{"title":"新生儿血糖测定:四种仪器的比较。","authors":"H Schlebusch, M Niesen, M Sorger, I Paffenholz, H Fahnenstich","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Four portable analyzers, HemoCue B-Glucose (I), Accu-Check III (II), One-touch II (III), and Glucometer Elite (IV), with different measuring principles were tested for their suitability for measuring blood glucose in neonates. Precision of all instruments is satisfactory. In the analysis of capillary blood from newborns, two instruments show an excellent accuracy; however, the scatter of the results for instrument (II) is about 1.6 times greater than for instrument (I). The inaccuracy of instruments (III) and (IV) is not acceptable from a clinical point of view. All devices show an influence of hematocrit, the magnitude of which varies between 5% (I) and 12% (III) for every 10% change of hematocrit. Instruments II and IV show that temperature has a marked influence on the readings; the same is true for oxygen in instrument IV. In conclusion, only instrument (I) has met the requirements of accurate and precise blood glucose determinations in neonates.</p>","PeriodicalId":79453,"journal":{"name":"Pediatric pathology & laboratory medicine : journal of the Society for Pediatric Pathology, affiliated with the International Paediatric Pathology Association","volume":"18 1","pages":"41-8"},"PeriodicalIF":0.0000,"publicationDate":"1998-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Blood glucose determinations in newborns: four instruments compared.\",\"authors\":\"H Schlebusch, M Niesen, M Sorger, I Paffenholz, H Fahnenstich\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Four portable analyzers, HemoCue B-Glucose (I), Accu-Check III (II), One-touch II (III), and Glucometer Elite (IV), with different measuring principles were tested for their suitability for measuring blood glucose in neonates. Precision of all instruments is satisfactory. In the analysis of capillary blood from newborns, two instruments show an excellent accuracy; however, the scatter of the results for instrument (II) is about 1.6 times greater than for instrument (I). The inaccuracy of instruments (III) and (IV) is not acceptable from a clinical point of view. All devices show an influence of hematocrit, the magnitude of which varies between 5% (I) and 12% (III) for every 10% change of hematocrit. Instruments II and IV show that temperature has a marked influence on the readings; the same is true for oxygen in instrument IV. In conclusion, only instrument (I) has met the requirements of accurate and precise blood glucose determinations in neonates.</p>\",\"PeriodicalId\":79453,\"journal\":{\"name\":\"Pediatric pathology & laboratory medicine : journal of the Society for Pediatric Pathology, affiliated with the International Paediatric Pathology Association\",\"volume\":\"18 1\",\"pages\":\"41-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pediatric pathology & laboratory medicine : journal of the Society for Pediatric Pathology, affiliated with the International Paediatric Pathology Association\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pediatric pathology & laboratory medicine : journal of the Society for Pediatric Pathology, affiliated with the International Paediatric Pathology Association","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
采用不同的测量原理,对HemoCue B-Glucose (I)、Accu-Check III (II)、One-touch II (III)和Glucometer Elite (IV)四种便携式分析仪进行了新生儿血糖测量的适用性测试。所有仪器的精度都令人满意。在新生儿毛细血管血的分析中,有两种仪器显示出极好的准确性;然而,仪器(II)的结果离散性比仪器(I)大1.6倍。从临床的角度来看,仪器(III)和(IV)的不准确性是不可接受的。所有设备都显示出对血细胞比容的影响,其幅度在5% (I)和12% (III)之间,每10%的血细胞比容变化。仪器二和仪器四显示温度对读数有显著影响;综上所述,只有仪器(I)满足了新生儿准确、精密血糖测定的要求。
Blood glucose determinations in newborns: four instruments compared.
Four portable analyzers, HemoCue B-Glucose (I), Accu-Check III (II), One-touch II (III), and Glucometer Elite (IV), with different measuring principles were tested for their suitability for measuring blood glucose in neonates. Precision of all instruments is satisfactory. In the analysis of capillary blood from newborns, two instruments show an excellent accuracy; however, the scatter of the results for instrument (II) is about 1.6 times greater than for instrument (I). The inaccuracy of instruments (III) and (IV) is not acceptable from a clinical point of view. All devices show an influence of hematocrit, the magnitude of which varies between 5% (I) and 12% (III) for every 10% change of hematocrit. Instruments II and IV show that temperature has a marked influence on the readings; the same is true for oxygen in instrument IV. In conclusion, only instrument (I) has met the requirements of accurate and precise blood glucose determinations in neonates.