降低无针连接器鲍曼不动杆菌 blaOXA-23 感染率的消毒程序的效果:一项体外研究

IF 1.8 Q3 INFECTIOUS DISEASES
C. Biazus-Dalcin , T.C.M. Sincero , C.P. Zamparette , D.C. Tartari , S. de Souza , T.L. Silva , A. Tomazoni , P.K. Rocha
{"title":"降低无针连接器鲍曼不动杆菌 blaOXA-23 感染率的消毒程序的效果:一项体外研究","authors":"C. Biazus-Dalcin ,&nbsp;T.C.M. Sincero ,&nbsp;C.P. Zamparette ,&nbsp;D.C. Tartari ,&nbsp;S. de Souza ,&nbsp;T.L. Silva ,&nbsp;A. Tomazoni ,&nbsp;P.K. Rocha","doi":"10.1016/j.infpip.2023.100328","DOIUrl":null,"url":null,"abstract":"<div><h3>Aim</h3><p>This study aimed to verify the efficacy of disinfection procedures to reduce <em>Acinetobacter baumannii bla</em><sub>OXA-23</sub> bacterial load in needleless connectors that had been experimentally contaminated.</p></div><div><h3>Methods</h3><p>Two-way intermediate extender's hub and needle-free valve were contaminated with <em>Acinetobacter baumannii bla</em><sub>OXA-23</sub>. To disinfect them, the following procedures were carried out: sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% isopropyl alcohol single-use cap, with eight times friction for 10 s, followed by 5 s drying time. The statistical tests Kruskal–Wallis and post-hoc Conover were performed using MedCalc®.</p></div><div><h3>Results</h3><p>A total of 82 experiments were conducted. All tested disinfection procedures were efficacious in reducing the A. <em>baumannii bla</em><sub>OXA-23</sub> load. The 70% IPA single-use cap was found to be the best method for disinfecting the two-way intermediate extender's hub (87.28%), while all the methods were efficacious for the disinfection of the needle-free valve (more than 90%). During the inoculation period, <em>A. baumannii bla</em><sub>OXA-23</sub> showed less adherence to the needle-free valve during the inoculation period, probably due to the device's design.</p></div><div><h3>Conclusion</h3><p>The three tested disinfection procedures using sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% IPA single-use cap were found to be efficacious in reducing the bacterial load of <em>A. baumanni bla</em><sub><em>OXA-23</em></sub> in needleless connectors. Proper disinfection of needleless connectors is a crucial nursing practice to prevent bloodstream infections, as it significantly reduces the bacterial load present in the device.</p></div>","PeriodicalId":33492,"journal":{"name":"Infection Prevention in Practice","volume":"6 1","pages":"Article 100328"},"PeriodicalIF":1.8000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2590088923000616/pdfft?md5=61d8e715701963a973546981466696b6&pid=1-s2.0-S2590088923000616-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Efficacy of disinfection procedures to reduce Acinetobacter baumanii blaOXA-23 contamination rate of needleless connectors: an in-vitro study\",\"authors\":\"C. Biazus-Dalcin ,&nbsp;T.C.M. Sincero ,&nbsp;C.P. Zamparette ,&nbsp;D.C. Tartari ,&nbsp;S. de Souza ,&nbsp;T.L. Silva ,&nbsp;A. Tomazoni ,&nbsp;P.K. Rocha\",\"doi\":\"10.1016/j.infpip.2023.100328\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Aim</h3><p>This study aimed to verify the efficacy of disinfection procedures to reduce <em>Acinetobacter baumannii bla</em><sub>OXA-23</sub> bacterial load in needleless connectors that had been experimentally contaminated.</p></div><div><h3>Methods</h3><p>Two-way intermediate extender's hub and needle-free valve were contaminated with <em>Acinetobacter baumannii bla</em><sub>OXA-23</sub>. To disinfect them, the following procedures were carried out: sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% isopropyl alcohol single-use cap, with eight times friction for 10 s, followed by 5 s drying time. The statistical tests Kruskal–Wallis and post-hoc Conover were performed using MedCalc®.</p></div><div><h3>Results</h3><p>A total of 82 experiments were conducted. All tested disinfection procedures were efficacious in reducing the A. <em>baumannii bla</em><sub>OXA-23</sub> load. The 70% IPA single-use cap was found to be the best method for disinfecting the two-way intermediate extender's hub (87.28%), while all the methods were efficacious for the disinfection of the needle-free valve (more than 90%). During the inoculation period, <em>A. baumannii bla</em><sub>OXA-23</sub> showed less adherence to the needle-free valve during the inoculation period, probably due to the device's design.</p></div><div><h3>Conclusion</h3><p>The three tested disinfection procedures using sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% IPA single-use cap were found to be efficacious in reducing the bacterial load of <em>A. baumanni bla</em><sub><em>OXA-23</em></sub> in needleless connectors. Proper disinfection of needleless connectors is a crucial nursing practice to prevent bloodstream infections, as it significantly reduces the bacterial load present in the device.</p></div>\",\"PeriodicalId\":33492,\"journal\":{\"name\":\"Infection Prevention in Practice\",\"volume\":\"6 1\",\"pages\":\"Article 100328\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2590088923000616/pdfft?md5=61d8e715701963a973546981466696b6&pid=1-s2.0-S2590088923000616-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Infection Prevention in Practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2590088923000616\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"INFECTIOUS DISEASES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Infection Prevention in Practice","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2590088923000616","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0

