中国数字乳腺 X 射线平均腺体剂量调查及国家诊断参考水平建议》。

IF 1 4区 医学 Q4 ENVIRONMENTAL SCIENCES
Hui Xu, Baorong Yue, Jinsheng Cheng, Jun Deng, Yinping Su, Qiaoqiao Zhao, Ke Xue, Zechen Feng, Yantao Niu, Quanfu Sun
{"title":"中国数字乳腺 X 射线平均腺体剂量调查及国家诊断参考水平建议》。","authors":"Hui Xu, Baorong Yue, Jinsheng Cheng, Jun Deng, Yinping Su, Qiaoqiao Zhao, Ke Xue, Zechen Feng, Yantao Niu, Quanfu Sun","doi":"10.1097/HP.0000000000001853","DOIUrl":null,"url":null,"abstract":"<p><strong>Abstract: </strong>The primary purpose of this study was to report the mean glandular doses and to determine the national diagnostic reference levels for digital mammography based on data between 2016 and 2018 in China. The data from 19,076 mammograms (4,769 examinations) by random sampling from 118 digital mammography systems were compiled. Exposure factors included age, compressed breast thickness, kVp, mAs, target/filter combination, entrance surface air kerma, and mean glandular doses, which were retrospectively surveyed and recorded from the monitor. The national diagnostic reference levels (75th percentiles) in mean glandular dose were calculated across median value obtained for all included data and stratified to specific compressed breast thickness ranges. The patients' ages ranged from 22 to 88 y, with a median age of 45. The applied voltage and output medians were 28 kVp and 75.1 mAs for all exposure, respectively. The median CBTs were 45 mm and 48 mm for craniocaudal views and mediolateral oblique views, and the corresponding median mean glandular doses were 1.32 mGy and 1.40 mGy, respectively. The national diagnostic reference level at compressed breast thickness of 40-50 mm was 1.67 mGy for CC views and 1.71 mGy for MLO views. The median mean glandular doses varied significantly and increased with compressed breast thickness, demonstrating the necessity of establishing DRL according to breast thickness and optimizing the clinic's digital mammography practice in China.</p>","PeriodicalId":12976,"journal":{"name":"Health physics","volume":" ","pages":""},"PeriodicalIF":1.0000,"publicationDate":"2024-07-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A Survey of Mean Glandular Doses and Suggestions on National Diagnostic Reference Levels for Digital Mammography in China.\",\"authors\":\"Hui Xu, Baorong Yue, Jinsheng Cheng, Jun Deng, Yinping Su, Qiaoqiao Zhao, Ke Xue, Zechen Feng, Yantao Niu, Quanfu Sun\",\"doi\":\"10.1097/HP.0000000000001853\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Abstract: </strong>The primary purpose of this study was to report the mean glandular doses and to determine the national diagnostic reference levels for digital mammography based on data between 2016 and 2018 in China. The data from 19,076 mammograms (4,769 examinations) by random sampling from 118 digital mammography systems were compiled. Exposure factors included age, compressed breast thickness, kVp, mAs, target/filter combination, entrance surface air kerma, and mean glandular doses, which were retrospectively surveyed and recorded from the monitor. The national diagnostic reference levels (75th percentiles) in mean glandular dose were calculated across median value obtained for all included data and stratified to specific compressed breast thickness ranges. The patients' ages ranged from 22 to 88 y, with a median age of 45. The applied voltage and output medians were 28 kVp and 75.1 mAs for all exposure, respectively. The median CBTs were 45 mm and 48 mm for craniocaudal views and mediolateral oblique views, and the corresponding median mean glandular doses were 1.32 mGy and 1.40 mGy, respectively. The national diagnostic reference level at compressed breast thickness of 40-50 mm was 1.67 mGy for CC views and 1.71 mGy for MLO views. The median mean glandular doses varied significantly and increased with compressed breast thickness, demonstrating the necessity of establishing DRL according to breast thickness and optimizing the clinic's digital mammography practice in China.</p>\",\"PeriodicalId\":12976,\"journal\":{\"name\":\"Health physics\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2024-07-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Health physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/HP.0000000000001853\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENVIRONMENTAL SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Health physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/HP.0000000000001853","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENVIRONMENTAL SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

