{"title":"基于b样条插值的水激励射频脉冲设计与优化的灵活框架。","authors":"Xavier Sieber, Ludovica Romanin, Jessica A M Bastiaansen, Christopher W Roy, Jérôme Yerly, Daniel Wenz, Jonas Richiardi, Matthias Stuber, Ruud B van Heeswijk","doi":"10.1002/mrm.30390","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To implement a flexible framework, named HydrOptiFrame, for the design and optimization of time-efficient water-excitation (WE) RF pulses using B-spline interpolation, and to characterize their lipid suppression performance.</p><p><strong>Methods: </strong>An evolutionary optimization algorithm was used to design WE RF pulses. The algorithm minimizes a composite loss function that quantifies the fat-water contrast using Bloch equation simulations. In a first study, B-spline interpolated optimized (BSIO) pulses designed with HydrOptiFrame with durations of 1 and 0.76 ms were generated for 3 T and characterized in healthy volunteers' knees. The femoral bone marrow SNR was compared to that obtained with to 1-1 WE and lipid insensitive binomial off resonant excitation (LIBRE) pulses. In a second study, in the heart at 1.5 T, the water-fat contrast ratio and coronary artery vessel length obtained with a 2.56 ms BSIO pulse was compared to 1-1 WE and LIBRE pulses in free-running cardiovascular MR.</p><p><strong>Results: </strong>The 1 ms BSIO pulse resulted in higher fat suppression and lower contrast ratio (CR) in the bone marrow than the state-of-the-art pulses (4.1 ± 0.2 vs. 4.7 ± 0.4 and 4.4 ± 0.3 for the BSIO, the 1-1 WE and LIBRE respectively, p < 0.05 vs. both) at 3 T. At 1.5 T, the BSIO pulse resulted in a higher blood-epicardial fat CR (3.8 ± 1.3 vs. 1.6 ± 0.6 and 2.4 ± 1.1 for the BSIO, 1-1 WE and LIBRE, respectively, p < 0.05 vs. both) and longer traceable left coronary artery vessel length (8.7 ± 1.4 cm vs. 7.0 ± 1.0 cm [p = 0.04] and 7.5 ± 1.2 cm [p = 0.09]).</p><p><strong>Conclusion: </strong>The HydrOptiFrame framework offers a new opportunity to design WE RF pulses that are robust to B<sub>0</sub> inhomogeneity at multiple magnetic field strengths and for variable RF pulse durations.</p>","PeriodicalId":18065,"journal":{"name":"Magnetic Resonance in Medicine","volume":" ","pages":""},"PeriodicalIF":3.0000,"publicationDate":"2024-12-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A flexible framework for the design and optimization of water-excitation RF pulses using B-spline interpolation.\",\"authors\":\"Xavier Sieber, Ludovica Romanin, Jessica A M Bastiaansen, Christopher W Roy, Jérôme Yerly, Daniel Wenz, Jonas Richiardi, Matthias Stuber, Ruud B van Heeswijk\",\"doi\":\"10.1002/mrm.30390\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To implement a flexible framework, named HydrOptiFrame, for the design and optimization of time-efficient water-excitation (WE) RF pulses using B-spline interpolation, and to characterize their lipid suppression performance.</p><p><strong>Methods: </strong>An evolutionary optimization algorithm was used to design WE RF pulses. The algorithm minimizes a composite loss function that quantifies the fat-water contrast using Bloch equation simulations. In a first study, B-spline interpolated optimized (BSIO) pulses designed with HydrOptiFrame with durations of 1 and 0.76 ms were generated for 3 T and characterized in healthy volunteers' knees. The femoral bone marrow SNR was compared to that obtained with to 1-1 WE and lipid insensitive binomial off resonant excitation (LIBRE) pulses. In a second study, in the heart at 1.5 T, the water-fat contrast ratio and coronary artery vessel length obtained with a 2.56 ms BSIO pulse was compared to 1-1 WE and LIBRE pulses in free-running cardiovascular MR.</p><p><strong>Results: </strong>The 1 ms BSIO pulse resulted in higher fat suppression and lower contrast ratio (CR) in the bone marrow than the state-of-the-art pulses (4.1 ± 0.2 vs. 4.7 ± 0.4 and 4.4 ± 0.3 for the BSIO, the 1-1 WE and LIBRE respectively, p < 0.05 vs. both) at 3 T. At 1.5 T, the BSIO pulse resulted in a higher blood-epicardial fat CR (3.8 ± 1.3 vs. 1.6 ± 0.6 and 2.4 ± 1.1 for the BSIO, 1-1 WE and LIBRE, respectively, p < 0.05 vs. both) and longer traceable left coronary artery vessel length (8.7 ± 1.4 cm vs. 7.0 ± 1.0 cm [p = 0.04] and 7.5 ± 1.2 cm [p = 0.09]).</p><p><strong>Conclusion: </strong>The HydrOptiFrame framework offers a new opportunity to design WE RF pulses that are robust to B<sub>0</sub> inhomogeneity at multiple magnetic field strengths and for variable RF pulse durations.</p>\",\"PeriodicalId\":18065,\"journal\":{\"name\":\"Magnetic Resonance in Medicine\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2024-12-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic Resonance in Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/mrm.30390\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance in Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/mrm.30390","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:采用一种名为HydrOptiFrame的灵活框架,利用B-样条插值法设计和优化具有时间效率的水激发(WE)射频脉冲,并鉴定其脂质抑制性能:方法:采用进化优化算法设计水激发射频脉冲。该算法通过布洛赫方程模拟,最小化量化脂水对比度的复合损失函数。在第一项研究中,使用 HydrOptiFrame 设计了持续时间为 1 和 0.76 毫秒的 B-样条插值优化(BSIO)脉冲,用于 3 T,并在健康志愿者膝关节中进行了表征。股骨髓信噪比与 1-1 WE 脉冲和脂质不敏感二叉离谐激励(LIBRE)脉冲的信噪比进行了比较。在第二项研究中,在 1.5 T 的心脏中,在自由运行的心血管磁共振中,将 2.56 毫秒 BSIO 脉冲与 1-1 WE 和 LIBRE 脉冲获得的水脂对比度和冠状动脉血管长度进行了比较:结果:与最先进的脉冲相比,1 毫秒 BSIO 脉冲在骨髓中的脂肪抑制更高,对比度 (CR) 更低(BSIO、1-1 WE 和 LIBRE 分别为 4.1 ± 0.2 vs. 4.7 ± 0.4 和 4.4 ± 0.3,p 结论:HydrOptiFrame框架为设计WE射频脉冲提供了一个新的机会,这种脉冲在多种磁场强度和可变射频脉冲持续时间下对B0不均匀性都很稳健。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A flexible framework for the design and optimization of water-excitation RF pulses using B-spline interpolation.

