勃林格曼海姆/日立717分析系统的多中心评估。

H Baadenhuijsen, P M Bayer, H Keller, M Knedel, N Montalbetti, S Brenna, L Prencipe, A Vassault, M Bailly, H T Phung
{"title":"勃林格曼海姆/日立717分析系统的多中心评估。","authors":"H Baadenhuijsen,&nbsp;P M Bayer,&nbsp;H Keller,&nbsp;M Knedel,&nbsp;N Montalbetti,&nbsp;S Brenna,&nbsp;L Prencipe,&nbsp;A Vassault,&nbsp;M Bailly,&nbsp;H T Phung","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>We conducted an European multicentre trial to assess the performance of the new Boehringer Mannheim/Hitachi 717 analysis system. The photometer response was linear up to an absorbance of 2.8. The maximal CV of photometric imprecision was 0.5% for the wavelength pair 340/405 nm within the absorbance range 0.9 to 2.4. For the 13 analytes in our study, mean within-run imprecision was less than 2%, and mean between-day imprecision less than 2.5%. The results obtained with the Hitachi 717 instrument correlated closely with those of comparison instruments. Linearity for the various tests was high and exceeded the manufacturer's claims. No drift was detected during an 8-hour work period; carry over could not be detected under the chosen experimental conditions. The new instrument was readily accepted by the evaluators because of its ease of handling and simple daily maintenance.</p>","PeriodicalId":15649,"journal":{"name":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","volume":"28 4","pages":"261-9"},"PeriodicalIF":0.0000,"publicationDate":"1990-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Multicentre evaluation of the Boehringer Mannheim/Hitachi 717 analysis system.\",\"authors\":\"H Baadenhuijsen,&nbsp;P M Bayer,&nbsp;H Keller,&nbsp;M Knedel,&nbsp;N Montalbetti,&nbsp;S Brenna,&nbsp;L Prencipe,&nbsp;A Vassault,&nbsp;M Bailly,&nbsp;H T Phung\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We conducted an European multicentre trial to assess the performance of the new Boehringer Mannheim/Hitachi 717 analysis system. The photometer response was linear up to an absorbance of 2.8. The maximal CV of photometric imprecision was 0.5% for the wavelength pair 340/405 nm within the absorbance range 0.9 to 2.4. For the 13 analytes in our study, mean within-run imprecision was less than 2%, and mean between-day imprecision less than 2.5%. The results obtained with the Hitachi 717 instrument correlated closely with those of comparison instruments. Linearity for the various tests was high and exceeded the manufacturer's claims. No drift was detected during an 8-hour work period; carry over could not be detected under the chosen experimental conditions. The new instrument was readily accepted by the evaluators because of its ease of handling and simple daily maintenance.</p>\",\"PeriodicalId\":15649,\"journal\":{\"name\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"volume\":\"28 4\",\"pages\":\"261-9\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

我们进行了一项欧洲多中心试验,以评估新的勃林格曼海姆/日立717分析系统的性能。在2.8的吸光度下,光度计的响应呈线性。在0.9 ~ 2.4的吸光度范围内,340/405 nm波长对的光度不精确CV值最大为0.5%。对于我们研究中的13种分析物,平均运行内不精密度小于2%,平均日间不精密度小于2.5%。用日立717仪器测得的结果与比较仪器的结果有密切的相关性。各种测试的线性度很高,超过了制造商的要求。8小时工作期间未检测到漂移;在选定的实验条件下,无法检测到结转。由于易于操作和日常维护,新仪器很容易被评估人员接受。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Multicentre evaluation of the Boehringer Mannheim/Hitachi 717 analysis system.

We conducted an European multicentre trial to assess the performance of the new Boehringer Mannheim/Hitachi 717 analysis system. The photometer response was linear up to an absorbance of 2.8. The maximal CV of photometric imprecision was 0.5% for the wavelength pair 340/405 nm within the absorbance range 0.9 to 2.4. For the 13 analytes in our study, mean within-run imprecision was less than 2%, and mean between-day imprecision less than 2.5%. The results obtained with the Hitachi 717 instrument correlated closely with those of comparison instruments. Linearity for the various tests was high and exceeded the manufacturer's claims. No drift was detected during an 8-hour work period; carry over could not be detected under the chosen experimental conditions. The new instrument was readily accepted by the evaluators because of its ease of handling and simple daily maintenance.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信