Beckton Dickinson和Kima 3.2%柠檬酸钠凝血管中优球蛋白凝块溶解时间参考区间的测定。

IF 1.2 4区 医学 Q4 HEMATOLOGY
Blood Coagulation & Fibrinolysis Pub Date : 2025-01-01 Epub Date: 2024-11-18 DOI:10.1097/MBC.0000000000001334
Ivona Kuktić, Ante Pašalić, Ognjen Čančarević, Edvard Galić, Andrea Saračević, Vanja Radišić Biljak
{"title":"Beckton Dickinson和Kima 3.2%柠檬酸钠凝血管中优球蛋白凝块溶解时间参考区间的测定。","authors":"Ivona Kuktić, Ante Pašalić, Ognjen Čančarević, Edvard Galić, Andrea Saračević, Vanja Radišić Biljak","doi":"10.1097/MBC.0000000000001334","DOIUrl":null,"url":null,"abstract":"<p><p>Enhanced fibrinolysis or hyperfibrinolysis may lead to life-threatening blood loss, while reduced activity may contribute to thrombosis. Euglobulin clot lysis time (ECLT) is a manual method that measures plasma fibrinolytic activity and is considered the gold standard. However, the data on reference interval is scarce and outdated. We have employed one-sided reference interval (>3 h) since the implementation of ECLT in our laboratory; therefore, we aimed to establish reliable ECLT reference interval and to explore the possible preanalytical influence of different blood collection tubes on the established ECLT reference interval. Establishing a reference interval for fibrinolysis was performed according to CLSI EP28-A3c guidelines by employing a posteriori direct sampling technique. The predefined exclusion criteria included a history of malignant or hepatobiliary disease, a history of deep vein thrombosis/pulmonary embolism (DVT/PE), an acute inflammatory state at the time of the venipuncture. We collected vein blood samples in Vacutest plastic coagulation tubes (Kima, Italy) and Vacutainer glass coagulation tubes (Beckton Dickinson, USA) containing 0.109 mol/l buffered trisodium citrate as an anticoagulant at a blood-to-anticoagulant ratio 9 : 1. We calculated two-sided reference interval and presented as 2.5th and 97.5th percentiles. ECLT values did not differ between sexes or the types of tubes enrolled in the study ( P  = 0.8979). The established reference interval ranged from 130 to 297 min for the KIMA Vacutest tube and from 120 to 292 min for the BD Vacutainer tube. The established ECLT reference interval differed significantly from the currently used cut-off value in our laboratory, thus enabling the assessment of hyperfibrinolysis by employing double-sided reference interval.</p>","PeriodicalId":8992,"journal":{"name":"Blood Coagulation & Fibrinolysis","volume":" ","pages":"14-17"},"PeriodicalIF":1.2000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The determination of euglobulin clot lysis time reference intervals in Beckton Dickinson and Kima 3.2% sodium citrate coagulation tubes.\",\"authors\":\"Ivona Kuktić, Ante Pašalić, Ognjen Čančarević, Edvard Galić, Andrea Saračević, Vanja Radišić Biljak\",\"doi\":\"10.1097/MBC.0000000000001334\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Enhanced fibrinolysis or hyperfibrinolysis may lead to life-threatening blood loss, while reduced activity may contribute to thrombosis. Euglobulin clot lysis time (ECLT) is a manual method that measures plasma fibrinolytic activity and is considered the gold standard. However, the data on reference interval is scarce and outdated. We have employed one-sided reference interval (>3 h) since the implementation of ECLT in our laboratory; therefore, we aimed to establish reliable ECLT reference interval and to explore the possible preanalytical influence of different blood collection tubes on the established ECLT reference interval. Establishing a reference interval for fibrinolysis was performed according to CLSI EP28-A3c guidelines by employing a posteriori direct sampling technique. The predefined exclusion criteria included a history of malignant or hepatobiliary disease, a history of deep vein thrombosis/pulmonary embolism (DVT/PE), an acute inflammatory state at the time of the venipuncture. We collected vein blood samples in Vacutest plastic coagulation tubes (Kima, Italy) and Vacutainer glass coagulation tubes (Beckton Dickinson, USA) containing 0.109 mol/l buffered trisodium citrate as an anticoagulant at a blood-to-anticoagulant ratio 9 : 1. We calculated two-sided reference interval and presented as 2.5th and 97.5th percentiles. ECLT values did not differ between sexes or the types of tubes enrolled in the study ( P  = 0.8979). The established reference interval ranged from 130 to 297 min for the KIMA Vacutest tube and from 120 to 292 min for the BD Vacutainer tube. The established ECLT reference interval differed significantly from the currently used cut-off value in our laboratory, thus enabling the assessment of hyperfibrinolysis by employing double-sided reference interval.</p>\",\"PeriodicalId\":8992,\"journal\":{\"name\":\"Blood Coagulation & Fibrinolysis\",\"volume\":\" \",\"pages\":\"14-17\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Coagulation & Fibrinolysis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBC.0000000000001334\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/11/18 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Coagulation & Fibrinolysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBC.0000000000001334","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/18 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