Purpose: To implement a flexible framework, named HydrOptiFrame, for the design and optimization of time-efficient water-excitation (WE) RF pulses using B-spline interpolation, and to characterize their lipid suppression performance.

Methods: An evolutionary optimization algorithm was used to design WE RF pulses. The algorithm minimizes a composite loss function that quantifies the fat-water contrast using Bloch equation simulations. In a first study, B-spline interpolated optimized (BSIO) pulses designed with HydrOptiFrame with durations of 1 and 0.76 ms were generated for 3 T and characterized in healthy volunteers' knees. The femoral bone marrow SNR was compared to that obtained with to 1-1 WE and lipid insensitive binomial off resonant excitation (LIBRE) pulses. In a second study, in the heart at 1.5 T, the water-fat contrast ratio and coronary artery vessel length obtained with a 2.56 ms BSIO pulse was compared to 1-1 WE and LIBRE pulses in free-running cardiovascular MR.

Results: The 1 ms BSIO pulse resulted in higher fat suppression and lower contrast ratio (CR) in the bone marrow than the state-of-the-art pulses (4.1 ± 0.2 vs. 4.7 ± 0.4 and 4.4 ± 0.3 for the BSIO, the 1-1 WE and LIBRE respectively, p < 0.05 vs. both) at 3 T. At 1.5 T, the BSIO pulse resulted in a higher blood-epicardial fat CR (3.8 ± 1.3 vs. 1.6 ± 0.6 and 2.4 ± 1.1 for the BSIO, 1-1 WE and LIBRE, respectively, p < 0.05 vs. both) and longer traceable left coronary artery vessel length (8.7 ± 1.4 cm vs. 7.0 ± 1.0 cm [p = 0.04] and 7.5 ± 1.2 cm [p = 0.09]).

Conclusion: The HydrOptiFrame framework offers a new opportunity to design WE RF pulses that are robust to B0 inhomogeneity at multiple magnetic field strengths and for variable RF pulse durations.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.70
自引率
24.20%
发文量
376
审稿时长
2-4 weeks
期刊介绍: Magnetic Resonance in Medicine (Magn Reson Med) is an international journal devoted to the publication of original investigations concerned with all aspects of the development and use of nuclear magnetic resonance and electron paramagnetic resonance techniques for medical applications. Reports of original investigations in the areas of mathematics, computing, engineering, physics, biophysics, chemistry, biochemistry, and physiology directly relevant to magnetic resonance will be accepted, as well as methodology-oriented clinical studies.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信