{"title":"使用六西格玛量表评估中国孕产妇第二胎产前筛查实践中的分析绩效","authors":"Jinming Zhang ,&nbsp;Xingtong Chen ,&nbsp;Jiaming Wu ,&nbsp;Penghui Feng ,&nbsp;Wei Wang ,&nbsp;Kun Zhong ,&nbsp;Shuai Yuan ,&nbsp;Yuxuan Du ,&nbsp;Chuanbao Zhang ,&nbsp;Falin He","doi":"10.1016/j.plabm.2024.e00422","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>We aimed to evaluate the analytical performance of second-trimester maternal serum screening in China, and to compare if there are differences in sigma levels across different methods and months.</p></div><div><h3>Methods</h3><p>A retrospective study was conducted to assess the analytical quality levels of laboratories by calculating the Sigma metrics with prenatal screening biomarkers: AFP, Total β-hCG, free β-hCG, uE3. Data from 591 laboratories were selected. Sigma metrics were computed using the formula: Sigma metrics(σ) = (%TEa - |%Bias|)/%CV. The Friedman test and Mann-Whitney test were used to compare differences across various methods and different months. The Hodges–Lehmann was used for determining 95 % confidence intervals of pseudo-medians.</p></div><div><h3>Results</h3><p>Only uE3 showed significant monthly variations in sigma calculations. However, around 8 % of laboratories across all four analytes demonstrated sigma levels both above 6 and below 3 in different months. Laboratories utilizing time-resolved fluorescence methods significantly outperformed those using chemiluminescence in sigma level. For AFP, the pseudo-median difference between these methods lies within a 95 % confidence interval of (−3.22, −1.93), while for uE3, it is at (−2.30, −1.40). Notably, the median sigma levels for all analytes reached the 4-sigma threshold, with free β-hCG even attaining the 6-sigma level.</p></div><div><h3>Conclusion</h3><p>With current standards, China's second-trimester maternal serum screening is of relatively high analytical quality, and variations in sigma levels exist across different months and methods.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"41 ","pages":"Article e00422"},"PeriodicalIF":1.7000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000684/pdfft?md5=94fce0bc72c07b868b95e4af86aeb411&pid=1-s2.0-S2352551724000684-main.pdf","citationCount":"0","resultStr":"{\"title\":\"An assessment of analytical performance using the six sigma scale in second-trimester maternal prenatal screening practices in China\",\"authors\":\"Jinming Zhang ,&nbsp;Xingtong Chen ,&nbsp;Jiaming Wu ,&nbsp;Penghui Feng ,&nbsp;Wei Wang ,&nbsp;Kun Zhong ,&nbsp;Shuai Yuan ,&nbsp;Yuxuan Du ,&nbsp;Chuanbao Zhang ,&nbsp;Falin He\",\"doi\":\"10.1016/j.plabm.2024.e00422\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><p>We aimed to evaluate the analytical performance of second-trimester maternal serum screening in China, and to compare if there are differences in sigma levels across different methods and months.</p></div><div><h3>Methods</h3><p>A retrospective study was conducted to assess the analytical quality levels of laboratories by calculating the Sigma metrics with prenatal screening biomarkers: AFP, Total β-hCG, free β-hCG, uE3. Data from 591 laboratories were selected. Sigma metrics were computed using the formula: Sigma metrics(σ) = (%TEa - |%Bias|)/%CV. The Friedman test and Mann-Whitney test were used to compare differences across various methods and different months. The Hodges–Lehmann was used for determining 95 % confidence intervals of pseudo-medians.</p></div><div><h3>Results</h3><p>Only uE3 showed significant monthly variations in sigma calculations. However, around 8 % of laboratories across all four analytes demonstrated sigma levels both above 6 and below 3 in different months. Laboratories utilizing time-resolved fluorescence methods significantly outperformed those using chemiluminescence in sigma level. For AFP, the pseudo-median difference between these methods lies within a 95 % confidence interval of (−3.22, −1.93), while for uE3, it is at (−2.30, −1.40). Notably, the median sigma levels for all analytes reached the 4-sigma threshold, with free β-hCG even attaining the 6-sigma level.</p></div><div><h3>Conclusion</h3><p>With current standards, China's second-trimester maternal serum screening is of relatively high analytical quality, and variations in sigma levels exist across different months and methods.</p></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":\"41 \",\"pages\":\"Article e00422\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000684/pdfft?md5=94fce0bc72c07b868b95e4af86aeb411&pid=1-s2.0-S2352551724000684-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2352551724000684\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000684","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的 我们旨在评估中国第二胎孕产妇血清筛查的分析性能,并比较不同方法和月份的西格玛水平是否存在差异。方法 我们进行了一项回顾性研究,通过计算产前筛查生物标志物的西格玛指标来评估实验室的分析质量水平:AFP、总β-hCG、游离β-hCG、uE3。选取了 591 个实验室的数据。西格玛指标的计算公式为Sigma metrics(σ) = (%TEa - |%Bias|)/%CV。弗里德曼检验和曼-惠特尼检验用于比较不同方法和不同月份之间的差异。结果只有 uE3 在西格玛计算中显示出显著的月度差异。然而,在所有四种分析物中,约有 8% 的实验室在不同月份的西格玛水平高于 6 和低于 3。使用时间分辨荧光方法的实验室在西格玛水平上明显优于使用化学发光方法的实验室。就甲胎蛋白而言,这些方法之间的假中值差异在 95 % 置信区间(-3.22,-1.93)内,而就 uE3 而言,则在 95 % 置信区间(-2.30,-1.40)内。值得注意的是,所有分析物的中位数西格玛水平都达到了 4 西格玛阈值,游离β-hCG 甚至达到了 6 西格玛水平。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An assessment of analytical performance using the six sigma scale in second-trimester maternal prenatal screening practices in China

Objectives

We aimed to evaluate the analytical performance of second-trimester maternal serum screening in China, and to compare if there are differences in sigma levels across different methods and months.

Methods

A retrospective study was conducted to assess the analytical quality levels of laboratories by calculating the Sigma metrics with prenatal screening biomarkers: AFP, Total β-hCG, free β-hCG, uE3. Data from 591 laboratories were selected. Sigma metrics were computed using the formula: Sigma metrics(σ) = (%TEa - |%Bias|)/%CV. The Friedman test and Mann-Whitney test were used to compare differences across various methods and different months. The Hodges–Lehmann was used for determining 95 % confidence intervals of pseudo-medians.

Results

Only uE3 showed significant monthly variations in sigma calculations. However, around 8 % of laboratories across all four analytes demonstrated sigma levels both above 6 and below 3 in different months. Laboratories utilizing time-resolved fluorescence methods significantly outperformed those using chemiluminescence in sigma level. For AFP, the pseudo-median difference between these methods lies within a 95 % confidence interval of (−3.22, −1.93), while for uE3, it is at (−2.30, −1.40). Notably, the median sigma levels for all analytes reached the 4-sigma threshold, with free β-hCG even attaining the 6-sigma level.

Conclusion

With current standards, China's second-trimester maternal serum screening is of relatively high analytical quality, and variations in sigma levels exist across different months and methods.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Practical Laboratory Medicine
Practical Laboratory Medicine Health Professions-Radiological and Ultrasound Technology
CiteScore
3.50
自引率
0.00%
发文量
40
审稿时长
7 weeks
期刊介绍: Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信