术后散光眼内透镜对位试点研究:比较 Casia2 和 Photoshop 成像技术。

IF 1.8 4区 医学 Q3 OPHTHALMOLOGY
Journal of Ophthalmology Pub Date : 2024-11-14 eCollection Date: 2024-01-01 DOI:10.1155/2024/1053914
Baodi Yang, Chunxin Lai, Yongjie Qin, Hongliang Lin, Sheng Wang, Hailan Liao, Hongyang Zhang
{"title":"术后散光眼内透镜对位试点研究:比较 Casia2 和 Photoshop 成像技术。","authors":"Baodi Yang, Chunxin Lai, Yongjie Qin, Hongliang Lin, Sheng Wang, Hailan Liao, Hongyang Zhang","doi":"10.1155/2024/1053914","DOIUrl":null,"url":null,"abstract":"<p><p><b>Purpose:</b> Assess the comparative accuracy and reliability of postoperative toric intraocular lens (TIOL) alignment measurement methods: Casia2 and Adobe Photoshop with digital slit lamp images (PS method). <b>Methods:</b> In a study of 41 subjects with 58 eyes postimplantation of TIOL, we independently measured TIOL alignment with Casia2 and PS methods. The intraclass correlation coefficient (ICC<sub>1,1</sub>) was employed to assess the repeatability of the Casia2 method. While ICC<sub>2,1</sub> for absolute agreement and Bland-Altman analysis were utilized to determine the interdevice agreement between the two methods, the regression analysis was conducted to identify any proportional bias. <b>Results:</b> Casia2 demonstrated excellent intradevice repeatability with an ICC<sub>1,1</sub> of 0.998. The absolute agreement between Casia2 and PS was very high with an ICC<sub>2,1</sub> of 0.999. The average discrepancy between the two measurement methods was -0.828°, with a 95% confidence interval (CI) ranging from -1.623° to -0.032°. The 95% limits of agreement (LoA) were between -6.761° and 5.105°, indicating a strong concordance in TIOL alignment measurements. Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of pupil diameters (PDs) of 4 mm or greater. <b>Conclusion:</b> Casia2 and PS demonstrated significant concordance in measuring postoperative TIOL alignment, with Casia2 offering a more straightforward and accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.</p>","PeriodicalId":16674,"journal":{"name":"Journal of Ophthalmology","volume":"2024 ","pages":"1053914"},"PeriodicalIF":1.8000,"publicationDate":"2024-11-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11581804/pdf/","citationCount":"0","resultStr":"{\"title\":\"Pilot Study on Postoperative Toric Intraocular Lens Alignment: Comparing Casia2 and Photoshop Imaging Techniques.\",\"authors\":\"Baodi Yang, Chunxin Lai, Yongjie Qin, Hongliang Lin, Sheng Wang, Hailan Liao, Hongyang Zhang\",\"doi\":\"10.1155/2024/1053914\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b>Purpose:</b> Assess the comparative accuracy and reliability of postoperative toric intraocular lens (TIOL) alignment measurement methods: Casia2 and Adobe Photoshop with digital slit lamp images (PS method). <b>Methods:</b> In a study of 41 subjects with 58 eyes postimplantation of TIOL, we independently measured TIOL alignment with Casia2 and PS methods. The intraclass correlation coefficient (ICC<sub>1,1</sub>) was employed to assess the repeatability of the Casia2 method. While ICC<sub>2,1</sub> for absolute agreement and Bland-Altman analysis were utilized to determine the interdevice agreement between the two methods, the regression analysis was conducted to identify any proportional bias. <b>Results:</b> Casia2 demonstrated excellent intradevice repeatability with an ICC<sub>1,1</sub> of 0.998. The absolute agreement between Casia2 and PS was very high with an ICC<sub>2,1</sub> of 0.999. The average discrepancy between the two measurement methods was -0.828°, with a 95% confidence interval (CI) ranging from -1.623° to -0.032°. The 95% limits of agreement (LoA) were between -6.761° and 5.105°, indicating a strong concordance in TIOL alignment measurements. Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of pupil diameters (PDs) of 4 mm or greater. <b>Conclusion:</b> Casia2 and PS demonstrated significant concordance in measuring postoperative TIOL alignment, with Casia2 offering a more straightforward and accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.</p>\",\"PeriodicalId\":16674,\"journal\":{\"name\":\"Journal of Ophthalmology\",\"volume\":\"2024 \",\"pages\":\"1053914\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-11-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11581804/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Ophthalmology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1155/2024/1053914\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1155/2024/1053914","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:评估散光人工晶体(TIOL)术后配准测量方法的准确性和可靠性:Casia2和Adobe Photoshop与数字裂隙灯图像(PS方法)。方法:在对 41 名受试者的 58 只植入 TIOL 后的眼睛进行的研究中,我们使用 Casia2 和 PS 方法独立测量了 TIOL 的对准情况。采用类内相关系数(ICC1,1)来评估 Casia2 方法的可重复性。绝对一致性的 ICC2,1 和 Bland-Altman 分析用于确定两种方法的器械间一致性,而回归分析则用于确定任何比例偏差。结果:Casia2 的器械内重复性极佳,ICC1,1 为 0.998。Casia2 和 PS 的绝对一致性非常高,ICC2,1 为 0.999。两种测量方法的平均差异为-0.828°,95% 置信区间 (CI) 为-1.623°至-0.032°。95% 的一致性极限 (LoA) 介于 -6.761° 和 5.105° 之间,表明 TIOL 对齐测量的一致性很高。在瞳孔直径 (PD) 为 4 毫米或更大的条件下,Casia2 能够准确测量 TIOL 轴对齐情况。结论Casia2和PS在测量术后TIOL轴对位方面表现出明显的一致性,Casia2提供了一种更直接、更方便的替代方法,对瞳孔散大不理想的患者尤其有益。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Pilot Study on Postoperative Toric Intraocular Lens Alignment: Comparing Casia2 and Photoshop Imaging Techniques.

Purpose: Assess the comparative accuracy and reliability of postoperative toric intraocular lens (TIOL) alignment measurement methods: Casia2 and Adobe Photoshop with digital slit lamp images (PS method). Methods: In a study of 41 subjects with 58 eyes postimplantation of TIOL, we independently measured TIOL alignment with Casia2 and PS methods. The intraclass correlation coefficient (ICC1,1) was employed to assess the repeatability of the Casia2 method. While ICC2,1 for absolute agreement and Bland-Altman analysis were utilized to determine the interdevice agreement between the two methods, the regression analysis was conducted to identify any proportional bias. Results: Casia2 demonstrated excellent intradevice repeatability with an ICC1,1 of 0.998. The absolute agreement between Casia2 and PS was very high with an ICC2,1 of 0.999. The average discrepancy between the two measurement methods was -0.828°, with a 95% confidence interval (CI) ranging from -1.623° to -0.032°. The 95% limits of agreement (LoA) were between -6.761° and 5.105°, indicating a strong concordance in TIOL alignment measurements. Casia2 was capable of accurately measuring the TIOL axis alignment under conditions of pupil diameters (PDs) of 4 mm or greater. Conclusion: Casia2 and PS demonstrated significant concordance in measuring postoperative TIOL alignment, with Casia2 offering a more straightforward and accessible alternative, particularly beneficial for patients with suboptimal pupil dilation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Ophthalmology
Journal of Ophthalmology MEDICINE, RESEARCH & EXPERIMENTAL-OPHTHALMOLOGY
CiteScore
4.30
自引率
5.30%
发文量
194
审稿时长
6-12 weeks
期刊介绍: Journal of Ophthalmology is a peer-reviewed, Open Access journal that publishes original research articles, review articles, and clinical studies related to the anatomy, physiology and diseases of the eye. Submissions should focus on new diagnostic and surgical techniques, instrument and therapy updates, as well as clinical trials and research findings.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信