两种无头螺钉固定肱骨小头骨折的比较。

Stuart J Elkowitz, Erik N Kubiak, Daniel Polatsch, John Cooper, Frederick J Kummer, Kenneth J Koval
{"title":"两种无头螺钉固定肱骨小头骨折的比较。","authors":"Stuart J Elkowitz,&nbsp;Erik N Kubiak,&nbsp;Daniel Polatsch,&nbsp;John Cooper,&nbsp;Frederick J Kummer,&nbsp;Kenneth J Koval","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>In order to determine the effects of two different headless screw designs on fixation of simulated capitellum fractures six matched pairs of embalmed humeri had simulated capitellum fractures created. Fixation with Acutrac compression screws was compared to Herbert screws in a matched pair experimental design. All specimens were cyclically tested with simulated physiologic loading. Both displacement of the capitellum as a function of the number of cycles and failure loads were determined. Fixation by the Acutrac screws was significantly more stable than Herbert screws at 2000 cycles, 0.17 mm wersus 1.57 mm (p < 0.02) The Acutrac fixation also had a higher failure load, 154 N versus 118 N (p < 0.05). The Acutrac screws tested in this biomechanical study provided more stable fixation of simulated capitellum fractures than Herbert screws. This appears to be related to the design of these screws.</p>","PeriodicalId":77050,"journal":{"name":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","volume":"61 3-4","pages":"123-6"},"PeriodicalIF":0.0000,"publicationDate":"2003-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of two headless screw designs for fixation of capitellum fractures.\",\"authors\":\"Stuart J Elkowitz,&nbsp;Erik N Kubiak,&nbsp;Daniel Polatsch,&nbsp;John Cooper,&nbsp;Frederick J Kummer,&nbsp;Kenneth J Koval\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In order to determine the effects of two different headless screw designs on fixation of simulated capitellum fractures six matched pairs of embalmed humeri had simulated capitellum fractures created. Fixation with Acutrac compression screws was compared to Herbert screws in a matched pair experimental design. All specimens were cyclically tested with simulated physiologic loading. Both displacement of the capitellum as a function of the number of cycles and failure loads were determined. Fixation by the Acutrac screws was significantly more stable than Herbert screws at 2000 cycles, 0.17 mm wersus 1.57 mm (p < 0.02) The Acutrac fixation also had a higher failure load, 154 N versus 118 N (p < 0.05). The Acutrac screws tested in this biomechanical study provided more stable fixation of simulated capitellum fractures than Herbert screws. This appears to be related to the design of these screws.</p>\",\"PeriodicalId\":77050,\"journal\":{\"name\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"volume\":\"61 3-4\",\"pages\":\"123-6\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

为了确定两种不同的无头螺钉设计对模拟肱骨小头骨折的固定效果,我们制作了6对匹配的肱骨小头骨折标本。在配对实验设计中将Acutrac加压螺钉与Herbert螺钉进行比较。所有的标本都进行了模拟生理负荷的循环试验。确定了小头的位移作为循环次数和破坏载荷的函数。在2000个周期时,Acutrac螺钉比Herbert螺钉更稳定,0.17 mm比1.57 mm (p < 0.02)。Acutrac固定也有更高的失效负荷,154 N比118 N (p < 0.05)。在本生物力学研究中,与Herbert螺钉相比,Acutrac螺钉对模拟肱骨小头骨折的固定更为稳定。这似乎与这些螺钉的设计有关。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparison of two headless screw designs for fixation of capitellum fractures.

In order to determine the effects of two different headless screw designs on fixation of simulated capitellum fractures six matched pairs of embalmed humeri had simulated capitellum fractures created. Fixation with Acutrac compression screws was compared to Herbert screws in a matched pair experimental design. All specimens were cyclically tested with simulated physiologic loading. Both displacement of the capitellum as a function of the number of cycles and failure loads were determined. Fixation by the Acutrac screws was significantly more stable than Herbert screws at 2000 cycles, 0.17 mm wersus 1.57 mm (p < 0.02) The Acutrac fixation also had a higher failure load, 154 N versus 118 N (p < 0.05). The Acutrac screws tested in this biomechanical study provided more stable fixation of simulated capitellum fractures than Herbert screws. This appears to be related to the design of these screws.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信