{"title":"呼吸机自动触发:脑死亡诊断中的一个警告。","authors":"Rebecca Dodd-Sullivan, Jamie Quirin, John Newhart","doi":"10.1177/152692481102100211","DOIUrl":null,"url":null,"abstract":"<p><p>The Uniform Determination of Death Act and the American Academy of Neurology's Guidelines for Determination of Death define procedures for diagnosis of brain death before organ procurement for transplantation, including clinical assessment and apnea test. Autotriggering of mechanical ventilators can occur in both brain dead and living patients. A case is reported in which procurement was delayed because ventilator self-cycling was confused for brainstem-mediated respiratory effort. Background knowledge to effectively troubleshoot ventilator self-cycling is offered so that unnecessary delays in brain death testing can be avoided.</p>","PeriodicalId":346415,"journal":{"name":"Progress in Transplantation (Aliso Viejo, Calif.)","volume":" ","pages":"152-5"},"PeriodicalIF":0.0000,"publicationDate":"2011-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1177/152692481102100211","citationCount":"10","resultStr":"{\"title\":\"Ventilator autotriggering: a caution in brain death diagnosis.\",\"authors\":\"Rebecca Dodd-Sullivan, Jamie Quirin, John Newhart\",\"doi\":\"10.1177/152692481102100211\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The Uniform Determination of Death Act and the American Academy of Neurology's Guidelines for Determination of Death define procedures for diagnosis of brain death before organ procurement for transplantation, including clinical assessment and apnea test. Autotriggering of mechanical ventilators can occur in both brain dead and living patients. A case is reported in which procurement was delayed because ventilator self-cycling was confused for brainstem-mediated respiratory effort. Background knowledge to effectively troubleshoot ventilator self-cycling is offered so that unnecessary delays in brain death testing can be avoided.</p>\",\"PeriodicalId\":346415,\"journal\":{\"name\":\"Progress in Transplantation (Aliso Viejo, Calif.)\",\"volume\":\" \",\"pages\":\"152-5\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2011-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1177/152692481102100211\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Progress in Transplantation (Aliso Viejo, Calif.)\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1177/152692481102100211\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Progress in Transplantation (Aliso Viejo, Calif.)","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/152692481102100211","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Ventilator autotriggering: a caution in brain death diagnosis.
The Uniform Determination of Death Act and the American Academy of Neurology's Guidelines for Determination of Death define procedures for diagnosis of brain death before organ procurement for transplantation, including clinical assessment and apnea test. Autotriggering of mechanical ventilators can occur in both brain dead and living patients. A case is reported in which procurement was delayed because ventilator self-cycling was confused for brainstem-mediated respiratory effort. Background knowledge to effectively troubleshoot ventilator self-cycling is offered so that unnecessary delays in brain death testing can be avoided.