Valeria Faccenda, Denis Panizza, Rita Marina Niespolo, Riccardo Ray Colciago, Giulia Rossano, Lorenzo De Sanctis, Davide Gandola, Davide Ippolito, Stefano Arcangeli, Elena De Ponti
{"title":"用于腹部 SBRT 靶区划分的同步对比度增强 4DCT 模拟。","authors":"Valeria Faccenda, Denis Panizza, Rita Marina Niespolo, Riccardo Ray Colciago, Giulia Rossano, Lorenzo De Sanctis, Davide Gandola, Davide Ippolito, Stefano Arcangeli, Elena De Ponti","doi":"10.3390/cancers16234066","DOIUrl":null,"url":null,"abstract":"<p><p><b>Background/Objectives:</b> To present the technical aspects of contrast-enhanced 4DCT (ce4DCT) simulation for abdominal SBRT. <b>Methods:</b> Twenty-two patients underwent two sequential 4DCT scans: one baseline and one contrast-enhanced with personalized delay time (<i>t<sub>delay</sub></i>) calculated to capture the tumor in the desired contrast phase, based on diagnostic triple-phase CT. The internal target volume (ITV) was delineated on ten contrast phases, and a panel of three experts qualitatively evaluated tumor visibility. Aortic HU values were measured to refine the <i>t<sub>delay</sub></i> for subsequent patients. The commonly used approach of combining triple-phase CT with unenhanced 4DCT was simulated, and differences in target delineation were evaluated by volume, centroid shift, Dice and Jaccard indices, and mean distance agreement (MDA). The margins required to account for motion were calculated. <b>Results:</b> The ce4DCT acquisitions substantially improved tumor visibility over the entire breathing cycle in 20 patients, according to the experts' unanimous evaluation. The median contrast peak time was 54.5 s, and the washout plateau was observed at 70.3 s, with mean peak and plateau HU values of 292 ± 59 and 169 ± 25. The volumes from the commonly used procedure (ITV2) were significantly smaller than the ce4DCT volumes (ITV1) (<i>p</i> = 0.045). The median centroid shift was 4.7 mm. The ITV1-ITV2 overlap was 69% (Dice index), 53% (Jaccard index), and 2.89 mm (MDA), with the liver volumes showing significantly lower indices compared to the pancreatic volumes (<i>p</i> ≤ 0.011). The margins required to better encompass ITV1 were highly variable, with mean values ≥ 4 mm in all directions except for the left-right axis. <b>Conclusions:</b> The ce4DCT simulation was feasible, resulting in optimal tumor enhancement with minimal resource investment, while significantly mitigating uncertainties in SBRT planning by addressing poor visibility and respiratory motion. Triple-phase 3DCT with unenhanced 4DCT led to high variability in target delineation, making the isotropic margins ineffective.</p>","PeriodicalId":9681,"journal":{"name":"Cancers","volume":"16 23","pages":""},"PeriodicalIF":4.5000,"publicationDate":"2024-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Synchronized Contrast-Enhanced 4DCT Simulation for Target Volume Delineation in Abdominal SBRT.\",\"authors\":\"Valeria Faccenda, Denis Panizza, Rita Marina Niespolo, Riccardo Ray Colciago, Giulia Rossano, Lorenzo De Sanctis, Davide Gandola, Davide Ippolito, Stefano Arcangeli, Elena De Ponti\",\"doi\":\"10.3390/cancers16234066\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b>Background/Objectives:</b> To present the technical aspects of contrast-enhanced 4DCT (ce4DCT) simulation for abdominal SBRT. <b>Methods:</b> Twenty-two patients underwent two sequential 4DCT scans: one baseline and one contrast-enhanced with personalized delay time (<i>t<sub>delay</sub></i>) calculated to capture the tumor in the desired contrast phase, based on diagnostic triple-phase CT. The internal target volume (ITV) was delineated on ten contrast phases, and a panel of three experts qualitatively evaluated tumor visibility. Aortic HU values were measured to refine the <i>t<sub>delay</sub></i> for subsequent patients. The commonly used approach of combining triple-phase CT with unenhanced 4DCT was simulated, and differences in target delineation were evaluated by volume, centroid shift, Dice and Jaccard indices, and mean distance agreement (MDA). The margins required to account for motion were calculated. <b>Results:</b> The ce4DCT acquisitions substantially improved tumor visibility over the entire