Janne Moilanen MD , Marika Pada MD , Pasi Ohtonen MSc , Timo Kaakinen MD, PhD , Matti Pokela MD, PhD , Eeva-Riitta Savolainen MD, PhD , Tiina Erkinaro MD, PhD
{"title":"血色素ACT- lr和ACT+试验皿监测中低肝素浓度的性能:一项体外研究。","authors":"Janne Moilanen MD ,&nbsp;Marika Pada MD ,&nbsp;Pasi Ohtonen MSc ,&nbsp;Timo Kaakinen MD, PhD ,&nbsp;Matti Pokela MD, PhD ,&nbsp;Eeva-Riitta Savolainen MD, PhD ,&nbsp;Tiina Erkinaro MD, PhD","doi":"10.1053/j.jvca.2024.12.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><div>According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.</div></div><div><h3>Design</h3><div>An in vitro study.</div></div><div><h3>Setting</h3><div>A tertiary care university hospital.</div></div><div><h3>Participants</h3><div>Twenty healthy adult volunteers.</div></div><div><h3>Interventions</h3><div>Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.</div></div><div><h3>Measurements and Main Results</h3><div>All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements ​​was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.</div></div><div><h3>Conclusions</h3><div>The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.</div></div>","PeriodicalId":15176,"journal":{"name":"Journal of cardiothoracic and vascular anesthesia","volume":"39 2","pages":"Pages 447-452"},"PeriodicalIF":2.3000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study\",\"authors\":\"Janne Moilanen MD ,&nbsp;Marika Pada MD ,&nbsp;Pasi Ohtonen MSc ,&nbsp;Timo Kaakinen MD, PhD ,&nbsp;Matti Pokela MD, PhD ,&nbsp;Eeva-Riitta Savolainen MD, PhD ,&nbsp;Tiina Erkinaro MD, PhD\",\"doi\":\"10.1053/j.jvca.2024.12.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><div>According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.</div></div><div><h3>Design</h3><div>An in vitro study.</div></div><div><h3>Setting</h3><div>A tertiary care university hospital.</div></div><div><h3>Participants</h3><div>Twenty healthy adult volunteers.</div></div><div><h3>Interventions</h3><div>Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.</div></div><div><h3>Measurements and Main Results</h3><div>All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements ​​was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.</div></div><div><h3>Conclusions</h3><div>The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.</div></div>\",\"PeriodicalId\":15176,\"journal\":{\"name\":\"Journal of cardiothoracic and vascular anesthesia\",\"volume\":\"39 2\",\"pages\":\"Pages 447-452\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of cardiothoracic and vascular anesthesia\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S105307702400939X\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ANESTHESIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cardiothoracic and vascular anesthesia","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S105307702400939X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:据生产商介绍,haemchron ACT- lr比色管的肝素浓度范围为0 - 2.5 IU/mL,而ACT+试剂盒的最佳浓度范围为1 - 6 IU/mL。我们假设在低至中等肝素浓度下,ACT- lr比ACT+更可靠。设计:体外研究。环境:三级保健大学医院。参与者:20名健康成人志愿者。干预措施:供体血液样品加入未分离肝素,浓度分别为0、0.5、1.0、1.5、2.0和2.5 IU/mL。从每个浓度开始,每个测试盒使用3种不同的haemchron Signature Elite设备进行3次测量,总共进行720次ACT测试。采用Bland-Altman方法和类内相关系数对数据进行分析。测量和主要结果:所有360 ACT+测试都产生了结果。360例中,38例(11%)ACT-LR测试高于最高测量范围;其中,36例来自肝素浓度≥2.0 IU/mL的120例检测。所有ACT- lr和ACT+测量之间的偏差为52 (95% CI, 48-56) celite秒,一致性的限制范围很广。从类内相关系数来看,ACT+的信度优于或等于ACT- lr。结论:在0.5 IU/mL浓度下,ACT+与ACT- lr的性能相当,高于0.5 IU/mL时,ACT+优于ACT- lr,对进行特异性低量程检测的合理性提出质疑。然而,如果从一种测试切换到另一种测试,ACT- lr和ACT+测量之间需要考虑明显的偏差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Performance of Hemochron ACT-LR and ACT+ Test Cuvettes in Monitoring Low to Moderate Heparin Concentrations: An In Vitro Study

Objectives

According to the manufacturer, the Hemochron ACT-LR cuvette is designated for heparin concentrations of 0 to 2.5 IU/mL, while the optimal concentration range for the ACT+ cartridge is 1 to 6 IU/mL. We hypothesized that at low to moderate heparin concentrations, the ACT-LR is more reliable than the ACT+.

Design

An in vitro study.

Setting

A tertiary care university hospital.

Participants

Twenty healthy adult volunteers.

Interventions

Donor blood samples were spiked with unfractionated heparin to concentrations of 0, 0.5, 1.0, 1.5, 2.0, and 2.5 IU/mL. From each concentration, 3 measurements were made with each test cassette using 3 different Hemochron Signature Elite devices, for a total of 720 ACT tests. The Bland-Altman method and intraclass correlation coefficients were used to analyze the data.

Measurements and Main Results

All 360 ACT+ tests yielded a result. Of 360, 38 (11%) ACT-LR tests ran above the upper measurement range; of these, 36 were from the 120 measurements at heparin concentrations ≥2.0 IU/mL. The bias between all ACT-LR and ACT+ measurements ​​was 52 (95% CI, 48-56) celite seconds, and the limits of agreement were wide. Based on the intraclass correlation coefficients, the reliability of ACT+ was better than or equal to that of ACT-LR.

Conclusions

The performance of ACT+ is equal to that of ACT-LR up to a concentration of 0.5 IU/mL, above which ACT+ is superior to ACT-LR, questioning the rationale for a specific low-range test. However, there is a significant bias between ACT-LR and ACT+ measurements that needs to be considered if switching from one test to the other.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.80
自引率
17.90%
发文量
606
审稿时长
37 days
期刊介绍: The Journal of Cardiothoracic and Vascular Anesthesia is primarily aimed at anesthesiologists who deal with patients undergoing cardiac, thoracic or vascular surgical procedures. JCVA features a multidisciplinary approach, with contributions from cardiac, vascular and thoracic surgeons, cardiologists, and other related specialists. Emphasis is placed on rapid publication of clinically relevant material.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信