新型氧气清除面罩在将局部氧气浓度降至手术火灾风险阈值以下方面的功效:概念验证实验研究。

IF 2.6 Q1 SURGERY
Christopher D Yang, Teresa H Chen, Jeremiah P Tao
{"title":"新型氧气清除面罩在将局部氧气浓度降至手术火灾风险阈值以下方面的功效:概念验证实验研究。","authors":"Christopher D Yang, Teresa H Chen, Jeremiah P Tao","doi":"10.1186/s13037-024-00411-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>This study aims to evaluate the efficacy of an oxygen scavenging mask device in reducing local oxygen concentrations from nasal cannula ventilation compared to a standard open facial surgical field.</p><p><strong>Methods: </strong>This is a controlled experiment using a custom-fabricated silicone midfacial oxygen scavenging device, SimMan airway management trainer manikin (Laerdal Medical, Stavanger, Norway), handheld oxygen detector (Forensics Detectors, Los Angeles, United States) and oxygen from a Datex Ohmeda Aisys Carestation anesthesia unit (GE HealthCare, Chicago, United States). Oxygen concentrations were measured at 18 facial landmarks (Fig. 1) with nasal cannula flow of 2, 4, and 6 L/min of 100% FiO2 in both masked and unmasked conditions (Fig. 2).</p><p><strong>Results: </strong>The mean oxygen concentration in the facial surgical field was 20.95% with the scavenger mask and 24.8% without (P < 0.001; two-tailed paired t-test). The unmasked condition was associated with suprathreshold oxygen concentration levels at 13 of 18 facial landmarks (Table 1). The device significantly reduced local oxygen concentration at 16 of 18 facial landmarks (Table 1). The device provided safe oxygen concentration levels at all three flow rates, and measured oxygen concentrations directly correlated with oxygen flow rate in the unmasked condition (Table 2).</p><p><strong>Conclusions: </strong>An oxygen scavenger mask device reduced local oxygen concentrations from nasal cannula ventilation to below the 23% fire threshold in the entire facial surgical field external to the mask in these experiments. The device may reduce intraoperative fire risk in patients that require supplementary oxygen during surgery.</p>","PeriodicalId":46782,"journal":{"name":"Patient Safety in Surgery","volume":null,"pages":null},"PeriodicalIF":2.6000,"publicationDate":"2024-09-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11391775/pdf/","citationCount":"0","resultStr":"{\"title\":\"Efficacy of a novel oxygen scavenger mask in reducing local oxygen concentrations below the surgical fire risk threshold: an experimental proof-of-concept study.\",\"authors\":\"Christopher D Yang, Teresa H Chen, Jeremiah P Tao\",\"doi\":\"10.1186/s13037-024-00411-1\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>This study aims to evaluate the efficacy of an oxygen scavenging mask device in reducing local oxygen concentrations from nasal cannula ventilation compared to a standard open facial surgical field.</p><p><strong>Methods: </strong>This is a controlled experiment using a custom-fabricated silicone midfacial oxygen scavenging device, SimMan airway management trainer manikin (Laerdal Medical, Stavanger, Norway), handheld oxygen detector (Forensics Detectors, Los Angeles, United States) and oxygen from a Datex Ohmeda Aisys Carestation anesthesia unit (GE HealthCare, Chicago, United States). Oxygen concentrations were measured at 18 facial landmarks (Fig. 1) with nasal cannula flow of 2, 4, and 6 L/min of 100% FiO2 in both masked and unmasked conditions (Fig. 2).</p><p><strong>Results: </strong>The mean oxygen concentration in the facial surgical field was 20.95% with the scavenger mask and 24.8% without (P < 0.001; two-tailed paired t-test). The unmasked condition was associated with suprathreshold oxygen concentration levels at 13 of 18 facial landmarks (Table 1). The device significantly reduced local oxygen concentration at 16 of 18 facial landmarks (Table 1). The device provided safe oxygen concentration levels at all three flow rates, and measured oxygen concentrations directly correlated with oxygen flow rate in the unmasked condition (Table 2).</p><p><strong>Conclusions: </strong>An oxygen scavenger mask device reduced local oxygen concentrations from nasal cannula ventilation to below the 23% fire threshold in the entire facial surgical field external to the mask in these experiments. The device may reduce intraoperative fire risk in patients that require supplementary oxygen during surgery.</p>\",\"PeriodicalId\":46782,\"journal\":{\"name\":\"Patient Safety in Surgery\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2024-09-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11391775/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Patient Safety in Surgery\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1186/s13037-024-00411-1\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"SURGERY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Patient Safety in Surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s13037-024-00411-1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"SURGERY","Score":null,"Total":0}
引用次数: 0

