M. Sanoh, A. Mikata, K. Tokuno, M. Hanazawa, H. Kakitani
{"title":"商用同轴分流器在大脉冲电流下的校准方法","authors":"M. Sanoh, A. Mikata, K. Tokuno, M. Hanazawa, H. Kakitani","doi":"10.1109/CPEM.2014.6898543","DOIUrl":null,"url":null,"abstract":"This paper describes a method to calibrate a commercial coaxial shunt resistor with 0.07% of standard uncertainty at 1200 A, 0.5 ms pulse width current. In the method, a special current shunt is used as the new working standard in Measurement Standards Center (MSC) of Agilent Technologies International Japan. The shunt was designed for currents up to 1500 A at both pulse and DC, and then characterized at DC using a known DC shunt. The impedance characteristic is evaluated at pulse by comparing the voltage across the shunt with the commercial coaxial shunt resistor at the pulse current.","PeriodicalId":256575,"journal":{"name":"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)","volume":"14 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-10-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":"{\"title\":\"A calibration method for a commercial coaxial shunt at high pulse current\",\"authors\":\"M. Sanoh, A. Mikata, K. Tokuno, M. Hanazawa, H. Kakitani\",\"doi\":\"10.1109/CPEM.2014.6898543\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper describes a method to calibrate a commercial coaxial shunt resistor with 0.07% of standard uncertainty at 1200 A, 0.5 ms pulse width current. In the method, a special current shunt is used as the new working standard in Measurement Standards Center (MSC) of Agilent Technologies International Japan. The shunt was designed for currents up to 1500 A at both pulse and DC, and then characterized at DC using a known DC shunt. The impedance characteristic is evaluated at pulse by comparing the voltage across the shunt with the commercial coaxial shunt resistor at the pulse current.\",\"PeriodicalId\":256575,\"journal\":{\"name\":\"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)\",\"volume\":\"14 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2014-10-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"5\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/CPEM.2014.6898543\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"29th Conference on Precision Electromagnetic Measurements (CPEM 2014)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM.2014.6898543","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A calibration method for a commercial coaxial shunt at high pulse current
This paper describes a method to calibrate a commercial coaxial shunt resistor with 0.07% of standard uncertainty at 1200 A, 0.5 ms pulse width current. In the method, a special current shunt is used as the new working standard in Measurement Standards Center (MSC) of Agilent Technologies International Japan. The shunt was designed for currents up to 1500 A at both pulse and DC, and then characterized at DC using a known DC shunt. The impedance characteristic is evaluated at pulse by comparing the voltage across the shunt with the commercial coaxial shunt resistor at the pulse current.