使用ID-LC-MS/MS对心肌肌钙蛋白I定量的肽与蛋白校准进行比较评价。

IF 3.8 2区 医学 Q1 MEDICAL LABORATORY TECHNOLOGY
Meltem Asicioglu, Claudia Swart, Evren Saban, Emrah Yurek, Nevin Gul Karaguler, Merve Oztug
{"title":"使用ID-LC-MS/MS对心肌肌钙蛋白I定量的肽与蛋白校准进行比较评价。","authors":"Meltem Asicioglu, Claudia Swart, Evren Saban, Emrah Yurek, Nevin Gul Karaguler, Merve Oztug","doi":"10.1515/cclm-2024-0999","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.</p><p><strong>Methods: </strong>To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.</p><p><strong>Results: </strong>The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).</p><p><strong>Conclusions: </strong>This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.</p>","PeriodicalId":10390,"journal":{"name":"Clinical chemistry and laboratory medicine","volume":" ","pages":""},"PeriodicalIF":3.8000,"publicationDate":"2025-01-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparative evaluation of peptide vs. protein-based calibration for quantification of cardiac troponin I using ID-LC-MS/MS.\",\"authors\":\"Meltem Asicioglu, Claudia Swart, Evren Saban, Emrah Yurek, Nevin Gul Karaguler, Merve Oztug\",\"doi\":\"10.1515/cclm-2024-0999\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.</p><p><strong>Methods: </strong>To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.</p><p><strong>Results: </strong>The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).</p><p><strong>Conclusions: </strong>This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.</p>\",\"PeriodicalId\":10390,\"journal\":{\"name\":\"Clinical chemistry and laboratory medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.8000,\"publicationDate\":\"2025-01-03\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical chemistry and laboratory medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1515/cclm-2024-0999\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical chemistry and laboratory medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1515/cclm-2024-0999","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:建立并验证了一种基于同位素稀释液相色谱-串联质谱(ID-LC-MS/MS)的分析方案,其中包括基于肽的校准策略,用于测定临床样品中心肌肌钙蛋白I (cTnI)的水平。此外,将开发的方法与基于蛋白质的校准策略进行了比较,使用cTnI作为低丰度蛋白质的模型。目的是评估复杂基质中蛋白质定量的新方法,支持计量界实施新方法和开发适合用途的SI可追溯肽或蛋白质初级校准器。方法:采用多肽杂质校正氨基酸分析(PICAA)测定一级标准品中特征肽的绝对含量,建立可追溯性。采用免疫亲和富集的方法从人血清中捕获cTnI,并比较微珠和纳米珠的富集效率。平行反应监测用于监测cTnI特异性的两个特征肽。优化了各种消化参数,以达到完全消化。结果:该分析方法具有较好的选择性和特异性,cTnI的定量范围为0.9 ~ 22.0 μg/L。各浓度下的中间精密度RSD < 28.9 %,重复性RSD < 5.8 %,加样回收率为87 % ~ 121 %。两种校准策略的LOQ值相近,但基于多肽的校准在回收率上存在显著的定量偏差。数据可通过ProteomeXchange (PXD055104)获得。结论:基于多肽定标的同位素稀释液相色谱-串联质谱法(ID-LC-MS/MS)可成功定量人血清中的cTnI。将其与基于蛋白质的校准进行比较,突出了基于肽的策略的优势和潜在局限性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparative evaluation of peptide vs. protein-based calibration for quantification of cardiac troponin I using ID-LC-MS/MS.

Objectives: An analytical protocol based on isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS), which includes a peptide-based calibration strategy, was developed and validated for the determination of cardiac troponin I (cTnI) levels in clinical samples. Additionally, the developed method was compared with a protein-based calibration strategy, using cTnI serving as a model for low-abundant proteins. The aim is to evaluate new approaches for protein quantification in complex matrices, supporting the metrology community in implementing new methods and developing fit-for-purpose SI- traceable peptide or protein primary calibrators.

Methods: To establish traceability to SI units, peptide impurity correction amino acid analysis (PICAA) was conducted to determine the absolute content of signature peptides in the primary standards. Immunoaffinity enrichment was used to capture cTnI from human serum, with a comparison between microbeads and nanobeads to improve enrichment efficiency. Parallel reaction monitoring was used to monitor two signature peptides specific to cTnI. Various digestion parameters were optimized to achieve complete digestion.

Results: The analytical method demonstrated selectivity and specificity, allowing the quantification of cTnI within 0.9-22.0 μg/L. The intermediate precision RSD was below 28.9 %, and the repeatability RSD was below 5.8 % at all concentration levels, with recovery rates ranging from 87 % to 121 %. The comparison of calibration strategies showed similar LOQ values, but the peptide-based calibration exhibited significant quantitative bias in recovery rates. The data are available via ProteomeXchange (PXD055104).

Conclusions: This isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method, based on peptide calibration, successfully quantified cTnI in human serum. Comparing this with protein-based calibration highlighted both the strengths and potential limitations of peptide-based strategies.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical chemistry and laboratory medicine
Clinical chemistry and laboratory medicine 医学-医学实验技术
CiteScore
11.30
自引率
16.20%
发文量
306
审稿时长
3 months
期刊介绍: Clinical Chemistry and Laboratory Medicine (CCLM) publishes articles on novel teaching and training methods applicable to laboratory medicine. CCLM welcomes contributions on the progress in fundamental and applied research and cutting-edge clinical laboratory medicine. It is one of the leading journals in the field, with an impact factor over 3. CCLM is issued monthly, and it is published in print and electronically. CCLM is the official journal of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) and publishes regularly EFLM recommendations and news. CCLM is the official journal of the National Societies from Austria (ÖGLMKC); Belgium (RBSLM); Germany (DGKL); Hungary (MLDT); Ireland (ACBI); Italy (SIBioC); Portugal (SPML); and Slovenia (SZKK); and it is affiliated to AACB (Australia) and SFBC (France). Topics: - clinical biochemistry - clinical genomics and molecular biology - clinical haematology and coagulation - clinical immunology and autoimmunity - clinical microbiology - drug monitoring and analysis - evaluation of diagnostic biomarkers - disease-oriented topics (cardiovascular disease, cancer diagnostics, diabetes) - new reagents, instrumentation and technologies - new methodologies - reference materials and methods - reference values and decision limits - quality and safety in laboratory medicine - translational laboratory medicine - clinical metrology Follow @cclm_degruyter on Twitter!
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信