[18F]-PSMA-1007正电子发射计算机断层成像的优化和评分者之间的可靠性--由四位放射科医生读取的三种不同重建的比较。

IF 1.3 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Nuclear Medicine Communications Pub Date : 2024-05-01 Epub Date: 2024-02-05 DOI:10.1097/MNM.0000000000001823
Olof Jonmarker, Ted Nilsson, Rimma Axelsson, Lovisa Hult Ericson, Thuy A Tran, Antonios Tzortzakakis, Irina Savitcheva, Maria Holstensson
{"title":"[18F]-PSMA-1007正电子发射计算机断层成像的优化和评分者之间的可靠性--由四位放射科医生读取的三种不同重建的比较。","authors":"Olof Jonmarker, Ted Nilsson, Rimma Axelsson, Lovisa Hult Ericson, Thuy A Tran, Antonios Tzortzakakis, Irina Savitcheva, Maria Holstensson","doi":"10.1097/MNM.0000000000001823","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To increase understanding of optimal imaging parameters [ 18 F]PSMA-1007 when imaging patients with prostate cancer and to determine interrater agreement using [ 18 F]PSMA-1007.</p><p><strong>Methods: </strong>In this observational study, four independent physicians read reconstruction sets using bedtimes of 1, 2 and 3 minutes of patients undergoing [ 18 F]PSMA-1007. positron emission topography. Clear and equivocal lesions and their locations were recorded. Image noise was rated on a four-point scale. Lesion counts were compared using inter-class correlation whereas noise ratings were compared using generalized estimating equations. Repeated cases were used to assess intra-rater agreement.</p><p><strong>Results: </strong>Sixty reconstruction sets of 16 consecutively examined participants were included. Participants had a mean age of 71.5 years, six of them were examined prior to any treatment, three had a history of radiotherapy and seven of prostatectomy. Median Gleason score of primary tumors was 7. Imaging was performed after a mean of 132 min using a mean 3.95 MBq/Kg body weight of [ 18 F] PSMA-1007. Neither the total number of lesions per location nor the proportion of equivocal lesions varied consistently between bedtimes. Inter-rater reliability scores varied depending on location from 0.40 to 1.0 and were similar for all bedtimes. Intra-rater reliability varied between 0.70 and 0.76 for the three different bedtimes. Noise ratings were significantly lower for 1 minute than 3 minutes per bed.</p><p><strong>Conclusion: </strong>In the setting of [ 18 F]PSMA-1007 PET CT, 1, 2 and 3 minutes per bed produce similar results unlikely to affect clinical interpretation. Image noise ratings favor 2 and 3 minutes per bed.</p>","PeriodicalId":19708,"journal":{"name":"Nuclear Medicine Communications","volume":" ","pages":"389-395"},"PeriodicalIF":1.3000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[ 18 F]-PSMA-1007 PET imaging optimization and inter-rater reliability - a comparison of three different reconstructions read by four radiologists.\",\"authors\":\"Olof Jonmarker, Ted Nilsson, Rimma Axelsson, Lovisa Hult Ericson, Thuy A Tran, Antonios Tzortzakakis, Irina Savitcheva, Maria Holstensson\",\"doi\":\"10.1097/MNM.0000000000001823\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>To increase understanding of optimal imaging parameters [ 18 F]PSMA-1007 when imaging patients with prostate cancer and to determine interrater agreement using [ 18 F]PSMA-1007.</p><p><strong>Methods: </strong>In this observational study, four independent physicians read reconstruction sets using bedtimes of 1, 2 and 3 minutes of patients undergoing [ 18 F]PSMA-1007. positron emission topography. Clear and equivocal lesions and their locations were recorded. Image noise was rated on a four-point scale. Lesion counts were compared using inter-class correlation whereas noise ratings were compared using generalized estimating equations. Repeated cases were used to assess intra-rater agreement.</p><p><strong>Results: </strong>Sixty reconstruction sets of 16 consecutively examined participants were included. Participants had a mean age of 71.5 years, six of them were examined prior to any treatment, three had a history of radiotherapy and seven of prostatectomy. Median Gleason score of primary tumors was 7. Imaging was performed after a mean of 132 min using a mean 3.95 MBq/Kg body weight of [ 18 F] PSMA-1007. Neither the total number of lesions per location nor the proportion of equivocal lesions varied consistently between bedtimes. Inter-rater reliability scores varied depending on location from 0.40 to 1.0 and were similar for all bedtimes. Intra-rater reliability varied between 0.70 and 0.76 for the three different bedtimes. Noise ratings were significantly lower for 1 minute than 3 minutes per bed.</p><p><strong>Conclusion: </strong>In the setting of [ 18 F]PSMA-1007 PET CT, 1, 2 and 3 minutes per bed produce similar results unlikely to affect clinical interpretation. Image noise ratings favor 2 and 3 minutes per bed.</p>\",\"PeriodicalId\":19708,\"journal\":{\"name\":\"Nuclear Medicine Communications\",\"volume\":\" \",\"pages\":\"389-395\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nuclear Medicine Communications\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MNM.0000000000001823\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/2/5 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Medicine Communications","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MNM.0000000000001823","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/5 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的加深对前列腺癌患者成像时最佳成像参数[18 F]PSMA-1007 的理解,并确定使用[18 F]PSMA-1007 的医生之间的一致性:在这项观察性研究中,四名独立的医生分别以 1、2 和 3 分钟的上床时间阅读接受[18 F]PSMA-1007 正电子发射地形图检查的患者的重建集。记录清晰和模糊病灶及其位置。图像噪音按四级评分。病灶计数使用类间相关性进行比较,而噪声评分则使用广义估计方程进行比较。重复病例用于评估评分者内部的一致性:结果:共纳入了连续检查的 16 名参与者的 60 组重建数据。参与者的平均年龄为 71.5 岁,其中 6 人在接受任何治疗前接受过检查,3 人接受过放射治疗,7 人接受过前列腺切除术。原发性肿瘤的格里森评分中位数为 7 分。造影平均在 132 分钟后进行,[18 F] PSMA-1007 的平均剂量为 3.95 MBq/Kg(体重)。每个部位的病变总数和等位病变比例在不同就寝时间的变化都不一致。评分员之间的信度评分因位置而异,从 0.40 到 1.0 不等,所有就寝时间的评分都相似。三个不同就寝时间的评分者内部信度在 0.70 到 0.76 之间。每床 1 分钟的噪音评分明显低于 3 分钟的评分:结论:在[18 F]PSMA-1007 PET CT中,每床1分钟、2分钟和3分钟产生的结果相似,不会影响临床解释。图像噪音评分倾向于每床 2 分钟和 3 分钟。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[ 18 F]-PSMA-1007 PET imaging optimization and inter-rater reliability - a comparison of three different reconstructions read by four radiologists.

