连续动脉正电子监测仪在PET成像中的定量。

A D Nelson, R F Muzic, F Miraldi, G J Muswick, G P Leisure, W Voelker
{"title":"连续动脉正电子监测仪在PET成像中的定量。","authors":"A D Nelson,&nbsp;R F Muzic,&nbsp;F Miraldi,&nbsp;G J Muswick,&nbsp;G P Leisure,&nbsp;W Voelker","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Quantitative measures of physiologic function with PET require continuous monitoring of arterial positron isotope concentration. A device has been developed that automates this process. This device has advantages over manual sampling techniques with syringes since fewer personnel are required, measurements are less error prone, and more continuous measures of arterial positron concentration are available. A constant flow infusion/withdrawal pump withdraws blood from the radial artery through a catheter connected to 0.5 mm inner diameter teflon tubing. This tubing is wrapped around a 50 mm thick by 50 mm diameter NaI(T1) crystal that is interfaced to a photomultiplier tube (PMT) and encased in a cylindrical lead shield. This crystal detects 511 Kev photons that result from positron annihilation. The device sensitivity is greater than 240 (cts/sec)/(microCi/ml) corresponding to a peak activity of approximately 10,000 cts/sec for a 50 mCi bolus injection in an adult. The system dynamic response has been measured and the true arterial input function is recovered by deconvolution. The system has been used clinically for more than 400 human PET studies and has been a reliable continuous monitor of arterial positron concentration.</p>","PeriodicalId":76992,"journal":{"name":"American journal of physiologic imaging","volume":"5 2","pages":"84-8"},"PeriodicalIF":0.0000,"publicationDate":"1990-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Continuous arterial positron monitor for quantitation in PET imaging.\",\"authors\":\"A D Nelson,&nbsp;R F Muzic,&nbsp;F Miraldi,&nbsp;G J Muswick,&nbsp;G P Leisure,&nbsp;W Voelker\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Quantitative measures of physiologic function with PET require continuous monitoring of arterial positron isotope concentration. A device has been developed that automates this process. This device has advantages over manual sampling techniques with syringes since fewer personnel are required, measurements are less error prone, and more continuous measures of arterial positron concentration are available. A constant flow infusion/withdrawal pump withdraws blood from the radial artery through a catheter connected to 0.5 mm inner diameter teflon tubing. This tubing is wrapped around a 50 mm thick by 50 mm diameter NaI(T1) crystal that is interfaced to a photomultiplier tube (PMT) and encased in a cylindrical lead shield. This crystal detects 511 Kev photons that result from positron annihilation. The device sensitivity is greater than 240 (cts/sec)/(microCi/ml) corresponding to a peak activity of approximately 10,000 cts/sec for a 50 mCi bolus injection in an adult. The system dynamic response has been measured and the true arterial input function is recovered by deconvolution. The system has been used clinically for more than 400 human PET studies and has been a reliable continuous monitor of arterial positron concentration.</p>\",\"PeriodicalId\":76992,\"journal\":{\"name\":\"American journal of physiologic imaging\",\"volume\":\"5 2\",\"pages\":\"84-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"American journal of physiologic imaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"American journal of physiologic imaging","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

用PET定量测量生理功能需要连续监测动脉正电子同位素浓度。已经研制出一种使这一过程自动化的装置。该设备与注射器手动采样技术相比具有优势,因为需要的人员更少,测量误差更小,并且可以更连续地测量动脉正电子浓度。恒流量输注/抽吸泵通过连接内径0.5 mm的聚四氟乙烯管的导管从桡动脉抽出血液。该管包裹在50mm厚,50mm直径的NaI(T1)晶体上,该晶体与光电倍增管(PMT)相连,并包裹在圆柱形铅屏蔽中。这种晶体能探测到511个由正电子湮灭产生的Kev光子。该装置的灵敏度大于240 (cts/sec)/(microCi/ml),对应于成人50 mCi注射的峰值活性约为10,000 cts/sec。测量了系统的动态响应,并通过反卷积恢复了真实的动脉输入函数。该系统已在临床上用于400多项人体PET研究,并已成为动脉正电子浓度的可靠连续监测。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Continuous arterial positron monitor for quantitation in PET imaging.

Quantitative measures of physiologic function with PET require continuous monitoring of arterial positron isotope concentration. A device has been developed that automates this process. This device has advantages over manual sampling techniques with syringes since fewer personnel are required, measurements are less error prone, and more continuous measures of arterial positron concentration are available. A constant flow infusion/withdrawal pump withdraws blood from the radial artery through a catheter connected to 0.5 mm inner diameter teflon tubing. This tubing is wrapped around a 50 mm thick by 50 mm diameter NaI(T1) crystal that is interfaced to a photomultiplier tube (PMT) and encased in a cylindrical lead shield. This crystal detects 511 Kev photons that result from positron annihilation. The device sensitivity is greater than 240 (cts/sec)/(microCi/ml) corresponding to a peak activity of approximately 10,000 cts/sec for a 50 mCi bolus injection in an adult. The system dynamic response has been measured and the true arterial input function is recovered by deconvolution. The system has been used clinically for more than 400 human PET studies and has been a reliable continuous monitor of arterial positron concentration.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信