. Unsatisfactory performance of these point-of-care testing (POC) assays for urinary hCG could have major consequences in a hospital setting, exposing the early pregnant woman to harmful diagnostic and therapeutic procedures. Although qualitative urine hCG tests are rapid and convenient, determination of hCG in blood remains the gold standard for the diagnosis of pregnancy. </p>","PeriodicalId":48865,"journal":{"name":"Acta Clinica Belgica","volume":"69 4","pages":"277-9"},"PeriodicalIF":1.1000,"publicationDate":"2014-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1179/0001551214Z.00000000069","citationCount":"9","resultStr":"{\"title\":\"Performance of qualitative urinary hCG assays.\",\"authors\":\"S Nickmans,&nbsp;P Vermeersch,&nbsp;J Van Eldere,&nbsp;J Billen\",\"doi\":\"10.1179/0001551214Z.00000000069\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Significant differences in the sensitivity of eight frequently used qualitative urine human chorionic gonadotropin (hCG) tests in Belgium were observed in this study. Although most manufacturers claimed to detect hCG levels as low as 25 mIU/ml, only two out of six tests for home use and one out of two tests for professional use only, achieved the claimed detection limit. According to a survey, we performed among 20 acute care hospitals, 80% of the surveyed hospitals claimed to use these types of hCG analysis in a diagnostic setting. Unsatisfactory performance of these point-of-care testing (POC) assays for urinary hCG could have major consequences in a hospital setting, exposing the early pregnant woman to harmful diagnostic and therapeutic procedures. Although qualitative urine hCG tests are rapid and convenient, determination of hCG in blood remains the gold standard for the diagnosis of pregnancy. </p>\",\"PeriodicalId\":48865,\"journal\":{\"name\":\"Acta Clinica Belgica\",\"volume\":\"69 4\",\"pages\":\"277-9\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2014-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1179/0001551214Z.00000000069\",\"citationCount\":\"9\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta Clinica Belgica\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1179/0001551214Z.00000000069\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2014/5/20 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Clinica Belgica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1179/0001551214Z.00000000069","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2014/5/20 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 9

摘要

在比利时的八个常用的定性尿人绒毛膜促性腺激素(hCG)测试的敏感性显著差异观察在本研究中。虽然大多数制造商声称能检测到低至25 mIU/ml的hCG水平,但只有六分之二的家庭使用测试和两分之一的专业使用测试达到了声称的检测极限。根据一项调查,我们对20家急症护理医院进行了调查,80%的被调查医院声称在诊断设置中使用这些类型的hCG分析。这些尿促性腺激素即时检测(POC)检测的不理想表现可能在医院环境中造成重大后果,使早期孕妇暴露于有害的诊断和治疗程序中。尽管定性尿液hCG测试快速方便,但血液中hCG的测定仍然是诊断妊娠的金标准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Performance of qualitative urinary hCG assays.

Significant differences in the sensitivity of eight frequently used qualitative urine human chorionic gonadotropin (hCG) tests in Belgium were observed in this study. Although most manufacturers claimed to detect hCG levels as low as 25 mIU/ml, only two out of six tests for home use and one out of two tests for professional use only, achieved the claimed detection limit. According to a survey, we performed among 20 acute care hospitals, 80% of the surveyed hospitals claimed to use these types of hCG analysis in a diagnostic setting. Unsatisfactory performance of these point-of-care testing (POC) assays for urinary hCG could have major consequences in a hospital setting, exposing the early pregnant woman to harmful diagnostic and therapeutic procedures. Although qualitative urine hCG tests are rapid and convenient, determination of hCG in blood remains the gold standard for the diagnosis of pregnancy.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Acta Clinica Belgica
Acta Clinica Belgica MEDICINE, GENERAL & INTERNAL-
CiteScore
3.50
自引率
0.00%
发文量
44
期刊介绍: Acta Clinica Belgica: International Journal of Clinical and Laboratory Medicine primarily publishes papers on clinical medicine, clinical chemistry, pathology and molecular biology, provided they describe results which contribute to our understanding of clinical problems or describe new methods applicable to clinical investigation. Readership includes physicians, pathologists, pharmacists and physicians working in non-academic and academic hospitals, practicing internal medicine and its subspecialties.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信