Objectives: To increase understanding of optimal imaging parameters [ 18 F]PSMA-1007 when imaging patients with prostate cancer and to determine interrater agreement using [ 18 F]PSMA-1007.

Methods: In this observational study, four independent physicians read reconstruction sets using bedtimes of 1, 2 and 3 minutes of patients undergoing [ 18 F]PSMA-1007. positron emission topography. Clear and equivocal lesions and their locations were recorded. Image noise was rated on a four-point scale. Lesion counts were compared using inter-class correlation whereas noise ratings were compared using generalized estimating equations. Repeated cases were used to assess intra-rater agreement.

Results: Sixty reconstruction sets of 16 consecutively examined participants were included. Participants had a mean age of 71.5 years, six of them were examined prior to any treatment, three had a history of radiotherapy and seven of prostatectomy. Median Gleason score of primary tumors was 7. Imaging was performed after a mean of 132 min using a mean 3.95 MBq/Kg body weight of [ 18 F] PSMA-1007. Neither the total number of lesions per location nor the proportion of equivocal lesions varied consistently between bedtimes. Inter-rater reliability scores varied depending on location from 0.40 to 1.0 and were similar for all bedtimes. Intra-rater reliability varied between 0.70 and 0.76 for the three different bedtimes. Noise ratings were significantly lower for 1 minute than 3 minutes per bed.

Conclusion: In the setting of [ 18 F]PSMA-1007 PET CT, 1, 2 and 3 minutes per bed produce similar results unlikely to affect clinical interpretation. Image noise ratings favor 2 and 3 minutes per bed.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.20
自引率
6.70%
发文量
212
审稿时长
3-8 weeks
期刊介绍: Nuclear Medicine Communications, the official journal of the British Nuclear Medicine Society, is a rapid communications journal covering nuclear medicine and molecular imaging with radionuclides, and the basic supporting sciences. As well as clinical research and commentary, manuscripts describing research on preclinical and basic sciences (radiochemistry, radiopharmacy, radiobiology, radiopharmacology, medical physics, computing and engineering, and technical and nursing professions involved in delivering nuclear medicine services) are welcomed, as the journal is intended to be of interest internationally to all members of the many medical and non-medical disciplines involved in nuclear medicine. In addition to papers reporting original studies, frankly written editorials and topical reviews are a regular feature of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信