{"title":"Spuncrit红外分析仪用于红细胞压积测量的评价。","authors":"M S Weatherall,&nbsp;K M Sherry","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>An open, single centre study was carried out to evaluate the accuracy of the Spuncrit (Micro Diagnostics, Bethlehem, PA, USA) infra-red analyser which can be used for near-patient testing to measure haematocrit and estimate haemoglobin concentration. The primary comparison was with the Sysmex NE1500 (Tao Medical) analyser situated in the main hospital laboratory. Secondary comparison was with the Ciba Corning 288 (Ciba Corning Diagnostics Ltd, Halstead, UK) blood gas analyser currently used for near-patient testing in the Northern General Hospital. A total of 217 samples from 50 patients was analysed. The Pearson's correlation coefficients for haematocrit and haemoglobin concentration between the Spuncrit and Sysmex NE1500 and between the Spuncrit and Ciba Corning 288 were all close, between 0.85 and 0.92. The method of Bland and Altman was used to assess agreement between the results of the Spuncrit and the Sysmex NE1500. The agreement for haematocrit was good with 2 SD of the Spuncrit results being between -5.66 and +4.42% of the measurement from the Sysmex NE1500. In conclusion, the Spuncrit haematocrit measurement agreed well with results from the central laboratory, but the estimated haemoglobin concentrations agreed less well and three reasons are discussed.</p>","PeriodicalId":10285,"journal":{"name":"Clinical and laboratory haematology","volume":"19 3","pages":"183-6"},"PeriodicalIF":0.0000,"publicationDate":"1997-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An evaluation of the Spuncrit infra-red analyser for measurement of haematocrit.\",\"authors\":\"M S Weatherall,&nbsp;K M Sherry\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>An open, single centre study was carried out to evaluate the accuracy of the Spuncrit (Micro Diagnostics, Bethlehem, PA, USA) infra-red analyser which can be used for near-patient testing to measure haematocrit and estimate haemoglobin concentration. The primary comparison was with the Sysmex NE1500 (Tao Medical) analyser situated in the main hospital laboratory. Secondary comparison was with the Ciba Corning 288 (Ciba Corning Diagnostics Ltd, Halstead, UK) blood gas analyser currently used for near-patient testing in the Northern General Hospital. A total of 217 samples from 50 patients was analysed. The Pearson's correlation coefficients for haematocrit and haemoglobin concentration between the Spuncrit and Sysmex NE1500 and between the Spuncrit and Ciba Corning 288 were all close, between 0.85 and 0.92. The method of Bland and Altman was used to assess agreement between the results of the Spuncrit and the Sysmex NE1500. The agreement for haematocrit was good with 2 SD of the Spuncrit results being between -5.66 and +4.42% of the measurement from the Sysmex NE1500. In conclusion, the Spuncrit haematocrit measurement agreed well with results from the central laboratory, but the estimated haemoglobin concentrations agreed less well and three reasons are discussed.</p>\",\"PeriodicalId\":10285,\"journal\":{\"name\":\"Clinical and laboratory haematology\",\"volume\":\"19 3\",\"pages\":\"183-6\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1997-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical and laboratory haematology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical and laboratory haematology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

开展了一项开放的单中心研究,以评估Spuncrit (Micro Diagnostics, Bethlehem, PA, USA)红外分析仪的准确性,该分析仪可用于近患者检测,以测量红细胞压差和估计血红蛋白浓度。主要比较是与位于医院主要实验室的Sysmex NE1500 (Tao Medical)分析仪进行比较。二次比较是与汽巴康宁288(汽巴康宁诊断有限公司,霍尔斯特德,英国)血气分析仪进行比较,目前在北方综合医院用于近病人检测。共分析了50例患者的217份样本。Spuncrit与Sysmex NE1500、Spuncrit与Ciba Corning 288红细胞压积和血红蛋白浓度的Pearson相关系数均接近,在0.85 ~ 0.92之间。使用Bland和Altman方法来评估Spuncrit和Sysmex NE1500结果之间的一致性。红细胞压积的一致性很好,Spuncrit结果的2个标准差在Sysmex NE1500测量值的-5.66和+4.42%之间。总之,Spuncrit红细胞压积测量与中心实验室的结果一致,但估计的血红蛋白浓度一致不太好,并讨论了三个原因。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An evaluation of the Spuncrit infra-red analyser for measurement of haematocrit.

An open, single centre study was carried out to evaluate the accuracy of the Spuncrit (Micro Diagnostics, Bethlehem, PA, USA) infra-red analyser which can be used for near-patient testing to measure haematocrit and estimate haemoglobin concentration. The primary comparison was with the Sysmex NE1500 (Tao Medical) analyser situated in the main hospital laboratory. Secondary comparison was with the Ciba Corning 288 (Ciba Corning Diagnostics Ltd, Halstead, UK) blood gas analyser currently used for near-patient testing in the Northern General Hospital. A total of 217 samples from 50 patients was analysed. The Pearson's correlation coefficients for haematocrit and haemoglobin concentration between the Spuncrit and Sysmex NE1500 and between the Spuncrit and Ciba Corning 288 were all close, between 0.85 and 0.92. The method of Bland and Altman was used to assess agreement between the results of the Spuncrit and the Sysmex NE1500. The agreement for haematocrit was good with 2 SD of the Spuncrit results being between -5.66 and +4.42% of the measurement from the Sysmex NE1500. In conclusion, the Spuncrit haematocrit measurement agreed well with results from the central laboratory, but the estimated haemoglobin concentrations agreed less well and three reasons are discussed.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信