{"title":"通过 Tomey OA-2000 Biometer 和 Spectralis OCT 获得的植入式准分子透镜中心穹窿的重复性和一致性。","authors":"Hao Wu, Zuocheng Wang, Pengfei Wang, Yifei Meng, Zengying Wang, Yuhong Xue, Bohua Jiang, Shuaixi Pan, Zhipeng Yan","doi":"10.1155/2024/3684626","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT).</p><p><strong>Methods: </strong>In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired <i>t</i>-test and Bland-Altman plots were used to analyze the differences and agreements of central vault measured by two devices.</p><p><strong>Results: </strong>Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired <i>t</i>-test showed that central vault measured by Tomey OA-2000 biometer was -7.25 ± 23.57 microns lower than that measured by Spectralis OCT (<i>P</i> = 0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was -38.94 to 53.44 <i>μ</i>m.</p><p><strong>Conclusion: </strong>Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.</p>","PeriodicalId":16674,"journal":{"name":"Journal of Ophthalmology","volume":"2024 ","pages":"3684626"},"PeriodicalIF":1.8000,"publicationDate":"2024-06-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11217569/pdf/","citationCount":"0","resultStr":"{\"title\":\"Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT.\",\"authors\":\"Hao Wu, Zuocheng Wang, Pengfei Wang, Yifei Meng, Zengying Wang, Yuhong Xue, Bohua Jiang, Shuaixi Pan, Zhipeng Yan\",\"doi\":\"10.1155/2024/3684626\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT).</p><p><strong>Methods: </strong>In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired <i>t</i>-test and Bland-Altman plots were used to analyze the differences and agreements of central vault measured by two devices.</p><p><strong>Results: </strong>Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired <i>t</i>-test showed that central vault measured by Tomey OA-2000 biometer was -7.25 ± 23.57 microns lower than that measured by Spectralis OCT (<i>P</i> = 0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was -38.94 to 53.44 <i>μ</i>m.</p><p><strong>Conclusion: </strong>Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.</p>\",\"PeriodicalId\":16674,\"journal\":{\"name\":\"Journal of Ophthalmology\",\"volume\":\"2024 \",\"pages\":\"3684626\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-06-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11217569/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Ophthalmology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1155/2024/3684626\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1155/2024/3684626","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
目的评估通过 Tomey OA-2000 生物测量仪和 Spectralis 光学相干断层扫描(OCT)测量的植入式有晶体眼(ICL)中央穹隆的重复性和一致性:在这项前瞻性研究中,使用Tomey OA-2000生物测量仪和Spectralis光学相干断层扫描测量了ICL植入术后六个月随访的84只眼睛(43名患者)的中央穹隆。由一名经验丰富的技术人员在同一天使用 Tomey OA-2000 和 Spectralis OCT 连续扫描三次。计算变异系数(CoV)、类内相关系数(ICC)、受试者内标准偏差(Sw)和 2.77 Sw,以评估重复性和再现性。采用配对 t 检验和 Bland-Altman 图分析两种仪器测量的中央穹隆的差异和一致性:结果:Tomey OA-2000 生物测量仪和 Spectralis OCT 测量的中央穹隆的重复性显示,CoV 分别为 2.71% 和 1.66%。两种设备的 ICC 分别为 0.996 和 0.999。配对 t 检验显示,Tomey OA-2000 生物测量仪测量的中心穹窿比 Spectralis OCT 测量的中心穹窿低 -7.25 ± 23.57 微米(P = 0.006)。Tomey OA-2000 和 ASM-OCT 测量值之间的平均差异(95% 的一致性限值)为 -38.94 至 53.44 微米:结论:Tomey OA-2000 生物测量仪和 Spectralis OCT 在测量 ICL 中央穹隆时均显示出良好的重复性。Tomey OA-2000 生物测量仪和 Spectralis OCT 之间具有良好的可靠性和一致性。这两种仪器在中心穹隆测量中都很有用,但不能相互替代。
Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT.
Objective: To assess repeatability and agreement of central vault for implantable collamer lens (ICL) measured by the Tomey OA-2000 biometry and Spectralis optical coherence tomography (OCT).
Methods: In this prospective study, the central vault was measured by the Tomey OA-2000 biometer and Spectralis OCT in 84 eyes (43 patients) after ICL implantation at six month follow-up. Three consecutive scans were obtained by one experienced technician using Tomey OA-2000 and the Spectralis OCT in the same day. The coefficient of variation (CoV), intraclass correlation coefficient (ICC), within-subject standard deviation (Sw), and 2.77 Sw were calculated to assess the repeatability and reproducibility. The paired t-test and Bland-Altman plots were used to analyze the differences and agreements of central vault measured by two devices.
Results: Repeatability of the central vault measured by Tomey OA-2000 biometer and Spectralis OCT showed that the CoV was 2.71% and 1.66%, respectively. The ICC for both devices was 0.996 and 0.999, respectively. The paired t-test showed that central vault measured by Tomey OA-2000 biometer was -7.25 ± 23.57 microns lower than that measured by Spectralis OCT (P = 0.006). The mean difference between measurements for Tomey OA-2000 and ASM-OCT with 95% limits of agreement (LoAs) was -38.94 to 53.44 μm.
Conclusion: Both Tomey OA-2000 biometer and Spectralis OCT displayed good repeatability for the measurement of central vault of ICL. Good reliability and agreement were observed between Tomey OA-2000 biometer and Spectralis OCT. Both instruments are useful but not replaced each other for central vault measurements.
期刊介绍:
Journal of Ophthalmology is a peer-reviewed, Open Access journal that publishes original research articles, review articles, and clinical studies related to the anatomy, physiology and diseases of the eye. Submissions should focus on new diagnostic and surgical techniques, instrument and therapy updates, as well as clinical trials and research findings.