{"title":"西罗莫司洗脱铁生物可吸收支架在犬膝下动脉中的临床前评估。","authors":"Guoyi Sun, Wanqian Zhang, Tao Li, Gui Zhang, Haifeng Li, Xian Gao, Guoxian Zhu, Deyuan Zhang, Wei Guo","doi":"10.1016/j.ijcard.2024.132722","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and aims: </strong>Infrapopliteal artery intervention has yielded suboptimal results in below-the-knee (BTK) arterial disease. However, bioresorbable scaffolds are a potential treatment for this condition. This study compared the support performance, anti-hyperplastic effects, and histological manifestations between a sirolimus-eluting iron bioresorbable scaffold (IBS) and an everolimus-eluting stent (EES) within 6 months and the corrosion profile during 12 months in canine BTK arteries.</p><p><strong>Methods: </strong>Eighteen IBS and twelve EES systems were implanted into nonatherosclerotic BTK arteries in dogs. Support performance and inhibit intimal hyperplasia, histological manifestations, and the corrosion profile were compared between the two systems using angiography, optical coherence tomography (OCT), micro-computed tomography, and histopathologic evaluation at 1, 3, 6, and 12 months.</p><p><strong>Results: </strong>All systems were successfully implanted. There was no significant difference in the area of stenosis between the IBS and EES groups within 6 months. Semi-quantitative OCT revealed degradation of the IBS at 3 months, with strut corrosion rates of 24.6 %, 45.0 %, and 69.2 % at 3, 6, and 12 months. Micro-computed tomography showed that the IBS maintained its integrity at 1 month without corrosion, with more struts fractured because of degradation at 6 and 12 months. Endothelialization was complete at 1 month in both groups. There was no significant between-group difference in cell responses during 6 months of follow-up, with the exception of macrophages.</p><p><strong>Conclusion: </strong>This preclinical study suggests that the IBS has performance support, anti-hyperplastic effects, and histological manifestations comparable to those of the EES within 6 months, with a reasonable corrosion profile over 12 months.</p>","PeriodicalId":13710,"journal":{"name":"International journal of cardiology","volume":null,"pages":null},"PeriodicalIF":3.2000,"publicationDate":"2024-11-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Preclinical evaluation of the sirolimus-eluting iron bioresorbable scaffold in canine below-the-knee artery.\",\"authors\":\"Guoyi Sun, Wanqian Zhang, Tao Li, Gui Zhang, Haifeng Li, Xian Gao, Guoxian Zhu, Deyuan Zhang, Wei Guo\",\"doi\":\"10.1016/j.ijcard.2024.132722\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background and aims: </strong>Infrapopliteal artery intervention has yielded suboptimal results in below-the-knee (BTK) arterial disease. However, bioresorbable scaffolds are a potential treatment for this condition. This study compared the support performance, anti-hyperplastic effects, and histological manifestations between a sirolimus-eluting iron bioresorbable scaffold (IBS) and an everolimus-eluting stent (EES) within 6 months and the corrosion profile during 12 months in canine BTK arteries.</p><p><strong>Methods: </strong>Eighteen IBS and twelve EES systems were implanted into nonatherosclerotic BTK arteries in dogs. Support performance and inhibit intimal hyperplasia, histological manifestations, and the corrosion profile were compared between the two systems using angiography, optical coherence tomography (OCT), micro-computed tomography, and histopathologic evaluation at 1, 3, 6, and 12 months.</p><p><strong>Results: </strong>All systems were successfully implanted. There was no significant difference in the area of stenosis between the IBS and EES groups within 6 months. Semi-quantitative OCT revealed degradation of the IBS at 3 months, with strut corrosion rates of 24.6 %, 45.0 %, and 69.2 % at 3, 6, and 12 months. Micro-computed tomography showed that the IBS maintained its integrity at 1 month without corrosion, with more struts fractured because of degradation at 6 and 12 months. Endothelialization was complete at 1 month in both groups. There was no significant between-group difference in cell responses during 6 months of follow-up, with the exception of macrophages.</p><p><strong>Conclusion: </strong>This preclinical study suggests that the IBS has performance support, anti-hyperplastic effects, and histological manifestations comparable to those of the EES within 6 months, with a reasonable corrosion profile over 12 months.</p>\",\"PeriodicalId\":13710,\"journal\":{\"name\":\"International journal of cardiology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2024-11-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of cardiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1016/j.ijcard.2024.132722\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of cardiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.ijcard.2024.132722","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

