{"title":"在正常、圆锥角膜和屈光后眼睛中使用标准设备的新型前段光学相干断层扫描的重复性和一致性。","authors":"Gairik Kundu, Naren Shetty, Durgalaxmi Modak, Luci Kaweri, Pooja Khamar, Vishal Arora, Rudy M M A Nuijts, Rohit Shetty","doi":"10.4103/IJO.IJO_805_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess repeatability and agreement of Eyestar 900 (Haag-Streit, Köniz, Switzerland) with Pentacam AXL Wave system (Oculus, Wetzlar, Germany), Anterion (Heidelberg Engineering GmBH, Heidelberg, Germany), and MS-39 (CSO, Florence, Italy) in measuring anterior segment parameters.</p><p><strong>Methods: </strong>Two hundred normal eyes, 50 keratoconus (KC) eyes, and 50 post-refractive surgery eyes underwent three sequential scans on each instrument. The order of scans was randomized. Repeatability was assessed using within-subject standard deviation (Sw) and intraclass correlation coefficient (ICC). Bland-Altman analysis was used to assess the 95% limits of agreement.</p><p><strong>Results: </strong>Eyestar 900 had comparable repeatability to Anterion and MS-39 across groups. ICC was overall >0.8. Pentacam AXL Wave had lower Sw (better repeatability) across all groups. In normal eyes, Eyestar 900 can be used interchangeably with MS-39 and Anterion (P > 0.23). Eyestar 900 and Pentacam AXL Wave did not show agreement on any parameter, except anterior chamber depth (ACD) (P = 0.92). Eyestar 900 in KC eyes can be used interchangeably with MS-39, except for posterior keratometry (pK) and thinnest corneal thickness (TCT) (P < 0.001), and with Anterion, except for TCT and ACD (P < 0.001). In post-refractive eyes, Eyestar 900 can be used interchangeably with MS-39, except for pK (P < 0.001), and with Anterion, except for ACD (P < 0.001). Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably in KC and post-refractive eyes.</p><p><strong>Conclusions: </strong>Newer devices such as Eyestar 900 can be used interchangeably with Anterion and MS-39 in normal eyes and partly in KC and post-refractive eyes. Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably across the various groups. However, these multifunctional hybrid tools of combined topography, AS-OCT, can be helpful in refractive clinics.</p>","PeriodicalId":13329,"journal":{"name":"Indian Journal of Ophthalmology","volume":"73 3","pages":"325-334"},"PeriodicalIF":2.1000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11994167/pdf/","citationCount":"0","resultStr":"{\"title\":\"Repeatability and agreement of a new anterior segment optical coherence tomography with standard devices in normal, keratoconus, and post-refractive eyes.\",\"authors\":\"Gairik Kundu, Naren Shetty, Durgalaxmi Modak, Luci Kaweri, Pooja Khamar, Vishal Arora, Rudy M M A Nuijts, Rohit Shetty\",\"doi\":\"10.4103/IJO.IJO_805_24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To assess repeatability and agreement of Eyestar 900 (Haag-Streit, Köniz, Switzerland) with Pentacam AXL Wave system (Oculus, Wetzlar, Germany), Anterion (Heidelberg Engineering GmBH, Heidelberg, Germany), and MS-39 (CSO, Florence, Italy) in measuring anterior segment parameters.</p><p><strong>Methods: </strong>Two hundred normal eyes, 50 keratoconus (KC) eyes, and 50 post-refractive surgery eyes underwent three sequential scans on each instrument. The order of scans was randomized. Repeatability was assessed using within-subject standard deviation (Sw) and intraclass correlation coefficient (ICC). Bland-Altman analysis was used to assess the 95% limits of agreement.</p><p><strong>Results: </strong>Eyestar 900 had comparable repeatability to Anterion and MS-39 across groups. ICC was overall >0.8. Pentacam AXL Wave had lower Sw (better repeatability) across all groups. In normal eyes, Eyestar 900 can be used interchangeably with MS-39 and Anterion (P > 0.23). Eyestar 900 and Pentacam AXL Wave did not show agreement on any parameter, except anterior chamber depth (ACD) (P = 0.92). Eyestar 900 in KC eyes can be used interchangeably with MS-39, except for posterior keratometry (pK) and thinnest corneal thickness (TCT) (P < 0.001), and with Anterion, except for TCT and ACD (P < 0.001). In post-refractive eyes, Eyestar 900 can be used interchangeably with MS-39, except for pK (P < 0.001), and with Anterion, except for ACD (P < 0.001). Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably in KC and post-refractive eyes.</p><p><strong>Conclusions: </strong>Newer devices such as Eyestar 900 can be used interchangeably with Anterion and MS-39 in normal eyes and partly in KC and post-refractive eyes. Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably across the various groups. However, these multifunctional hybrid tools of combined topography, AS-OCT, can be helpful in refractive clinics.</p>\",\"PeriodicalId\":13329,\"journal\":{\"name\":\"Indian Journal of Ophthalmology\",\"volume\":\"73 3\",\"pages\":\"325-334\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2025-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11994167/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Indian Journal of Ophthalmology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.4103/IJO.IJO_805_24\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/2/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4103/IJO.IJO_805_24","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/26 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
目的:评估Eyestar 900 (Haag-Streit, Köniz,瑞士)与Pentacam AXL Wave系统(Oculus, Wetzlar,德国)、Anterion (Heidelberg Engineering GmBH,海德堡,德国)和MS-39 (CSO,佛罗伦萨,意大利)测量前段参数的重复性和一致性。方法:200只正常眼、50只圆锥角膜(KC)眼和50只屈光术后眼,分别在每台仪器上进行3次连续扫描。扫描的顺序是随机的。使用受试者内标准差(Sw)和类内相关系数(ICC)评估重复性。采用Bland-Altman分析评估95%的一致性限度。结果:Eyestar 900与Anterion和MS-39具有相当的重复性。ICC总体为bb0.8。Pentacam AXL Wave在所有组中具有较低的Sw(更好的重复性)。在正常眼睛中,Eyestar 900可与MS-39和Anterion交替使用(P < 0.23)。Eyestar 900和Pentacam AXL Wave在除前房深度(ACD)外的任何参数上均不一致(P = 0.92)。Eyestar 900在KC眼中可与MS-39互换使用,但用于后角膜测量(pK)和最薄角膜厚度(TCT) (P < 0.001);与Anterion互换使用,但用于TCT和ACD (P < 0.001)。在屈光后眼中,Eyestar 900除pK (P < 0.001)外可与MS-39互换使用,除ACD (P < 0.001)外可与Anterion互换使用。Pentacam AXL Wave和Eyestar 900不能在KC和屈光后眼睛中互换使用。结论:Eyestar 900等新设备可与Anterion和MS-39交替用于正常眼,部分用于KC和屈光后眼。Pentacam AXL Wave和Eyestar 900不能在不同的组中互换使用。然而,这些多功能混合工具的综合地形,AS-OCT,可以帮助屈光临床。
Repeatability and agreement of a new anterior segment optical coherence tomography with standard devices in normal, keratoconus, and post-refractive eyes.
Purpose: To assess repeatability and agreement of Eyestar 900 (Haag-Streit, Köniz, Switzerland) with Pentacam AXL Wave system (Oculus, Wetzlar, Germany), Anterion (Heidelberg Engineering GmBH, Heidelberg, Germany), and MS-39 (CSO, Florence, Italy) in measuring anterior segment parameters.
Methods: Two hundred normal eyes, 50 keratoconus (KC) eyes, and 50 post-refractive surgery eyes underwent three sequential scans on each instrument. The order of scans was randomized. Repeatability was assessed using within-subject standard deviation (Sw) and intraclass correlation coefficient (ICC). Bland-Altman analysis was used to assess the 95% limits of agreement.
Results: Eyestar 900 had comparable repeatability to Anterion and MS-39 across groups. ICC was overall >0.8. Pentacam AXL Wave had lower Sw (better repeatability) across all groups. In normal eyes, Eyestar 900 can be used interchangeably with MS-39 and Anterion (P > 0.23). Eyestar 900 and Pentacam AXL Wave did not show agreement on any parameter, except anterior chamber depth (ACD) (P = 0.92). Eyestar 900 in KC eyes can be used interchangeably with MS-39, except for posterior keratometry (pK) and thinnest corneal thickness (TCT) (P < 0.001), and with Anterion, except for TCT and ACD (P < 0.001). In post-refractive eyes, Eyestar 900 can be used interchangeably with MS-39, except for pK (P < 0.001), and with Anterion, except for ACD (P < 0.001). Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably in KC and post-refractive eyes.
Conclusions: Newer devices such as Eyestar 900 can be used interchangeably with Anterion and MS-39 in normal eyes and partly in KC and post-refractive eyes. Pentacam AXL Wave and Eyestar 900 cannot be used interchangeably across the various groups. However, these multifunctional hybrid tools of combined topography, AS-OCT, can be helpful in refractive clinics.