{"title":"低剂量肺部 CT:优化诊断辐射剂量 - 一项模型研究","authors":"Michael Zellner , Sebastian Tschauner , Mathias S. Weyland , Peter Eggenberger Hotz , Stephan Scheidegger , Christian J. Kellenberger","doi":"10.1016/j.ejro.2024.100614","DOIUrl":null,"url":null,"abstract":"<div><h3>Background/purpose</h3><div>To investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images.</div></div><div><h3>Methods</h3><div>Axial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings.</div></div><div><h3>Results</h3><div>Adequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA.</div></div><div><h3>Conclusions</h3><div>Determination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.</div></div>","PeriodicalId":38076,"journal":{"name":"European Journal of Radiology Open","volume":"13 ","pages":"Article 100614"},"PeriodicalIF":1.8000,"publicationDate":"2024-11-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Low-dose lung CT: Optimizing diagnostic radiation dose – A phantom study\",\"authors\":\"Michael Zellner , Sebastian Tschauner , Mathias S. Weyland , Peter Eggenberger Hotz , Stephan Scheidegger , Christian J. Kellenberger\",\"doi\":\"10.1016/j.ejro.2024.100614\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background/purpose</h3><div>To investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images.</div></div><div><h3>Methods</h3><div>Axial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings.</div></div><div><h3>Results</h3><div>Adequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA.</div></div><div><h3>Conclusions</h3><div>Determination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.</div></div>\",\"PeriodicalId\":38076,\"journal\":{\"name\":\"European Journal of Radiology Open\",\"volume\":\"13 \",\"pages\":\"Article 100614\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-11-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Journal of Radiology Open\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2352047724000698\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Radiology Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352047724000698","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Low-dose lung CT: Optimizing diagnostic radiation dose – A phantom study
Background/purpose
To investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images.
Methods
Axial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings.
Results
Adequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA.
Conclusions
Determination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.