{"title":"[使用防护设备对结构性心脏病患者头侧医护人员降低辐射剂量的评估]。","authors":"Hirotoshi Murakami, Naoki Hosoda, Natsuki Oyama, Ryoya Hatsuoka, Itaru Sudo","doi":"10.6009/jjrt.25-1505","DOIUrl":null,"url":null,"abstract":"<p><p>In interventions for structural heart disease (SHD), an echocardiologist and an anesthesiologist are also present close to the side of the bed at the patient's head. The purpose of this study was to investigate the effectiveness of X-ray protective devices in reducing exposure to scatter radiation for medical staff at the side of the head end of the angiography system bed during interventions for structural heart disease. The ambient radiation dosage was measured, with and without X-ray protective equipment, to determine the rate of reduction of scatter radiation. The dosage tended to be higher at the working angle compared to the PA direction. When protective equipment was not used, the dosage tended to be higher at the side of the head end of the bed than at the operator's position. The dose was reduced by more than 90% compared to that without protective equipment by installing a protective device at the side of the head end of the bed. It is important to understand the spatial dose distribution at the working angle, and the installation of protective equipment at the side of the head end of the bed during SHD interventions can reduce the exposure of attending staff.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":"81 3","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Assessment of Radiation Dose Reduction for Medical Staff at the Patient's Head Side in Structural Heart Disease Intervention Using Protective Equipment].\",\"authors\":\"Hirotoshi Murakami, Naoki Hosoda, Natsuki Oyama, Ryoya Hatsuoka, Itaru Sudo\",\"doi\":\"10.6009/jjrt.25-1505\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In interventions for structural heart disease (SHD), an echocardiologist and an anesthesiologist are also present close to the side of the bed at the patient's head. The purpose of this study was to investigate the effectiveness of X-ray protective devices in reducing exposure to scatter radiation for medical staff at the side of the head end of the angiography system bed during interventions for structural heart disease. The ambient radiation dosage was measured, with and without X-ray protective equipment, to determine the rate of reduction of scatter radiation. The dosage tended to be higher at the working angle compared to the PA direction. When protective equipment was not used, the dosage tended to be higher at the side of the head end of the bed than at the operator's position. The dose was reduced by more than 90% compared to that without protective equipment by installing a protective device at the side of the head end of the bed. It is important to understand the spatial dose distribution at the working angle, and the installation of protective equipment at the side of the head end of the bed during SHD interventions can reduce the exposure of attending staff.</p>\",\"PeriodicalId\":74309,\"journal\":{\"name\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"volume\":\"81 3\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/jjrt.25-1505\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.25-1505","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在结构性心脏病(SHD)的干预中,超声心动图专家和麻醉师也会出现在靠近患者头部的床边。本研究的目的是探讨x射线防护装置在减少血管造影系统床头端侧医务人员在结构性心脏病干预期间散射辐射暴露的有效性。测量了有x射线防护设备和没有x射线防护设备的环境辐射剂量,以确定散射辐射的减少率。与PA方向相比,工作角方向的投加量有较高的趋势。在不使用防护设备的情况下,床头端一侧的剂量往往高于操作人员位置。通过在床头一侧安装防护装置,与没有防护设备的情况下相比,剂量减少了90%以上。了解工作角度下的空间剂量分布是很重要的,在SHD干预期间,在床头端侧安装防护设备可以减少护理人员的暴露。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Assessment of Radiation Dose Reduction for Medical Staff at the Patient's Head Side in Structural Heart Disease Intervention Using Protective Equipment].

In interventions for structural heart disease (SHD), an echocardiologist and an anesthesiologist are also present close to the side of the bed at the patient's head. The purpose of this study was to investigate the effectiveness of X-ray protective devices in reducing exposure to scatter radiation for medical staff at the side of the head end of the angiography system bed during interventions for structural heart disease. The ambient radiation dosage was measured, with and without X-ray protective equipment, to determine the rate of reduction of scatter radiation. The dosage tended to be higher at the working angle compared to the PA direction. When protective equipment was not used, the dosage tended to be higher at the side of the head end of the bed than at the operator's position. The dose was reduced by more than 90% compared to that without protective equipment by installing a protective device at the side of the head end of the bed. It is important to understand the spatial dose distribution at the working angle, and the installation of protective equipment at the side of the head end of the bed during SHD interventions can reduce the exposure of attending staff.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信