{"title":"使用具有两种可切换分析模式的新一代离子交换高效液相色谱系统检测主要血红蛋白变体和次要血红蛋白变体及其干扰情况","authors":"Daisuke Manita ,&nbsp;Shinji Ogino ,&nbsp;Stefaan Marivoet ,&nbsp;Masatsune Ogura","doi":"10.1016/j.plabm.2023.e00346","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>High-performance liquid chromatography (HPLC) is commonly used to measure hemoglobin A<sub>1c</sub> (HbA<sub>1c</sub>) levels and detect hemoglobin variants (Hb-Vars). HLC-723GR01 (GR01) is a new-generation automated ion-exchange HPLC system with two switchable analysis modes, namely short (30 s/test) and long modes (50 s/test). We evaluated the general performance of both analysis modes of GR01 for quantifying HbA<sub>1c</sub> and detecting Hb-Vars.</p></div><div><h3>Design and methods</h3><p>We evaluated the instrument's precision based on CLSI protocol EP-05-A3. A comparison of the two analysis modes of GR01 against the standard mode of HLC-723G11 was performed on 100 whole blood samples. The GR01 long mode was compared with affinity HPLC (AF-HPLC) for detecting common Hb-Vars (HbE, HbD, HbS, and HbC, &gt;20 samples). To examine the detection capability for minor Hb-Vars, we analyzed 26 Hb-Vars using multiple analyzers, including both analysis modes of GR01.</p></div><div><h3>Results</h3><p>Both modes of GR01 had within-laboratory coefficients of variation of ≤1.0 % from four samples with HbA<sub>1c</sub> concentrations of 32–86 mmol/mol. Good correlation was observed between GR01 and HLC-723G11. The results for HbA<sub>1c</sub> detection in the presence of the major variants revealed a strong correlation between the long mode of GR01 and AF-HPLC (r = 0.986–0.998), and the difference biases ranged 0.1–1.9 mmol/mol. In the long mode, only one variant had a difference bias exceeding 14 % [10 % (%NGSP)].</p></div><div><h3>Conclusion</h3><p>The two analysis modes of GR01 were fast and had high accuracy and reproducibility, indicating their utility for routine clinical use in measuring HbA<sub>1c</sub> samples with Hb-Vars.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":"38 ","pages":"Article e00346"},"PeriodicalIF":1.7000,"publicationDate":"2023-12-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551723000409/pdfft?md5=0aab6be12d0810eb283e909b2347b95e&pid=1-s2.0-S2352551723000409-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Detectability of and interference by major and minor hemoglobin variants using a new-generation ion-exchange HPLC system with two switchable analysis modes\",\"authors\":\"Daisuke Manita ,&nbsp;Shinji Ogino ,&nbsp;Stefaan Marivoet ,&nbsp;Masatsune Ogura\",\"doi\":\"10.1016/j.plabm.2023.e00346\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><p>High-performance liquid chromatography (HPLC) is commonly used to measure hemoglobin A<sub>1c</sub> (HbA<sub>1c</sub>) levels and detect hemoglobin variants (Hb-Vars). HLC-723GR01 (GR01) is a new-generation automated ion-exchange HPLC system with two switchable analysis modes, namely short (30 s/test) and long modes (50 s/test). We evaluated the general performance of both analysis modes of GR01 for quantifying HbA<sub>1c</sub> and detecting Hb-Vars.</p></div><div><h3>Design and methods</h3><p>We evaluated the instrument's precision based on CLSI protocol EP-05-A3. A comparison of the two analysis modes of GR01 against the standard mode of HLC-723G11 was performed on 100 whole blood samples. The GR01 long mode was compared with affinity HPLC (AF-HPLC) for detecting common Hb-Vars (HbE, HbD, HbS, and HbC, &gt;20 samples). To examine the detection capability for minor Hb-Vars, we analyzed 26 Hb-Vars using multiple analyzers, including both analysis modes of GR01.</p></div><div><h3>Results</h3><p>Both modes of GR01 had within-laboratory coefficients of variation of ≤1.0 % from four samples with HbA<sub>1c</sub> concentrations of 32–86 mmol/mol. Good correlation was observed between GR01 and HLC-723G11. The results for HbA<sub>1c</sub> detection in the presence of the major variants revealed a strong correlation between the long mode of GR01 and AF-HPLC (r = 0.986–0.998), and the difference biases ranged 0.1–1.9 mmol/mol. In the long mode, only one variant had a difference bias exceeding 14 % [10 % (%NGSP)].</p></div><div><h3>Conclusion</h3><p>The two analysis modes of GR01 were fast and had high accuracy and reproducibility, indicating their utility for routine clinical use in measuring HbA<sub>1c</sub> samples with Hb-Vars.</p></div>\",\"PeriodicalId\":20421,\"journal\":{\"name\":\"Practical Laboratory Medicine\",\"volume\":\"38 \",\"pages\":\"Article e00346\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2023-12-09\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2352551723000409/pdfft?md5=0aab6be12d0810eb283e909b2347b95e&pid=1-s2.0-S2352551723000409-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Practical Laboratory Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2352551723000409\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551723000409","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的高效液相色谱 (HPLC) 常用于测量血红蛋白 A1c (HbA1c) 水平和检测血红蛋白变异体 (Hb-Vars)。HLC-723GR01 (GR01) 是新一代自动离子交换 HPLC 系统,具有两种可切换的分析模式,即短模式(30 秒/次)和长模式(50 秒/次)。我们评估了 GR01 两种分析模式在定量 HbA1c 和检测 Hb-Vars 方面的总体性能。在 100 份全血样本中,将 GR01 的两种分析模式与 HLC-723G11 的标准模式进行了比较。在检测常见的 Hb-Vars(HbE、HbD、HbS 和 HbC,共 20 份样本)时,将 GR01 长模式与亲和 HPLC(AF-HPLC)进行了比较。结果从 HbA1c 浓度为 32-86 mmol/mol 的四个样本中,GR01 的两种分析模式在实验室内的变异系数均小于 1.0%。GR01 和 HLC-723G11 之间具有良好的相关性。主要变异体存在时的 HbA1c 检测结果显示,GR01 的长模式与 AF-HPLC 之间具有很强的相关性(r = 0.986-0.998),差异偏差范围为 0.1-1.9 mmol/mol。结论 GR01 的两种分析模式分析速度快、准确度高、重现性好,可用于临床常规 Hb-Vars 测量 HbA1c 样品。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Detectability of and interference by major and minor hemoglobin variants using a new-generation ion-exchange HPLC system with two switchable analysis modes

