G Pernod, B Bouabana, B Desrousseau, P Mossuz, J M Drevait, B Maljournal, D Rouhan, L Kolodié, B Polack
{"title":"同时评价三种多参数凝血仪:BFA, HEMOLAB和STA。","authors":"G Pernod, B Bouabana, B Desrousseau, P Mossuz, J M Drevait, B Maljournal, D Rouhan, L Kolodié, B Polack","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>In the present paper, three multiparametric coagulation instruments were evaluated with regard to chronometric tests for aPTT (CK-Prest and automated APTT), PT (Recombiplastin and Thromborel), fibrinogen and factors of the prothrombin complex. Analysis of within-run precision and linearity and comparative studies showed the analytical performances of the instruments to differ according to the reagents used and emphasized the difficulty of finding the best compromise between instrument and reagent. On the basis of this study, the mechanical instrument appeared to be more versatile than the optical machines. This conclusion could however be modified after further evaluation of the recent new generation coagulation instruments.</p>","PeriodicalId":19366,"journal":{"name":"Nouvelle revue francaise d'hematologie","volume":"36 6","pages":"409-18"},"PeriodicalIF":0.0000,"publicationDate":"1994-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A simultaneous evaluation of three multiparametric coagulation instruments: BFA, HEMOLAB and STA.\",\"authors\":\"G Pernod, B Bouabana, B Desrousseau, P Mossuz, J M Drevait, B Maljournal, D Rouhan, L Kolodié, B Polack\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In the present paper, three multiparametric coagulation instruments were evaluated with regard to chronometric tests for aPTT (CK-Prest and automated APTT), PT (Recombiplastin and Thromborel), fibrinogen and factors of the prothrombin complex. Analysis of within-run precision and linearity and comparative studies showed the analytical performances of the instruments to differ according to the reagents used and emphasized the difficulty of finding the best compromise between instrument and reagent. On the basis of this study, the mechanical instrument appeared to be more versatile than the optical machines. This conclusion could however be modified after further evaluation of the recent new generation coagulation instruments.</p>\",\"PeriodicalId\":19366,\"journal\":{\"name\":\"Nouvelle revue francaise d'hematologie\",\"volume\":\"36 6\",\"pages\":\"409-18\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1994-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nouvelle revue francaise d'hematologie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nouvelle revue francaise d'hematologie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A simultaneous evaluation of three multiparametric coagulation instruments: BFA, HEMOLAB and STA.
In the present paper, three multiparametric coagulation instruments were evaluated with regard to chronometric tests for aPTT (CK-Prest and automated APTT), PT (Recombiplastin and Thromborel), fibrinogen and factors of the prothrombin complex. Analysis of within-run precision and linearity and comparative studies showed the analytical performances of the instruments to differ according to the reagents used and emphasized the difficulty of finding the best compromise between instrument and reagent. On the basis of this study, the mechanical instrument appeared to be more versatile than the optical machines. This conclusion could however be modified after further evaluation of the recent new generation coagulation instruments.