摘要

背景:本研究旨在评估氧气清除面罩装置与标准开放式面部手术野相比,在降低鼻插管通气的局部氧气浓度方面的功效:本研究旨在评估与标准开放式面部手术野相比,氧气清除面罩装置在降低鼻插管通气产生的局部氧气浓度方面的功效:这是一项对照实验,使用了定制的硅胶中面部氧气清除装置、SimMan 气道管理训练模拟人(Laerdal Medical,挪威斯塔万格)、手持式氧气检测器(Forensics Detectors,美国洛杉矶)和来自 Datex Ohmeda Aisys Carestation 麻醉装置(GE HealthCare,美国芝加哥)的氧气。在戴面罩和不戴面罩的情况下(图 2),鼻插管流量为 2、4 和 6 升/分钟的 100% FiO2 时,在 18 个面部地标(图 1)处测量了氧气浓度:结果:使用清氧面罩时,面部手术区域的平均氧气浓度为 20.95%,而不使用清氧面罩时为 24.8%(P 结论:清氧面罩装置降低了面部手术区域的平均氧气浓度:在这些实验中,氧气清除器面罩装置将鼻插管通气产生的局部氧气浓度降至面罩外部整个面部手术区域的 23% 火阈值以下。该装置可降低手术期间需要补充氧气的患者的术中起火风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Efficacy of a novel oxygen scavenger mask in reducing local oxygen concentrations below the surgical fire risk threshold: an experimental proof-of-concept study.

Background: This study aims to evaluate the efficacy of an oxygen scavenging mask device in reducing local oxygen concentrations from nasal cannula ventilation compared to a standard open facial surgical field.

Methods: This is a controlled experiment using a custom-fabricated silicone midfacial oxygen scavenging device, SimMan airway management trainer manikin (Laerdal Medical, Stavanger, Norway), handheld oxygen detector (Forensics Detectors, Los Angeles, United States) and oxygen from a Datex Ohmeda Aisys Carestation anesthesia unit (GE HealthCare, Chicago, United States). Oxygen concentrations were measured at 18 facial landmarks (Fig. 1) with nasal cannula flow of 2, 4, and 6 L/min of 100% FiO2 in both masked and unmasked conditions (Fig. 2).

Results: The mean oxygen concentration in the facial surgical field was 20.95% with the scavenger mask and 24.8% without (P < 0.001; two-tailed paired t-test). The unmasked condition was associated with suprathreshold oxygen concentration levels at 13 of 18 facial landmarks (Table 1). The device significantly reduced local oxygen concentration at 16 of 18 facial landmarks (Table 1). The device provided safe oxygen concentration levels at all three flow rates, and measured oxygen concentrations directly correlated with oxygen flow rate in the unmasked condition (Table 2).

Conclusions: An oxygen scavenger mask device reduced local oxygen concentrations from nasal cannula ventilation to below the 23% fire threshold in the entire facial surgical field external to the mask in these experiments. The device may reduce intraoperative fire risk in patients that require supplementary oxygen during surgery.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.80
自引率
8.10%
发文量
37
审稿时长
9 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信