摘要

目的 本研究旨在验证消毒程序对减少受实验污染的无针接头中鲍曼不动杆菌 blaOXA-23 细菌量的效果。方法 用鲍曼不动杆菌 blaOXA-23 污染双向中间体延伸器的毂和无针阀。为了对其进行消毒,采用了以下方法:无菌纱布与 70% 乙醇、无菌纱布与 Incidin® 和 70% 异丙醇一次性使用帽,摩擦 8 次,每次 10 秒,然后干燥 5 秒。结果共进行了 82 次实验。所有测试的消毒程序都能有效减少鲍曼不动杆菌 blaOXA-23 的负荷。70%的IPA一次性使用瓶盖是消毒双向中间扩展器毂的最佳方法(87.28%),而所有方法对无针瓣膜的消毒效果都很好(90%以上)。结论 使用无菌纱布加 70% 乙醇、无菌纱布加 Incidin® 和 70% IPA 一次性使用瓶盖的三种测试消毒程序可有效减少无针接头中鲍曼不动杆菌 blaOXA-23 的细菌负荷。对无针接头进行适当消毒是预防血流感染的关键护理措施,因为它能显著减少装置中的细菌负荷。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Efficacy of disinfection procedures to reduce Acinetobacter baumanii blaOXA-23 contamination rate of needleless connectors: an in-vitro study

Aim

This study aimed to verify the efficacy of disinfection procedures to reduce Acinetobacter baumannii blaOXA-23 bacterial load in needleless connectors that had been experimentally contaminated.

Methods

Two-way intermediate extender's hub and needle-free valve were contaminated with Acinetobacter baumannii blaOXA-23. To disinfect them, the following procedures were carried out: sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% isopropyl alcohol single-use cap, with eight times friction for 10 s, followed by 5 s drying time. The statistical tests Kruskal–Wallis and post-hoc Conover were performed using MedCalc®.

Results

A total of 82 experiments were conducted. All tested disinfection procedures were efficacious in reducing the A. baumannii blaOXA-23 load. The 70% IPA single-use cap was found to be the best method for disinfecting the two-way intermediate extender's hub (87.28%), while all the methods were efficacious for the disinfection of the needle-free valve (more than 90%). During the inoculation period, A. baumannii blaOXA-23 showed less adherence to the needle-free valve during the inoculation period, probably due to the device's design.

Conclusion

The three tested disinfection procedures using sterile gauze with 70% ethanol, sterile gauze with Incidin®, and 70% IPA single-use cap were found to be efficacious in reducing the bacterial load of A. baumanni blaOXA-23 in needleless connectors. Proper disinfection of needleless connectors is a crucial nursing practice to prevent bloodstream infections, as it significantly reduces the bacterial load present in the device.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Infection Prevention in Practice
Infection Prevention in Practice Medicine-Public Health, Environmental and Occupational Health
CiteScore
4.80
自引率
0.00%
发文量
58
审稿时长
61 days
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信