维多利亚州用于证据使用的呼气酒精测量装置的评价。

Acta medicinae legalis et socialis Pub Date : 1990-01-01
C M Bell, S J Gutowski
{"title":"维多利亚州用于证据使用的呼气酒精测量装置的评价。","authors":"C M Bell,&nbsp;S J Gutowski","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The accuracy, precision and specificity for ethanol of eight different evidential breath alcohol measuring devices were evaluated. Other factors of importance, such as susceptibility to radio frequency interference and variation in the power supply voltage were also examined. On the basis of the above criteria, three instruments were identified as more suitable than the remainder for evidential use. These three instruments are based on the principle of absorption of infra-red radiation in the 9.5 um region and are the Drager Alcotest 7110, Siemens Alcomat V5.2F and Seres Ethylometre 679. Although the specificity of these instruments was adequate for most substances tested, it was not absolute. As a result, if these devices were introduced for evidential purposes, arguments may arise in the courts with respect to specificity of the devices for ethanol. Such arguments could not be resolved without recourse to physiological and toxicological data.</p>","PeriodicalId":75386,"journal":{"name":"Acta medicinae legalis et socialis","volume":"40 ","pages":"95-105"},"PeriodicalIF":0.0000,"publicationDate":"1990-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of breath alcohol measuring devices intended for evidential use in Victoria.\",\"authors\":\"C M Bell,&nbsp;S J Gutowski\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The accuracy, precision and specificity for ethanol of eight different evidential breath alcohol measuring devices were evaluated. Other factors of importance, such as susceptibility to radio frequency interference and variation in the power supply voltage were also examined. On the basis of the above criteria, three instruments were identified as more suitable than the remainder for evidential use. These three instruments are based on the principle of absorption of infra-red radiation in the 9.5 um region and are the Drager Alcotest 7110, Siemens Alcomat V5.2F and Seres Ethylometre 679. Although the specificity of these instruments was adequate for most substances tested, it was not absolute. As a result, if these devices were introduced for evidential purposes, arguments may arise in the courts with respect to specificity of the devices for ethanol. Such arguments could not be resolved without recourse to physiological and toxicological data.</p>\",\"PeriodicalId\":75386,\"journal\":{\"name\":\"Acta medicinae legalis et socialis\",\"volume\":\"40 \",\"pages\":\"95-105\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta medicinae legalis et socialis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta medicinae legalis et socialis","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

评价了8种不同的证据呼气酒精测量装置对乙醇的准确度、精密度和特异性。其他重要的因素,如对射频干扰的敏感性和电源电压的变化也进行了检查。根据上述标准,确定了三种仪器比其他仪器更适合作为证据使用。这三种仪器基于9.5 um区域红外辐射吸收原理,分别是Drager Alcotest 7110, Siemens Alcomat V5.2F和Seres Ethylometre 679。虽然这些仪器的特异性对大多数被测物质是足够的,但它不是绝对的。因此,如果这些装置是为了证据的目的而引入的,法庭上可能会出现关于乙醇装置的特殊性的争论。不借助生理学和毒理学数据,就无法解决这些争论。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of breath alcohol measuring devices intended for evidential use in Victoria.

The accuracy, precision and specificity for ethanol of eight different evidential breath alcohol measuring devices were evaluated. Other factors of importance, such as susceptibility to radio frequency interference and variation in the power supply voltage were also examined. On the basis of the above criteria, three instruments were identified as more suitable than the remainder for evidential use. These three instruments are based on the principle of absorption of infra-red radiation in the 9.5 um region and are the Drager Alcotest 7110, Siemens Alcomat V5.2F and Seres Ethylometre 679. Although the specificity of these instruments was adequate for most substances tested, it was not absolute. As a result, if these devices were introduced for evidential purposes, arguments may arise in the courts with respect to specificity of the devices for ethanol. Such arguments could not be resolved without recourse to physiological and toxicological data.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信