{"title":"单个中心使用无腔导联进行左束支起搏的早期经验:病例系列。","authors":"Evan Jim Gunawan, Johan Johan, Dian Andina Munawar, Dian Larasati Munawar, Beny Hartono, Muhammad Munawar","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Left bundle branch pacing (LBBP) has been subject to increasing interest over the last few years due to its capacity for physiological conduction and its advantages compared to His bundle pacing. His bundle pacing has certain limitations, such as a small pacing area for the His bundle, a high threshold that leads to battery depletion, a low R-wave amplitude that may result in atrial or His oversensing, and ventricular signal undersensing. In this case series, four patients (two female and two male) aged 62.2 ± 8.4 years old with symptomatic sick sinus disease and no scar tissue in the interventricular septum underwent LBBP. All LBBPs were done with standard LBBP using a lumenless SelectSecure 3830 lead (Medtronic®, Minneapolis, USA) with a fixed helix. The lead parameters showed a good R-wave amplitudes (13 ± 7.4 mV) and a low threshold  (0.77 ± 0.17 V @ 0.4 ms). All patients were discharged on the next day. During follow-up period of 13.3 ± 12.9 months, all patients were well and no complications were noted. In conclusion, LBBP may be as an alternative of novel conduction pacing techniques and can be done relatively easy and safe, even with limited experience center.</p>","PeriodicalId":6889,"journal":{"name":"Acta medica Indonesiana","volume":null,"pages":null},"PeriodicalIF":0.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Early Experience of Left Bundle Branch Pacing with Lumenless Lead in a Single Center: A Case Series.\",\"authors\":\"Evan Jim Gunawan, Johan Johan, Dian Andina Munawar, Dian Larasati Munawar, Beny Hartono, Muhammad Munawar\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Left bundle branch pacing (LBBP) has been subject to increasing interest over the last few years due to its capacity for physiological conduction and its advantages compared to His bundle pacing. His bundle pacing has certain limitations, such as a small pacing area for the His bundle, a high threshold that leads to battery depletion, a low R-wave amplitude that may result in atrial or His oversensing, and ventricular signal undersensing. In this case series, four patients (two female and two male) aged 62.2 ± 8.4 years old with symptomatic sick sinus disease and no scar tissue in the interventricular septum underwent LBBP. All LBBPs were done with standard LBBP using a lumenless SelectSecure 3830 lead (Medtronic®, Minneapolis, USA) with a fixed helix. The lead parameters showed a good R-wave amplitudes (13 ± 7.4 mV) and a low threshold  (0.77 ± 0.17 V @ 0.4 ms). All patients were discharged on the next day. During follow-up period of 13.3 ± 12.9 months, all patients were well and no complications were noted. In conclusion, LBBP may be as an alternative of novel conduction pacing techniques and can be done relatively easy and safe, even with limited experience center.</p>\",\"PeriodicalId\":6889,\"journal\":{\"name\":\"Acta medica Indonesiana\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta medica Indonesiana\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta medica Indonesiana","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

摘要

左束支起搏(LBBP)因其生理传导能力和与 His 束起搏相比的优势,在过去几年中受到越来越多的关注。His 束起搏有一定的局限性,例如 His 束起搏区域较小、阈值较高导致电池耗尽、R 波振幅较低可能导致心房或 His 束过感应以及心室信号感应不足。在本病例系列中,四名患者(两女两男)接受了 LBBP 检查,他们的年龄为 62.2 ± 8.4 岁,均有症状性病窦疾病,室间隔无瘢痕组织。所有 LBBP 均使用固定螺旋的无腔 SelectSecure 3830 导联(Medtronic®,美国明尼阿波利斯)进行标准 LBBP。导联参数显示 R 波振幅良好(13 ± 7.4 mV),阈值较低(0.77 ± 0.17 V @ 0.4 ms)。所有患者均于第二天出院。在 13.3 ± 12.9 个月的随访期间,所有患者均无并发症。总之,LBBP 可以作为新型传导起搏技术的一种替代方法,而且即使在经验有限的中心也可以相对容易和安全地完成。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Early Experience of Left Bundle Branch Pacing with Lumenless Lead in a Single Center: A Case Series.

Left bundle branch pacing (LBBP) has been subject to increasing interest over the last few years due to its capacity for physiological conduction and its advantages compared to His bundle pacing. His bundle pacing has certain limitations, such as a small pacing area for the His bundle, a high threshold that leads to battery depletion, a low R-wave amplitude that may result in atrial or His oversensing, and ventricular signal undersensing. In this case series, four patients (two female and two male) aged 62.2 ± 8.4 years old with symptomatic sick sinus disease and no scar tissue in the interventricular septum underwent LBBP. All LBBPs were done with standard LBBP using a lumenless SelectSecure 3830 lead (Medtronic®, Minneapolis, USA) with a fixed helix. The lead parameters showed a good R-wave amplitudes (13 ± 7.4 mV) and a low threshold  (0.77 ± 0.17 V @ 0.4 ms). All patients were discharged on the next day. During follow-up period of 13.3 ± 12.9 months, all patients were well and no complications were noted. In conclusion, LBBP may be as an alternative of novel conduction pacing techniques and can be done relatively easy and safe, even with limited experience center.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Acta medica Indonesiana
Acta medica Indonesiana MEDICINE, GENERAL & INTERNAL-
CiteScore
2.30
自引率
0.00%
发文量
61
审稿时长
12 weeks
期刊介绍: Acta Medica Indonesiana – The Indonesian Journal of Internal Medicine is an open accessed online journal and comprehensive peer-reviewed medical journal published by the Indonesian Society of Internal Medicine since 1968. Our main mission is to encourage the novel and important science in the clinical area in internal medicine. We welcome authors for original articles (research), review articles, interesting case reports, special articles, clinical practices, and medical illustrations that focus on the clinical area of internal medicine. Subjects suitable for publication include, but are not limited to the following fields of: -Allergy and immunology -Emergency medicine -Cancer and stem cells -Cardiovascular -Endocrinology and Metabolism -Gastroenterology -Gerontology -Hematology -Hepatology -Tropical and Infectious Disease -Virology -Internal medicine -Psychosomatic -Pulmonology -Rheumatology -Renal and Hypertension -Thyroid
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信