{"title":"不同品牌假体毛刺行截骨术时骨温度变化的体外研究。","authors":"Ömer Faruk Şarkbay, Ahmet Mihmanli, Hakan Cora","doi":"10.1186/s40729-025-00588-9","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of this study was to investigate in-vitro the temperature changes occurring in the bone during drilling with implant drills manufactured by different companies. Bone blocks obtained from fresh bovine ribs were used in the study. Bone blocks were drilled with drills manufactured by Ankylos, Astra Tech, Nobel Biocare, Bredent and Straumann implant brands at an ambient temperature of 30 ± 2° C under a constant pressure of 2 kg. Two K-type thermocouple sensors were placed on the bone blocks at 5th and 10th mm depths and the temperature changes were measured at a distance of 1 mm from the implant drill. In the study, working models were created under different conditions for implant socket preparation. In group 1, the first time drills were used at 150 rpm without irrigation, in group 2, the first time drills were used at 1200 rpm with 40 ml/min irrigation, in group 3, the 30th time drills were used at 150 rpm without irrigation, and in group 4, the 30th time drills were used at 1200 rpm with 40 ml/min irrigation. All osteotomy procedures were performed with 3.5 ± 0.3 mm diameter burs for a period of 8 s and the temperature values obtained at equal time intervals were recorded. Repeated Measures and Kruskall Wallis-H tests were used for statistical analysis of the data. No significant difference was observed between the implant drills and the temperature changes in the bone during drilling (p < 0.05). None of the groups reached critical temperature values (47° C+) throughout the study. At the 5th and 10th mm depths, the temperature changes in the sensors used were close to each other. It was also calculated that although the average temperatures were close to each other in the non-irrigated and irrigated systems, the difference values obtained by subtracting the initial temperature were significantly higher in the non-irrigated systems. The results showed that implant drills did not cause significant temperature increases in bone blocks depending on the difference in manufacturers (Ankylos, Astra Tech, Nobel Biocare, Bredent, Straumann) and the number of uses. It was also concluded that irrigated and non-irrigated systems are safe as long as they are used under the recommended conditions.</p>","PeriodicalId":14076,"journal":{"name":"International Journal of Implant Dentistry","volume":"11 1","pages":"11"},"PeriodicalIF":3.1000,"publicationDate":"2025-02-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11813834/pdf/","citationCount":"0","resultStr":"{\"title\":\"In-vitro investigation of bone temperature changes in osteotomies performed with different brands of implant burs.\",\"authors\":\"Ömer Faruk Şarkbay, Ahmet Mihmanli, Hakan Cora\",\"doi\":\"10.1186/s40729-025-00588-9\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The aim of this study was to investigate in-vitro the temperature changes occurring in the bone during drilling with implant drills manufactured by different companies. Bone blocks obtained from fresh bovine ribs were used in the study. Bone blocks were drilled with drills manufactured by Ankylos, Astra Tech, Nobel Biocare, Bredent and Straumann implant brands at an ambient temperature of 30 ± 2° C under a constant pressure of 2 kg. Two K-type thermocouple sensors were placed on the bone blocks at 5th and 10th mm depths and the temperature changes were measured at a distance of 1 mm from the implant drill. In the study, working models were created under different conditions for implant socket preparation. In group 1, the first time drills were used at 150 rpm without irrigation, in group 2, the first time drills were used at 1200 rpm with 40 ml/min irrigation, in group 3, the 30th time drills were used at 150 rpm without irrigation, and in group 4, the 30th time drills were used at 1200 rpm with 40 ml/min irrigation. All osteotomy procedures were performed with 3.5 ± 0.3 mm diameter burs for a period of 8 s and the temperature values obtained at equal time intervals were recorded. Repeated Measures and Kruskall Wallis-H tests were used for statistical analysis of the data. No significant difference was observed between the implant drills and the temperature changes in the bone during drilling (p < 0.05). None of the groups reached critical temperature values (47° C+) throughout the study. At the 5th and 10th mm depths, the temperature changes in the sensors used were close to each other. It was also calculated that although the average temperatures were close to each other in the non-irrigated and irrigated systems, the difference values obtained by subtracting the initial temperature were significantly higher in the non-irrigated systems. The results showed that implant drills did not cause significant temperature increases in bone blocks depending on the difference in manufacturers (Ankylos, Astra Tech, Nobel Biocare, Bredent, Straumann) and the number of uses. It was also concluded that irrigated and non-irrigated systems are safe as long as they are used under the recommended conditions.