芬兰放射治疗病人定位中的锥形束 CT 剂量--前列腺治疗。

Pub Date : 2024-06-18 DOI:10.1093/rpd/ncae133
Teemu Siiskonen, Saara Alenius, Tiina Seppälä, Joonas Tikkanen, Miia Nadhum, Jarkko Ojala
{"title":"芬兰放射治疗病人定位中的锥形束 CT 剂量--前列腺治疗。","authors":"Teemu Siiskonen, Saara Alenius, Tiina Seppälä, Joonas Tikkanen, Miia Nadhum, Jarkko Ojala","doi":"10.1093/rpd/ncae133","DOIUrl":null,"url":null,"abstract":"<p><p>Imaging parameters, frequencies and resulting patient organ doses in treatments of prostate cancer were assessed in Finnish radiotherapy centres. Based on a questionnaire to the clinics, Monte Carlo method was used to estimate organ doses in International Commission on Radiological Protection standard phantom for prostate, bladder, rectum and femoral head. The results show that doses from cone beam computed tomography imaging have reduced compared to earlier studies and are between 3.6 and 34.5 mGy per image for the above-mentioned organs and for normal sized patients. There still is room for further optimization of the patient exposure, as many centres use the default imaging parameters, and the length of the imaged region may not be optimal for the purpose.</p>","PeriodicalId":0,"journal":{"name":"","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2024-06-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Cone beam CT doses in radiotherapy patient positioning in Finland-prostate treatments.\",\"authors\":\"Teemu Siiskonen, Saara Alenius, Tiina Seppälä, Joonas Tikkanen, Miia Nadhum, Jarkko Ojala\",\"doi\":\"10.1093/rpd/ncae133\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Imaging parameters, frequencies and resulting patient organ doses in treatments of prostate cancer were assessed in Finnish radiotherapy centres. Based on a questionnaire to the clinics, Monte Carlo method was used to estimate organ doses in International Commission on Radiological Protection standard phantom for prostate, bladder, rectum and femoral head. The results show that doses from cone beam computed tomography imaging have reduced compared to earlier studies and are between 3.6 and 34.5 mGy per image for the above-mentioned organs and for normal sized patients. There still is room for further optimization of the patient exposure, as many centres use the default imaging parameters, and the length of the imaged region may not be optimal for the purpose.</p>\",\"PeriodicalId\":0,\"journal\":{\"name\":\"\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0,\"publicationDate\":\"2024-06-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"\",\"FirstCategoryId\":\"93\",\"ListUrlMain\":\"https://doi.org/10.1093/rpd/ncae133\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncae133","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

芬兰放射治疗中心对前列腺癌治疗中的成像参数、频率和由此产生的患者器官剂量进行了评估。根据对诊所的问卷调查,采用蒙特卡洛法估算了国际放射防护委员会标准模型中前列腺、膀胱、直肠和股骨头的器官剂量。结果表明,锥形束计算机断层扫描成像的剂量与之前的研究相比有所降低,对于上述器官和正常体型的患者,每幅图像的剂量介于 3.6 至 34.5 mGy 之间。由于许多中心使用默认的成像参数,而且成像区域的长度可能并非最佳,因此仍有进一步优化患者照射剂量的余地。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享
查看原文
Cone beam CT doses in radiotherapy patient positioning in Finland-prostate treatments.

Imaging parameters, frequencies and resulting patient organ doses in treatments of prostate cancer were assessed in Finnish radiotherapy centres. Based on a questionnaire to the clinics, Monte Carlo method was used to estimate organ doses in International Commission on Radiological Protection standard phantom for prostate, bladder, rectum and femoral head. The results show that doses from cone beam computed tomography imaging have reduced compared to earlier studies and are between 3.6 and 34.5 mGy per image for the above-mentioned organs and for normal sized patients. There still is room for further optimization of the patient exposure, as many centres use the default imaging parameters, and the length of the imaged region may not be optimal for the purpose.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信