静脉注射CT造影剂对确定腰椎骨小梁骨密度的内部校准技术的影响。

IF 3.2 3区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Stefan Bartenschlager , Alexander Cavallaro , Tobias Pogarell , Michael Uder , Oliver Chaudry , Georg Schett , Klaus Engelke
{"title":"静脉注射CT造影剂对确定腰椎骨小梁骨密度的内部校准技术的影响。","authors":"Stefan Bartenschlager ,&nbsp;Alexander Cavallaro ,&nbsp;Tobias Pogarell ,&nbsp;Michael Uder ,&nbsp;Oliver Chaudry ,&nbsp;Georg Schett ,&nbsp;Klaus Engelke","doi":"10.1016/j.ejrad.2025.111923","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><div>Contrast agents are frequently administered in computed tomography (CT) scans used for opportunistic screening of osteoporosis. The objective of this study is to compare the impact of contrast-related bone mineral density (BMD) increase between phantom-based and internal CT calibration techniques.</div></div><div><h3>Materials and Methods</h3><div>Phantom-based and internal CT calibration techniques were used to determine trabecular BMD in 93 existing clinical CT scans of the lumbar spine of 34 subjects, scanned before and after administration of contrast agents. BMD differences between native (Nat) and contrast-enhanced scans in arterial (Art) and portal venous (PV) phases were compared among calibration methods. Three pairs of internal reference materials were investigated: blood/air, subcutaneous adipose tissue (SAT)/air and muscle tissue/air.</div></div><div><h3>Results</h3><div>Contrast agents increased CT values of blood in the aorta and the inferior vena cava in the Art phase by 210 ± 99HU and 47 ± 35HU, respectively and in the PV phase by 110 ± 49HU and 73 ± 28HU, respectively. Effects on CT values of muscle and SAT were &lt; 5HU in the Art phase and &lt; 10HU in the PV phase.</div><div>BMD obtained with phantom calibration increased significantly (p &lt; 0.01) in Art and PV phases by 8.6 ± 14.7 mg/cm<sup>3</sup> and 14.4 ± 16.3 mg/cm<sup>3</sup>, respectively. If SAT/air or muscle/air were used as internal reference materials, increases in internally calibrated BMD (&lt;12 mg/cm<sup>3</sup> in AT phase and &lt; 17 mg/cm<sup>3</sup> in PV phase) did not significantly differ (p &gt; 0.1 for Art and PV phases) from increases of phantom-calibrated BMD.</div></div><div><h3>Conclusions</h3><div>Although CT values of internal reference materials increased following the administration of contrast, internal calibration was unable to mitigate the contrast-related BMD increases observed in phantom-based calibration.</div></div>","PeriodicalId":12063,"journal":{"name":"European Journal of Radiology","volume":"183 ","pages":"Article 111923"},"PeriodicalIF":3.2000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Impact of intravenous CT contrast agents on internal calibration techniques to determine trabecular BMD of the lumbar spine\",\"authors\":\"Stefan Bartenschlager ,&nbsp;Alexander Cavallaro ,&nbsp;Tobias Pogarell ,&nbsp;Michael Uder ,&nbsp;Oliver Chaudry ,&nbsp;Georg Schett ,&nbsp;Klaus Engelke\",\"doi\":\"10.1016/j.ejrad.2025.111923\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><div>Contrast agents are frequently administered in computed tomography (CT) scans used for opportunistic screening of osteoporosis. The objective of this study is to compare the impact of contrast-related bone mineral density (BMD) increase between phantom-based and internal CT calibration techniques.</div></div><div><h3>Materials and Methods</h3><div>Phantom-based and internal CT calibration techniques were used to determine trabecular BMD in 93 existing clinical CT scans of the lumbar spine of 34 subjects, scanned before and after administration of contrast agents. BMD differences between native (Nat) and contrast-enhanced scans in arterial (Art) and portal venous (PV) phases were compared among calibration methods. Three pairs of internal reference materials were investigated: blood/air, subcutaneous adipose tissue (SAT)/air and muscle tissue/air.</div></div><div><h3>Results</h3><div>Contrast agents increased CT values of blood in the aorta and the inferior vena cava in the Art phase by 210 ± 99HU and 47 ± 35HU, respectively and in the PV phase by 110 ± 49HU and 73 ± 28HU, respectively. Effects on CT values of muscle and SAT were &lt; 5HU in the Art phase and &lt; 10HU in the PV phase.</div><div>BMD obtained with phantom calibration increased significantly (p &lt; 0.01) in Art and PV phases by 8.6 ± 14.7 mg/cm<sup>3</sup> and 14.4 ± 16.3 mg/cm<sup>3</sup>, respectively. If SAT/air or muscle/air were used as internal reference materials, increases in internally calibrated BMD (&lt;12 mg/cm<sup>3</sup> in AT phase and &lt; 17 mg/cm<sup>3</sup> in PV phase) did not significantly differ (p &gt; 0.1 for Art and PV phases) from increases of phantom-calibrated BMD.</div></div><div><h3>Conclusions</h3><div>Although CT values of internal reference materials increased following the administration of contrast, internal calibration was unable to mitigate the contrast-related BMD increases observed in phantom-based calibration.</div></div>\",\"PeriodicalId\":12063,\"journal\":{\"name\":\"European Journal of Radiology\",\"volume\":\"183 \",\"pages\":\"Article 111923\"},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2025-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Journal of Radiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0720048X25000099\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Radiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0720048X25000099","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:造影剂经常用于计算机断层扫描(CT),用于骨质疏松症的机会性筛查。本研究的目的是比较基于幻影和内部CT校准技术对对比相关骨矿物质密度(BMD)增加的影响。材料和方法:采用基于幻影和内部CT校准技术,对34名受试者的93份腰椎临床CT扫描进行对比剂前后扫描,以确定骨小梁骨密度。比较两种校准方法在动脉(Art)和门静脉(PV)期的天然(Nat)和增强扫描的骨密度差异。研究了血液/空气、皮下脂肪组织(SAT)/空气和肌肉组织/空气三对内参物。结果:造影剂使Art期主动脉和下腔静脉血流CT值分别增加210±99HU和47±35HU,使PV期血流CT值分别增加110±49HU和73±28HU。对肌肉CT值和SAT值的影响分别为3和14.4±16.3 mg/cm3。如果使用SAT/air或肌肉/air作为内参比物质,内部校准BMD的增加(AT期3个,PV期3个)与幻影校准BMD的增加没有显著差异(Art和PV期p >.1)。结论:虽然内参比物的CT值在使用造影剂后增加,但内校准无法减轻基于幻象的校准中观察到的与造影剂相关的BMD增加。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Impact of intravenous CT contrast agents on internal calibration techniques to determine trabecular BMD of the lumbar spine