Objectives

High-performance liquid chromatography (HPLC) is commonly used to measure hemoglobin A1c (HbA1c) levels and detect hemoglobin variants (Hb-Vars). HLC-723GR01 (GR01) is a new-generation automated ion-exchange HPLC system with two switchable analysis modes, namely short (30 s/test) and long modes (50 s/test). We evaluated the general performance of both analysis modes of GR01 for quantifying HbA1c and detecting Hb-Vars.

Design and methods

We evaluated the instrument's precision based on CLSI protocol EP-05-A3. A comparison of the two analysis modes of GR01 against the standard mode of HLC-723G11 was performed on 100 whole blood samples. The GR01 long mode was compared with affinity HPLC (AF-HPLC) for detecting common Hb-Vars (HbE, HbD, HbS, and HbC, >20 samples). To examine the detection capability for minor Hb-Vars, we analyzed 26 Hb-Vars using multiple analyzers, including both analysis modes of GR01.

Results

Both modes of GR01 had within-laboratory coefficients of variation of ≤1.0 % from four samples with HbA1c concentrations of 32–86 mmol/mol. Good correlation was observed between GR01 and HLC-723G11. The results for HbA1c detection in the presence of the major variants revealed a strong correlation between the long mode of GR01 and AF-HPLC (r = 0.986–0.998), and the difference biases ranged 0.1–1.9 mmol/mol. In the long mode, only one variant had a difference bias exceeding 14 % [10 % (%NGSP)].

Conclusion

The two analysis modes of GR01 were fast and had high accuracy and reproducibility, indicating their utility for routine clinical use in measuring HbA1c samples with Hb-Vars.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Practical Laboratory Medicine
Practical Laboratory Medicine Health Professions-Radiological and Ultrasound Technology
CiteScore
3.50
自引率
0.00%
发文量
40
审稿时长
7 weeks
期刊介绍: Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信