摘要:本研究的主要目的是根据中国2016年至2018年的数据,报告平均腺体剂量并确定国家数字乳腺X光摄影诊断参考水平。研究汇编了从118套数字乳腺X光摄影系统中随机抽样的19076例乳腺X光检查(4769次检查)的数据。暴露因素包括年龄、压缩乳腺厚度、kVp、mAs、靶/滤光片组合、入口表面空气热玛和平均腺体剂量,这些因素都是通过监控器进行回顾性调查和记录的。平均腺体剂量的国家诊断参考水平(第 75 百分位数)是根据所有纳入数据的中位值计算得出的,并按照特定的压缩乳腺厚度范围进行了分层。患者年龄从 22 岁到 88 岁不等,中位年龄为 45 岁。所有暴露的应用电压和输出中值分别为 28 kVp 和 75.1 mAs。颅尾切面和内外侧斜切面的 CBT 中位数分别为 45 mm 和 48 mm,相应的平均腺体剂量中位数分别为 1.32 mGy 和 1.40 mGy。在压缩乳房厚度为 40-50 毫米时,CC 切面的国家诊断参考水平为 1.67 mGy,MLO 切面为 1.71 mGy。中位平均腺体剂量差异显著,并随压缩乳腺厚度的增加而增加,这表明根据乳腺厚度确定DRL并优化中国临床数字乳腺X光检查实践的必要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A Survey of Mean Glandular Doses and Suggestions on National Diagnostic Reference Levels for Digital Mammography in China.

Abstract: The primary purpose of this study was to report the mean glandular doses and to determine the national diagnostic reference levels for digital mammography based on data between 2016 and 2018 in China. The data from 19,076 mammograms (4,769 examinations) by random sampling from 118 digital mammography systems were compiled. Exposure factors included age, compressed breast thickness, kVp, mAs, target/filter combination, entrance surface air kerma, and mean glandular doses, which were retrospectively surveyed and recorded from the monitor. The national diagnostic reference levels (75th percentiles) in mean glandular dose were calculated across median value obtained for all included data and stratified to specific compressed breast thickness ranges. The patients' ages ranged from 22 to 88 y, with a median age of 45. The applied voltage and output medians were 28 kVp and 75.1 mAs for all exposure, respectively. The median CBTs were 45 mm and 48 mm for craniocaudal views and mediolateral oblique views, and the corresponding median mean glandular doses were 1.32 mGy and 1.40 mGy, respectively. The national diagnostic reference level at compressed breast thickness of 40-50 mm was 1.67 mGy for CC views and 1.71 mGy for MLO views. The median mean glandular doses varied significantly and increased with compressed breast thickness, demonstrating the necessity of establishing DRL according to breast thickness and optimizing the clinic's digital mammography practice in China.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Health physics
Health physics 医学-公共卫生、环境卫生与职业卫生
CiteScore
4.20
自引率
0.00%
发文量
324
审稿时长
3-8 weeks
期刊介绍: Health Physics, first published in 1958, provides the latest research to a wide variety of radiation safety professionals including health physicists, nuclear chemists, medical physicists, and radiation safety officers with interests in nuclear and radiation science. The Journal allows professionals in these and other disciplines in science and engineering to stay on the cutting edge of scientific and technological advances in the field of radiation safety. The Journal publishes original papers, technical notes, articles on advances in practical applications, editorials, and correspondence. Journal articles report on the latest findings in theoretical, practical, and applied disciplines of epidemiology and radiation effects, radiation biology and radiation science, radiation ecology, and related fields.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信