Objectives

Contrast agents are frequently administered in computed tomography (CT) scans used for opportunistic screening of osteoporosis. The objective of this study is to compare the impact of contrast-related bone mineral density (BMD) increase between phantom-based and internal CT calibration techniques.

Materials and Methods

Phantom-based and internal CT calibration techniques were used to determine trabecular BMD in 93 existing clinical CT scans of the lumbar spine of 34 subjects, scanned before and after administration of contrast agents. BMD differences between native (Nat) and contrast-enhanced scans in arterial (Art) and portal venous (PV) phases were compared among calibration methods. Three pairs of internal reference materials were investigated: blood/air, subcutaneous adipose tissue (SAT)/air and muscle tissue/air.

Results

Contrast agents increased CT values of blood in the aorta and the inferior vena cava in the Art phase by 210 ± 99HU and 47 ± 35HU, respectively and in the PV phase by 110 ± 49HU and 73 ± 28HU, respectively. Effects on CT values of muscle and SAT were < 5HU in the Art phase and < 10HU in the PV phase.
BMD obtained with phantom calibration increased significantly (p < 0.01) in Art and PV phases by 8.6 ± 14.7 mg/cm3 and 14.4 ± 16.3 mg/cm3, respectively. If SAT/air or muscle/air were used as internal reference materials, increases in internally calibrated BMD (<12 mg/cm3 in AT phase and < 17 mg/cm3 in PV phase) did not significantly differ (p > 0.1 for Art and PV phases) from increases of phantom-calibrated BMD.

Conclusions

Although CT values of internal reference materials increased following the administration of contrast, internal calibration was unable to mitigate the contrast-related BMD increases observed in phantom-based calibration.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.70
自引率
3.00%
发文量
398
审稿时长
42 days
期刊介绍: European Journal of Radiology is an international journal which aims to communicate to its readers, state-of-the-art information on imaging developments in the form of high quality original research articles and timely reviews on current developments in the field. Its audience includes clinicians at all levels of training including radiology trainees, newly qualified imaging specialists and the experienced radiologist. Its aim is to inform efficient, appropriate and evidence-based imaging practice to the benefit of patients worldwide.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信