红细胞沉降率分析前变量的挑战:CUBE 30 touch 评估。

IF 1.3 4区 医学 Q4 MEDICINE, RESEARCH & EXPERIMENTAL
Flaminia Tomassetti, Roberto Guerranti, Roberto Leoncini, Carolina Pieroni, Daniela Diamanti, Michele Cirianni, Caterina Silvestrini, Lucrezia Galasso, Martina Pelagalli, Eleonora Nicolai, Alfredo Giovannelli, Massimo Pieri, Sergio Bernardini
{"title":"红细胞沉降率分析前变量的挑战:CUBE 30 touch 评估。","authors":"Flaminia Tomassetti, Roberto Guerranti, Roberto Leoncini, Carolina Pieroni, Daniela Diamanti, Michele Cirianni, Caterina Silvestrini, Lucrezia Galasso, Martina Pelagalli, Eleonora Nicolai, Alfredo Giovannelli, Massimo Pieri, Sergio Bernardini","doi":"10.1080/00365513.2024.2422397","DOIUrl":null,"url":null,"abstract":"<p><p>The erythrocyte sedimentation rate (ESR) is a widely used diagnostic test, influenced by all physiological and pathological conditions that can bias blood rheology by interfering factors. This study aimed to evaluate the performance of the CUBE 30 touch ESR analyzer in samples with preanalytical variables, as lipemia, hemolysis, and icterus or in presence of fibrinogen., Moreover we focused to define the maximum time limits to ensure a reliable ESR measure. Accuracy, intra-run and inter-run precision, and stability studies were performed. Moreover, hemolytic, jaundiced, lipemic samples and fibrinogen sensitivity were analyzed for interference study. Statistical analyses were performed. CUBE 30 touch and Westergren method comparison showed no statistical differences (Spearman Coefficient, R<sup>2</sup>=0,95). In the intra-run precision, the CV% mean obtained on samples with normal ESR level was 8,9%; with middle ESR level was 5,9% and with high ESR level the CV% was 4,3%. Inter-run precision test showed CV% of for single samples and overall samples in the range (12,3% for normal level and 4,8% for abnormal level). The samples stored at 4 °C showed good stability up to 3 h from collecting time. ESR samples showing lipemia, hemolysis or jaundice showed good correlations with the gold standard method (R<sup>2</sup> 0,901, 0,940, 0,911; <i>p</i> < 0,0001), however, Westergren tests were more sensitive than CUBE 30 touch to fibrinogen additions. The high comparability with the Westergren method, both in normal and interfering samples, and the good precision, support the usefulness of CUBE 30 touch in the clinical routine laboratory.</p>","PeriodicalId":21474,"journal":{"name":"Scandinavian Journal of Clinical & Laboratory Investigation","volume":" ","pages":"1-9"},"PeriodicalIF":1.3000,"publicationDate":"2024-11-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Challenges of preanalytical variables in erythrocyte sedimentation rate: a CUBE 30 touch evaluation.\",\"authors\":\"Flaminia Tomassetti, Roberto Guerranti, Roberto Leoncini, Carolina Pieroni, Daniela Diamanti, Michele Cirianni, Caterina Silvestrini, Lucrezia Galasso, Martina Pelagalli, Eleonora Nicolai, Alfredo Giovannelli, Massimo Pieri, Sergio Bernardini\",\"doi\":\"10.1080/00365513.2024.2422397\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The erythrocyte sedimentation rate (ESR) is a widely used diagnostic test, influenced by all physiological and pathological conditions that can bias blood rheology by interfering factors. This study aimed to evaluate the performance of the CUBE 30 touch ESR analyzer in samples with preanalytical variables, as lipemia, hemolysis, and icterus or in presence of fibrinogen., Moreover we focused to define the maximum time limits to ensure a reliable ESR measure. Accuracy, intra-run and inter-run precision, and stability studies were performed. Moreover, hemolytic, jaundiced, lipemic samples and fibrinogen sensitivity were analyzed for interference study. Statistical analyses were performed. CUBE 30 touch and Westergren method comparison showed no statistical differences (Spearman Coefficient, R<sup>2</sup>=0,95). In the intra-run precision, the CV% mean obtained on samples with normal ESR level was 8,9%; with middle ESR level was 5,9% and with high ESR level the CV% was 4,3%. Inter-run precision test showed CV% of for single samples and overall samples in the range (12,3% for normal level and 4,8% for abnormal level). The samples stored at 4 °C showed good stability up to 3 h from collecting time. ESR samples showing lipemia, hemolysis or jaundice showed good correlations with the gold standard method (R<sup>2</sup> 0,901, 0,940, 0,911; <i>p</i> < 0,0001), however, Westergren tests were more sensitive than CUBE 30 touch to fibrinogen additions. The high comparability with the Westergren method, both in normal and interfering samples, and the good precision, support the usefulness of CUBE 30 touch in the clinical routine laboratory.</p>\",\"PeriodicalId\":21474,\"journal\":{\"name\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"volume\":\" \",\"pages\":\"1-9\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-11-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Scandinavian Journal of Clinical & Laboratory Investigation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/00365513.2024.2422397\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICINE, RESEARCH & EXPERIMENTAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Scandinavian Journal of Clinical & Laboratory Investigation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/00365513.2024.2422397","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0