纤溶增强或纤溶过高可导致危及生命的失血,而活动降低可导致血栓形成。Euglobulin凝块溶解时间(ECLT)是测量血浆纤溶活性的手工方法,被认为是金标准。然而,关于参考区间的数据很少且过时。自我们实验室实施ECLT以来,我们采用了单侧参考区间(> - 3小时);因此,我们旨在建立可靠的ECLT参考区间,并探讨不同采血管对建立的ECLT参考区间可能的分析前影响。根据CLSI EP28-A3c指南,采用后验直接抽样技术建立纤溶的参考区间。预先确定的排除标准包括恶性或肝胆疾病史,深静脉血栓/肺栓塞史(DVT/PE),静脉穿刺时急性炎症状态。我们在Vacutest塑料凝血管(Kima,意大利)和Vacutainer玻璃凝血管(Beckton Dickinson,美国)中采集静脉血液样本,其中含有0.109 mol/l缓冲柠檬酸三钠作为抗凝剂,血液与抗凝剂的比例为9:1。我们计算了双侧参考区间,并以2.5和97.5百分位表示。ECLT值在性别或纳入研究的试管类型之间没有差异(P = 0.8979)。KIMA真空管的参考间隔为130 - 297分钟,BD真空管的参考间隔为120 - 292分钟。建立的ECLT参考区间与我们实验室目前使用的临界值有明显差异,因此可以采用双面参考区间来评估高纤溶。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The determination of euglobulin clot lysis time reference intervals in Beckton Dickinson and Kima 3.2% sodium citrate coagulation tubes.

Enhanced fibrinolysis or hyperfibrinolysis may lead to life-threatening blood loss, while reduced activity may contribute to thrombosis. Euglobulin clot lysis time (ECLT) is a manual method that measures plasma fibrinolytic activity and is considered the gold standard. However, the data on reference interval is scarce and outdated. We have employed one-sided reference interval (>3 h) since the implementation of ECLT in our laboratory; therefore, we aimed to establish reliable ECLT reference interval and to explore the possible preanalytical influence of different blood collection tubes on the established ECLT reference interval. Establishing a reference interval for fibrinolysis was performed according to CLSI EP28-A3c guidelines by employing a posteriori direct sampling technique. The predefined exclusion criteria included a history of malignant or hepatobiliary disease, a history of deep vein thrombosis/pulmonary embolism (DVT/PE), an acute inflammatory state at the time of the venipuncture. We collected vein blood samples in Vacutest plastic coagulation tubes (Kima, Italy) and Vacutainer glass coagulation tubes (Beckton Dickinson, USA) containing 0.109 mol/l buffered trisodium citrate as an anticoagulant at a blood-to-anticoagulant ratio 9 : 1. We calculated two-sided reference interval and presented as 2.5th and 97.5th percentiles. ECLT values did not differ between sexes or the types of tubes enrolled in the study ( P  = 0.8979). The established reference interval ranged from 130 to 297 min for the KIMA Vacutest tube and from 120 to 292 min for the BD Vacutainer tube. The established ECLT reference interval differed significantly from the currently used cut-off value in our laboratory, thus enabling the assessment of hyperfibrinolysis by employing double-sided reference interval.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.90
自引率
0.00%
发文量
111
审稿时长
4-8 weeks
期刊介绍: Blood Coagulation & Fibrinolysis is an international fully refereed journal that features review and original research articles on all clinical, laboratory and experimental aspects of haemostasis and thrombosis. The journal is devoted to publishing significant developments worldwide in the field of blood coagulation, fibrinolysis, thrombosis, platelets and the kininogen-kinin system, as well as dealing with those aspects of blood rheology relevant to haemostasis and the effects of drugs on haemostatic components
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信