{"title":"延展焦距眼内透镜与单焦点透镜的术后视觉功能。","authors":"Koji Komatsu, Yoichiro Masuda, Sei Tachibana, Kei Sano, Masanobu Iida, Kosuke Ichihara, Tetutaro Oki, Kota Fukai, Masayuki Tatemichi, Tadashi Nakano","doi":"10.3928/1081597X-20240522-01","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the effects of intraocular lens (IOL) decentration and tilt, as well as age, on postoperative visual function (corrected distance visual acuity [CDVA] and contrast sensitivity) by comparing an extended depth-of-focus IOL using higher order aspheric optics against a monofocal IOL from the same platform.</p><p><strong>Methods: </strong>This retrospective observational study targeted patients without other eye diseases who underwent surgery to implant the Tecnis Eyhance OptiBlue or the monofocal IOL Tecnis OptiBlue 1-Piece (J&J Vision) during cataract surgery from November 2021 to December 2022. The effects of age, axial length, IOL decentration, tilt, and corneal higher order aberrations (HOAs) on the postoperative 5 m CDVA and area under log contrast sensitivity function (AULCSF) under photopic and scotopic conditions were evaluated within 3 months of surgery.</p><p><strong>Results: </strong>No significant difference was found in postoperative CDVA between the Tecnis Eyhance OptiBlue group (n = 61 eyes) and the Tecnis OptiBlue 1-Piece group (n = 35 eyes), but AULCSF was significantly better in the Tecnis Eyhance OptiBlue group for photopic (1.58 ± 0.13 vs 1.46 ± 0.18; <i>P</i> = .002) and scotopic (1.71 ± 0.11 vs 1.59 ± 0.19; <i>P</i> = .002) eyes. Multivariate analysis showed a negative correlation between AULCSF and IOL decentration and age in the Tecnis Eyhance OptiBlue group (<i>P</i> < .01), with no significant correlation with tilt, axial length, and corneal HOAs.</p><p><strong>Conclusions: </strong>The Tecnis Eyhance OptiBlue yielded significantly better contrast sensitivity under photopic and scotopic conditions than the Tecnis OptiBlue 1-Piece. However, it is important to consider the effects of IOL decentration and age when evaluating the contrast sensitivity of the Tecnis Eyhance OptiBlue. <b>[<i>J Refract Surg</i>. 2024;40(7):e499-e505.]</b>.</p>","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"40 7","pages":"e499-e505"},"PeriodicalIF":2.9000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Postoperative Visual Function of Extended Depth-of-Focus Intraocular Lenses Versus Monofocal Lenses.\",\"authors\":\"Koji Komatsu, Yoichiro Masuda, Sei Tachibana, Kei Sano, Masanobu Iida, Kosuke Ichihara, Tetutaro Oki, Kota Fukai, Masayuki Tatemichi, Tadashi Nakano\",\"doi\":\"10.3928/1081597X-20240522-01\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To assess the effects of intraocular lens (IOL) decentration and tilt, as well as age, on postoperative visual function (corrected distance visual acuity [CDVA] and contrast sensitivity) by comparing an extended depth-of-focus IOL using higher order aspheric optics against a monofocal IOL from the same platform.</p><p><strong>Methods: </strong>This retrospective observational study targeted patients without other eye diseases who underwent surgery to implant the Tecnis Eyhance OptiBlue or the monofocal IOL Tecnis OptiBlue 1-Piece (J&J Vision) during cataract surgery from November 2021 to December 2022. The effects of age, axial length, IOL decentration, tilt, and corneal higher order aberrations (HOAs) on the postoperative 5 m CDVA and area under log contrast sensitivity function (AULCSF) under photopic and scotopic conditions were evaluated within 3 months of surgery.</p><p><strong>Results: </strong>No significant difference was found in postoperative CDVA between the Tecnis Eyhance OptiBlue group (n = 61 eyes) and the Tecnis OptiBlue 1-Piece group (n = 35 eyes), but AULCSF was significantly better in the Tecnis Eyhance OptiBlue group for photopic (1.58 ± 0.13 vs 1.46 ± 0.18; <i>P</i> = .002) and scotopic (1.71 ± 0.11 vs 1.59 ± 0.19; <i>P</i> = .002) eyes. Multivariate analysis showed a negative correlation between AULCSF and IOL decentration and age in the Tecnis Eyhance OptiBlue group (<i>P</i> < .01), with no significant correlation with tilt, axial length, and corneal HOAs.</p><p><strong>Conclusions: </strong>The Tecnis Eyhance OptiBlue yielded significantly better contrast sensitivity under photopic and scotopic conditions than the Tecnis OptiBlue 1-Piece. However, it is important to consider the effects of IOL decentration and age when evaluating the contrast sensitivity of the Tecnis Eyhance OptiBlue. <b>[<i>J Refract Surg</i>. 2024;40(7):e499-e505.]</b>.</p>\",\"PeriodicalId\":16951,\"journal\":{\"name\":\"Journal of refractive surgery\",\"volume\":\"40 7\",\"pages\":\"e499-e505\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of refractive surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3928/1081597X-20240522-01\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597X-20240522-01","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:通过比较使用高阶非球面光学技术的扩展焦深人工晶体与同一平台的单焦点人工晶体,评估眼内晶体(IOL)的分散和倾斜以及年龄对术后视觉功能(矫正远距离视力[CDVA]和对比敏感度)的影响:这项回顾性观察研究的对象是 2021 年 11 月至 2022 年 12 月期间接受白内障手术植入 Tecnis Eyhance OptiBlue 或单焦人工晶体 Tecnis OptiBlue 1-Piece(J&J Vision)的无其他眼疾患者。在手术后 3 个月内,评估了年龄、轴向长度、人工晶体分散度、倾斜度和角膜高阶像差(HOAs)对术后 5 米 CDVA 和对数对比敏感度函数下面积(AULCSF)的影响:Tecnis Eyhance OptiBlue 组(n = 61 只眼睛)和 Tecnis OptiBlue 1 片组(n = 35 只眼睛)的术后 CDVA 没有明显差异,但 Tecnis Eyhance OptiBlue 组的 AULCSF 在近视眼(1.58 ± 0.13 vs 1.46 ± 0.18;P = .002)和散光眼(1.71 ± 0.11 vs 1.59 ± 0.19;P = .002)明显更好。多变量分析显示,在Tecnis Eyhance OptiBlue组中,AULCSF和人工晶体分散度与年龄呈负相关(P < .01),与倾斜度、轴向长度和角膜HOA无显著相关性:结论:Tecnis Eyhance OptiBlue 与 Tecnis OptiBlue 1 片式眼镜相比,在光视和散光条件下的对比敏感度明显更高。不过,在评估 Tecnis Eyhance OptiBlue 的对比敏感度时,必须考虑人工晶体分散和年龄的影响。[J Refract Surg. 2024;40(7):e499-e505]。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Postoperative Visual Function of Extended Depth-of-Focus Intraocular Lenses Versus Monofocal Lenses.

