6个月大的婴儿植入“心外”内转复除颤器。

T Kriebel, W Ruschewski, T Paul
{"title":"6个月大的婴儿植入“心外”内转复除颤器。","authors":"T Kriebel,&nbsp;W Ruschewski,&nbsp;T Paul","doi":"10.1007/s00392-005-0236-z","DOIUrl":null,"url":null,"abstract":"<p><p>In infants and small children, ICD implantation is a challenge due to technical limitations and a significant number of complications. This report describes ICD implantation in a 6-month-old infant (body weight 5.5 kg). A completely extracardiac defibrillation system was implanted using a transvenous lead subcutaneously in the back below the left scapula as the defibrillation electrode and an active-can device in the right upper abdomen. Defibrillation threshold of implantation was < or =10 J. During the follow-up of 3 months, 8 adequate ICD discharges were noted. The technique described seems feasible to facilitate ICD implantation in small infants.</p>","PeriodicalId":23757,"journal":{"name":"Zeitschrift fur Kardiologie","volume":"94 6","pages":"415-8"},"PeriodicalIF":0.0000,"publicationDate":"2005-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/s00392-005-0236-z","citationCount":"10","resultStr":"{\"title\":\"Implantation of an \\\"extracardiac\\\" internal cardioverter defibrillator in a 6-month-old infant.\",\"authors\":\"T Kriebel,&nbsp;W Ruschewski,&nbsp;T Paul\",\"doi\":\"10.1007/s00392-005-0236-z\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In infants and small children, ICD implantation is a challenge due to technical limitations and a significant number of complications. This report describes ICD implantation in a 6-month-old infant (body weight 5.5 kg). A completely extracardiac defibrillation system was implanted using a transvenous lead subcutaneously in the back below the left scapula as the defibrillation electrode and an active-can device in the right upper abdomen. Defibrillation threshold of implantation was < or =10 J. During the follow-up of 3 months, 8 adequate ICD discharges were noted. The technique described seems feasible to facilitate ICD implantation in small infants.</p>\",\"PeriodicalId\":23757,\"journal\":{\"name\":\"Zeitschrift fur Kardiologie\",\"volume\":\"94 6\",\"pages\":\"415-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2005-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1007/s00392-005-0236-z\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Zeitschrift fur Kardiologie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s00392-005-0236-z\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zeitschrift fur Kardiologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s00392-005-0236-z","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

摘要

由于技术限制和大量并发症,在婴儿和幼儿中,ICD植入是一项挑战。本报告描述了一个6个月大婴儿(体重5.5 kg)的ICD植入。一个完全的心外除颤系统通过经静脉导线在左肩胛骨下方皮下植入作为除颤电极,并在右上腹部植入活性罐装置。植入术除颤阈值<或= 10j,随访3个月,有8例ICD出院。所描述的技术似乎是可行的,以促进ICD植入小婴儿。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Implantation of an "extracardiac" internal cardioverter defibrillator in a 6-month-old infant.

In infants and small children, ICD implantation is a challenge due to technical limitations and a significant number of complications. This report describes ICD implantation in a 6-month-old infant (body weight 5.5 kg). A completely extracardiac defibrillation system was implanted using a transvenous lead subcutaneously in the back below the left scapula as the defibrillation electrode and an active-can device in the right upper abdomen. Defibrillation threshold of implantation was < or =10 J. During the follow-up of 3 months, 8 adequate ICD discharges were noted. The technique described seems feasible to facilitate ICD implantation in small infants.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信