背景和目的:膝下(BTK)动脉疾病的膝下动脉介入治疗效果并不理想。然而,生物可吸收支架是治疗这种疾病的一种潜在方法。本研究比较了西罗莫司洗脱铁生物可吸收支架(IBS)和依维莫司洗脱支架(EES)在犬BTK动脉中6个月内的支撑性能、抗过度增生效果和组织学表现,以及12个月内的腐蚀情况:将 18 个 IBS 和 12 个 EES 系统植入狗的非动脉粥样硬化 BTK 动脉。在 1、3、6 和 12 个月时,使用血管造影术、光学相干断层扫描(OCT)、微型计算机断层扫描和组织病理学评估,比较两种系统的支持性能和抑制内膜增生、组织学表现和腐蚀情况:结果:所有系统均成功植入。在 6 个月内,IBS 组和 EES 组的血管狭窄面积没有明显差异。半定量 OCT 显示,IBS 在 3 个月时出现退化,在 3、6 和 12 个月时支柱腐蚀率分别为 24.6%、45.0% 和 69.2%。显微计算机断层扫描显示,IBS 在 1 个月时保持了完整性,没有发生腐蚀,而在 6 个月和 12 个月时,由于降解而断裂的支柱较多。两组的内皮化在 1 个月时都已完成。随访 6 个月期间,除巨噬细胞外,细胞反应在组间无明显差异:这项临床前研究表明,IBS 在 6 个月内具有与 EES 相当的性能支持、抗增生作用和组织学表现,在 12 个月内具有合理的腐蚀性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Preclinical evaluation of the sirolimus-eluting iron bioresorbable scaffold in canine below-the-knee artery.

Background and aims: Infrapopliteal artery intervention has yielded suboptimal results in below-the-knee (BTK) arterial disease. However, bioresorbable scaffolds are a potential treatment for this condition. This study compared the support performance, anti-hyperplastic effects, and histological manifestations between a sirolimus-eluting iron bioresorbable scaffold (IBS) and an everolimus-eluting stent (EES) within 6 months and the corrosion profile during 12 months in canine BTK arteries.

Methods: Eighteen IBS and twelve EES systems were implanted into nonatherosclerotic BTK arteries in dogs. Support performance and inhibit intimal hyperplasia, histological manifestations, and the corrosion profile were compared between the two systems using angiography, optical coherence tomography (OCT), micro-computed tomography, and histopathologic evaluation at 1, 3, 6, and 12 months.

Results: All systems were successfully implanted. There was no significant difference in the area of stenosis between the IBS and EES groups within 6 months. Semi-quantitative OCT revealed degradation of the IBS at 3 months, with strut corrosion rates of 24.6 %, 45.0 %, and 69.2 % at 3, 6, and 12 months. Micro-computed tomography showed that the IBS maintained its integrity at 1 month without corrosion, with more struts fractured because of degradation at 6 and 12 months. Endothelialization was complete at 1 month in both groups. There was no significant between-group difference in cell responses during 6 months of follow-up, with the exception of macrophages.

Conclusion: This preclinical study suggests that the IBS has performance support, anti-hyperplastic effects, and histological manifestations comparable to those of the EES within 6 months, with a reasonable corrosion profile over 12 months.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
International journal of cardiology
International journal of cardiology 医学-心血管系统
CiteScore
6.80
自引率
5.70%
发文量
758
审稿时长
44 days
期刊介绍: The International Journal of Cardiology is devoted to cardiology in the broadest sense. Both basic research and clinical papers can be submitted. The journal serves the interest of both practicing clinicians and researchers. In addition to original papers, we are launching a range of new manuscript types, including Consensus and Position Papers, Systematic Reviews, Meta-analyses, and Short communications. Case reports are no longer acceptable. Controversial techniques, issues on health policy and social medicine are discussed and serve as useful tools for encouraging debate.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信