Purpose: To assess the effects of intraocular lens (IOL) decentration and tilt, as well as age, on postoperative visual function (corrected distance visual acuity [CDVA] and contrast sensitivity) by comparing an extended depth-of-focus IOL using higher order aspheric optics against a monofocal IOL from the same platform.

Methods: This retrospective observational study targeted patients without other eye diseases who underwent surgery to implant the Tecnis Eyhance OptiBlue or the monofocal IOL Tecnis OptiBlue 1-Piece (J&J Vision) during cataract surgery from November 2021 to December 2022. The effects of age, axial length, IOL decentration, tilt, and corneal higher order aberrations (HOAs) on the postoperative 5 m CDVA and area under log contrast sensitivity function (AULCSF) under photopic and scotopic conditions were evaluated within 3 months of surgery.

Results: No significant difference was found in postoperative CDVA between the Tecnis Eyhance OptiBlue group (n = 61 eyes) and the Tecnis OptiBlue 1-Piece group (n = 35 eyes), but AULCSF was significantly better in the Tecnis Eyhance OptiBlue group for photopic (1.58 ± 0.13 vs 1.46 ± 0.18; P = .002) and scotopic (1.71 ± 0.11 vs 1.59 ± 0.19; P = .002) eyes. Multivariate analysis showed a negative correlation between AULCSF and IOL decentration and age in the Tecnis Eyhance OptiBlue group (P < .01), with no significant correlation with tilt, axial length, and corneal HOAs.

Conclusions: The Tecnis Eyhance OptiBlue yielded significantly better contrast sensitivity under photopic and scotopic conditions than the Tecnis OptiBlue 1-Piece. However, it is important to consider the effects of IOL decentration and age when evaluating the contrast sensitivity of the Tecnis Eyhance OptiBlue. [J Refract Surg. 2024;40(7):e499-e505.].

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.10
自引率
12.50%
发文量
160
审稿时长
4-8 weeks
期刊介绍: The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as: • Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics” • Supplemental videos and materials available for many articles • Access to current articles, as well as several years of archived content • Articles posted online just 2 months after acceptance.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信