</p>\",\"PeriodicalId\":14076,\"journal\":{\"name\":\"International Journal of Implant Dentistry\",\"volume\":\"11 1\",\"pages\":\"11\"},\"PeriodicalIF\":3.1000,\"publicationDate\":\"2025-02-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11813834/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Implant Dentistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1186/s40729-025-00588-9\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Implant Dentistry","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40729-025-00588-9","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

本研究的目的是研究不同公司生产的种植体钻头在体外钻孔过程中骨内温度的变化。研究中使用了从新鲜牛肋骨中获得的骨块。骨块采用Ankylos、Astra Tech、Nobel Biocare、brendent和Straumann种植体品牌生产的钻头,在30±2°C的环境温度下,在2 kg的恒压下钻孔。将两个k型热电偶传感器分别放置在5 mm和10 mm深度的骨块上,在距离种植体钻头1mm处测量温度变化。在本研究中,建立了不同条件下的工作模型,用于种植槽的制备。第1组第一次钻孔,转速为150 rpm,不灌水;第2组第一次钻孔,转速为1200 rpm,灌水40 ml/min;第3组第30次钻孔,转速为150 rpm,不灌水;第4组第30次钻孔,转速为1200 rpm,灌水40 ml/min。所有截骨手术均采用直径3.5±0.3 mm的毛刺进行,持续8 s,并记录在相同时间间隔内获得的温度值。采用重复测量法和Kruskall Wallis-H检验对数据进行统计分析。种植体钻孔和钻孔过程中骨内温度变化之间无显著差异(p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
In-vitro investigation of bone temperature changes in osteotomies performed with different brands of implant burs.

The aim of this study was to investigate in-vitro the temperature changes occurring in the bone during drilling with implant drills manufactured by different companies. Bone blocks obtained from fresh bovine ribs were used in the study. Bone blocks were drilled with drills manufactured by Ankylos, Astra Tech, Nobel Biocare, Bredent and Straumann implant brands at an ambient temperature of 30 ± 2° C under a constant pressure of 2 kg. Two K-type thermocouple sensors were placed on the bone blocks at 5th and 10th mm depths and the temperature changes were measured at a distance of 1 mm from the implant drill. In the study, working models were created under different conditions for implant socket preparation. In group 1, the first time drills were used at 150 rpm without irrigation, in group 2, the first time drills were used at 1200 rpm with 40 ml/min irrigation, in group 3, the 30th time drills were used at 150 rpm without irrigation, and in group 4, the 30th time drills were used at 1200 rpm with 40 ml/min irrigation. All osteotomy procedures were performed with 3.5 ± 0.3 mm diameter burs for a period of 8 s and the temperature values obtained at equal time intervals were recorded. Repeated Measures and Kruskall Wallis-H tests were used for statistical analysis of the data. No significant difference was observed between the implant drills and the temperature changes in the bone during drilling (p < 0.05). None of the groups reached critical temperature values (47° C+) throughout the study. At the 5th and 10th mm depths, the temperature changes in the sensors used were close to each other. It was also calculated that although the average temperatures were close to each other in the non-irrigated and irrigated systems, the difference values obtained by subtracting the initial temperature were significantly higher in the non-irrigated systems. The results showed that implant drills did not cause significant temperature increases in bone blocks depending on the difference in manufacturers (Ankylos, Astra Tech, Nobel Biocare, Bredent, Straumann) and the number of uses. It was also concluded that irrigated and non-irrigated systems are safe as long as they are used under the recommended conditions.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
International Journal of Implant Dentistry
International Journal of Implant Dentistry DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
1.70
自引率
7.40%
发文量
53
审稿时长
13 weeks
期刊介绍: The International Journal of Implant Dentistry is a peer-reviewed open access journal published under the SpringerOpen brand. The journal is dedicated to promoting the exchange and discussion of all research areas relevant to implant dentistry in the form of systematic literature or invited reviews, prospective and retrospective clinical studies, clinical case reports, basic laboratory and animal research, and articles on material research and engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信