旋涡混合器振动对红细胞的影响。

IF 1.4 4区 医学 Q4 HEMATOLOGY
Chuang Xiang, Liang Wang
{"title":"旋涡混合器振动对红细胞的影响。","authors":"Chuang Xiang,&nbsp;Liang Wang","doi":"10.1016/j.tracli.2023.11.003","DOIUrl":null,"url":null,"abstract":"<div><h3>Purposes</h3><p>Red blood cells (RBCs) are often subject to vibration during processing, transfusion, and transport. Further research is necessary to understand the effects of vibration on human RBCs and to reduce experimental deviations caused by device vibration.</p></div><div><h3>Methods</h3><p>Flow cytometry was used in this study to observe the cytokine expression of IgG and IgA and deformation of human red blood cells affected by the vibration of a vortex mixer with varying frequency (750 rpm and 1500 rpm), duration (5 min and 10 min), and container volume (96 well plate and 48 well plate).</p></div><div><h3>Results</h3><p>The size of RBCs in duration of 10 min is obviously smaller than the duration of 5 min. The 10-minute duration led to visibly smaller RBC sizes compared to the 5-minute duration. There was little effect on the size of RBCs in the 10-minute groups from differences in frequency and container volume. However, decreased RBC size can be observed in the 5-minute groups, where frequency is increased or container volume is decreased. Echinocytes were present in photomicrographs of all 10-minute groups, but microstructure of the RBCs was not impacted by vortex mixer vibration. The elevated frequency or reduced container volume results in an increased cytokine expression of IgG within the 5-minute groupings.</p></div><div><h3>Conclusion</h3><p>It can be inferred that vibration must not be overlooked due to its potential impact on the shape and cytokine expression of RBCs. Hence, the inclusion of vibration must be taken into consideration in experiments and devices pertaining to RBCs.</p></div>","PeriodicalId":23262,"journal":{"name":"Transfusion Clinique et Biologique","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S1246782023005463/pdfft?md5=934d074120c374e7f07d8ec0fb47b894&pid=1-s2.0-S1246782023005463-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Effect of vibration of the vortex mixer on the red blood cells\",\"authors\":\"Chuang Xiang,&nbsp;Liang Wang\",\"doi\":\"10.1016/j.tracli.2023.11.003\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purposes</h3><p>Red blood cells (RBCs) are often subject to vibration during processing, transfusion, and transport. Further research is necessary to understand the effects of vibration on human RBCs and to reduce experimental deviations caused by device vibration.</p></div><div><h3>Methods</h3><p>Flow cytometry was used in this study to observe the cytokine expression of IgG and IgA and deformation of human red blood cells affected by the vibration of a vortex mixer with varying frequency (750 rpm and 1500 rpm), duration (5 min and 10 min), and container volume (96 well plate and 48 well plate).</p></div><div><h3>Results</h3><p>The size of RBCs in duration of 10 min is obviously smaller than the duration of 5 min. The 10-minute duration led to visibly smaller RBC sizes compared to the 5-minute duration. There was little effect on the size of RBCs in the 10-minute groups from differences in frequency and container volume. However, decreased RBC size can be observed in the 5-minute groups, where frequency is increased or container volume is decreased. Echinocytes were present in photomicrographs of all 10-minute groups, but microstructure of the RBCs was not impacted by vortex mixer vibration. The elevated frequency or reduced container volume results in an increased cytokine expression of IgG within the 5-minute groupings.</p></div><div><h3>Conclusion</h3><p>It can be inferred that vibration must not be overlooked due to its potential impact on the shape and cytokine expression of RBCs. Hence, the inclusion of vibration must be taken into consideration in experiments and devices pertaining to RBCs.</p></div>\",\"PeriodicalId\":23262,\"journal\":{\"name\":\"Transfusion Clinique et Biologique\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2024-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S1246782023005463/pdfft?md5=934d074120c374e7f07d8ec0fb47b894&pid=1-s2.0-S1246782023005463-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Transfusion Clinique et Biologique\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1246782023005463\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transfusion Clinique et Biologique","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1246782023005463","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:红细胞(rbc)在加工、输血和运输过程中经常受到振动。进一步研究振动对人体红细胞的影响,减少设备振动引起的实验偏差是必要的。方法:采用流式细胞术观察不同频率(750 rpm和1500 rpm)、持续时间(5 min和10 min)、容器体积(96孔板和48孔板)振动对人红细胞IgG和IgA细胞因子表达及变形的影响。结果:10min时红细胞体积明显小于5min时,10min时红细胞体积明显小于5min时。在10分钟组中,频率和容器体积的差异对红细胞大小的影响很小。然而,在5分钟组中可以观察到RBC大小减少,其中频率增加或容器体积减少。在所有10分钟组的显微照片中都有棘细胞,但红细胞的微观结构没有受到涡流混合器振动的影响。频率升高或容器体积减小导致5分钟内IgG细胞因子表达增加。结论:振动对红细胞形态及细胞因子表达的影响不容忽视。因此,在与红细胞有关的实验和设备中必须考虑振动的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Effect of vibration of the vortex mixer on the red blood cells

Purposes

Red blood cells (RBCs) are often subject to vibration during processing, transfusion, and transport. Further research is necessary to understand the effects of vibration on human RBCs and to reduce experimental deviations caused by device vibration.

Methods

Flow cytometry was used in this study to observe the cytokine expression of IgG and IgA and deformation of human red blood cells affected by the vibration of a vortex mixer with varying frequency (750 rpm and 1500 rpm), duration (5 min and 10 min), and container volume (96 well plate and 48 well plate).

Results

The size of RBCs in duration of 10 min is obviously smaller than the duration of 5 min. The 10-minute duration led to visibly smaller RBC sizes compared to the 5-minute duration. There was little effect on the size of RBCs in the 10-minute groups from differences in frequency and container volume. However, decreased RBC size can be observed in the 5-minute groups, where frequency is increased or container volume is decreased. Echinocytes were present in photomicrographs of all 10-minute groups, but microstructure of the RBCs was not impacted by vortex mixer vibration. The elevated frequency or reduced container volume results in an increased cytokine expression of IgG within the 5-minute groupings.

Conclusion

It can be inferred that vibration must not be overlooked due to its potential impact on the shape and cytokine expression of RBCs. Hence, the inclusion of vibration must be taken into consideration in experiments and devices pertaining to RBCs.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.50
自引率
11.80%
发文量
234
审稿时长
36 days
期刊介绍: Transfusion Clinique et Biologique, the official journal of the French Society of Blood Transfusion (SFTS): - an aid to training, at a European level - the only French journal indexed in the hematology and immunology sections of Current Contents Transfusion Clinique et Biologique spans fundamental research and everyday practice, with articles coming from both sides. Articles, reviews, case reports, letters to the editor and editorials are published in 4 editions a year, in French or in English, covering all scientific and medical aspects of transfusion: immunology, hematology, infectious diseases, genetics, molecular biology, etc. And finally, a convivial cross-disciplinary section on training and information offers practical updates. Readership: "Transfusers" are many and various: anesthetists, biologists, hematologists, and blood-bank, ICU and mobile emergency specialists...
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信