摘要

红细胞沉降率(ESR)是一种广泛使用的诊断测试,受各种生理和病理条件的影响,干扰因素会使血液流变学产生偏差。本研究旨在评估 CUBE 30 touch 血沉分析仪在分析前变量(如脂血、溶血和黄疸)或存在纤维蛋白原的样本中的性能。我们还进行了准确度、运行内和运行间精密度以及稳定性研究。此外,我们还对溶血、黄疸、脂血样本和纤维蛋白原敏感性进行了干扰研究分析。进行了统计分析。CUBE 30 touch 与 Westergren 方法的比较显示没有统计学差异(Spearman Coefficient,R2=0,95)。在运行内精确度方面,血沉正常样本的 CV% 平均值为 8.9%;血沉中等样本的 CV% 平均值为 5.9%;血沉高样本的 CV% 平均值为 4.3%。运行间精密度测试表明,单个样品和整体样品的 CV%在一定范围内(正常水平为 12.3%,异常水平为 4.8%)。在 4 °C 下保存的样本在采集后 3 小时内稳定性良好。出现脂血、溶血或黄疸的血沉样本与金标准方法有良好的相关性(R2 0,901, 0,940, 0,911; p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Challenges of preanalytical variables in erythrocyte sedimentation rate: a CUBE 30 touch evaluation.

The erythrocyte sedimentation rate (ESR) is a widely used diagnostic test, influenced by all physiological and pathological conditions that can bias blood rheology by interfering factors. This study aimed to evaluate the performance of the CUBE 30 touch ESR analyzer in samples with preanalytical variables, as lipemia, hemolysis, and icterus or in presence of fibrinogen., Moreover we focused to define the maximum time limits to ensure a reliable ESR measure. Accuracy, intra-run and inter-run precision, and stability studies were performed. Moreover, hemolytic, jaundiced, lipemic samples and fibrinogen sensitivity were analyzed for interference study. Statistical analyses were performed. CUBE 30 touch and Westergren method comparison showed no statistical differences (Spearman Coefficient, R2=0,95). In the intra-run precision, the CV% mean obtained on samples with normal ESR level was 8,9%; with middle ESR level was 5,9% and with high ESR level the CV% was 4,3%. Inter-run precision test showed CV% of for single samples and overall samples in the range (12,3% for normal level and 4,8% for abnormal level). The samples stored at 4 °C showed good stability up to 3 h from collecting time. ESR samples showing lipemia, hemolysis or jaundice showed good correlations with the gold standard method (R2 0,901, 0,940, 0,911; p < 0,0001), however, Westergren tests were more sensitive than CUBE 30 touch to fibrinogen additions. The high comparability with the Westergren method, both in normal and interfering samples, and the good precision, support the usefulness of CUBE 30 touch in the clinical routine laboratory.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.50
自引率
4.80%
发文量
85
审稿时长
4-8 weeks
期刊介绍: The Scandinavian Journal of Clinical and Laboratory Investigation is an international scientific journal covering clinically oriented biochemical and physiological research. Since the launch of the journal in 1949, it has been a forum for international laboratory medicine, closely related to, and edited by, The Scandinavian Society for Clinical Chemistry. The journal contains peer-reviewed articles, editorials, invited reviews, and short technical notes, as well as several supplements each year. Supplements consist of monographs, and symposium and congress reports covering subjects within clinical chemistry and clinical physiology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信