breathing cycle in 20 patients, according to the experts' unanimous evaluation. The median contrast peak time was 54.5 s, and the washout plateau was observed at 70.3 s, with mean peak and plateau HU values of 292 ± 59 and 169 ± 25. The volumes from the commonly used procedure (ITV2) were significantly smaller than the ce4DCT volumes (ITV1) (<i>p</i> = 0.045). The median centroid shift was 4.7 mm. The ITV1-ITV2 overlap was 69% (Dice index), 53% (Jaccard index), and 2.89 mm (MDA), with the liver volumes showing significantly lower indices compared to the pancreatic volumes (<i>p</i> ≤ 0.011). The margins required to better encompass ITV1 were highly variable, with mean values ≥ 4 mm in all directions except for the left-right axis. <b>Conclusions:</b> The ce4DCT simulation was feasible, resulting in optimal tumor enhancement with minimal resource investment, while significantly mitigating uncertainties in SBRT planning by addressing poor visibility and respiratory motion. Triple-phase 3DCT with unenhanced 4DCT led to high variability in target delineation, making the isotropic margins ineffective.</p>\",\"PeriodicalId\":9681,\"journal\":{\"name\":\"Cancers\",\"volume\":\"16 23\",\"pages\":\"\"},\"PeriodicalIF\":4.5000,\"publicationDate\":\"2024-12-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cancers\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3390/cancers16234066\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cancers","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3390/cancers16234066","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ONCOLOGY","Score":null,"Total":0}
Synchronized Contrast-Enhanced 4DCT Simulation for Target Volume Delineation in Abdominal SBRT.
Background/Objectives: To present the technical aspects of contrast-enhanced 4DCT (ce4DCT) simulation for abdominal SBRT. Methods: Twenty-two patients underwent two sequential 4DCT scans: one baseline and one contrast-enhanced with personalized delay time (tdelay) calculated to capture the tumor in the desired contrast phase, based on diagnostic triple-phase CT. The internal target volume (ITV) was delineated on ten contrast phases, and a panel of three experts qualitatively evaluated tumor visibility. Aortic HU values were measured to refine the tdelay for subsequent patients. The commonly used approach of combining triple-phase CT with unenhanced 4DCT was simulated, and differences in target delineation were evaluated by volume, centroid shift, Dice and Jaccard indices, and mean distance agreement (MDA). The margins required to account for motion were calculated. Results: The ce4DCT acquisitions substantially improved tumor visibility over the entire breathing cycle in 20 patients, according to the experts' unanimous evaluation. The median contrast peak time was 54.5 s, and the washout plateau was observed at 70.3 s, with mean peak and plateau HU values of 292 ± 59 and 169 ± 25. The volumes from the commonly used procedure (ITV2) were significantly smaller than the ce4DCT volumes (ITV1) (p = 0.045). The median centroid shift was 4.7 mm. The ITV1-ITV2 overlap was 69% (Dice index), 53% (Jaccard index), and 2.89 mm (MDA), with the liver volumes showing significantly lower indices compared to the pancreatic volumes (p ≤ 0.011). The margins required to better encompass ITV1 were highly variable, with mean values ≥ 4 mm in all directions except for the left-right axis. Conclusions: The ce4DCT simulation was feasible, resulting in optimal tumor enhancement with minimal resource investment, while significantly mitigating uncertainties in SBRT planning by addressing poor visibility and respiratory motion. Triple-phase 3DCT with unenhanced 4DCT led to high variability in target delineation, making the isotropic margins ineffective.
期刊介绍:
Cancers (ISSN 2072-6694) is an international, peer-reviewed open access journal on oncology. It publishes reviews, regular research papers and short communications. Our aim is to encourage scientists to publish their experimental and theoretical results in as much detail as possible. There is no restriction on the length of the papers. The full experimental details must be provided so that the results can be reproduced.