{"title":"使用正弦图一致性标准校正PET的透射和发射扫描以纠正受试者运动","authors":"Yu. Chin-Lung, Sung-Cheng Huang, M. Dahlbom, M. Bergsneider","doi":"10.1109/NSSMIC.2003.1352487","DOIUrl":null,"url":null,"abstract":"Subject movement leads to misalignment between transmission and emission PET scans acquired at different times. This leads to inaccurate attenuation correction and causes errors in PET images. We propose a new method, which seeks to minimize the variance of the total counts of attenuation corrected emission sinogram among all projection angles. The underlying basis of the method is that if the transmission and emission scans are aligned, the total counts (after attenuation correction) at different projection angles should be equal (i.e., with zero variance). Therefore, by adjusting the spatial transformation parameters between the transmission and emission images to minimize such a variance as a cost function, the two scans would be aligned. The effectiveness of this method is demonstrated in two-dimensional situations with computer simulated data and with a physical phantom PET study.","PeriodicalId":186175,"journal":{"name":"2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515)","volume":"4 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2003-10-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Alignment of transmission and emission scans of PET to correct for subject movement using a sinogram consistency criterion\",\"authors\":\"Yu. Chin-Lung, Sung-Cheng Huang, M. Dahlbom, M. Bergsneider\",\"doi\":\"10.1109/NSSMIC.2003.1352487\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Subject movement leads to misalignment between transmission and emission PET scans acquired at different times. This leads to inaccurate attenuation correction and causes errors in PET images. We propose a new method, which seeks to minimize the variance of the total counts of attenuation corrected emission sinogram among all projection angles. The underlying basis of the method is that if the transmission and emission scans are aligned, the total counts (after attenuation correction) at different projection angles should be equal (i.e., with zero variance). Therefore, by adjusting the spatial transformation parameters between the transmission and emission images to minimize such a variance as a cost function, the two scans would be aligned. The effectiveness of this method is demonstrated in two-dimensional situations with computer simulated data and with a physical phantom PET study.\",\"PeriodicalId\":186175,\"journal\":{\"name\":\"2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515)\",\"volume\":\"4 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-10-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.2003.1352487\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2003 IEEE Nuclear Science Symposium. Conference Record (IEEE Cat. No.03CH37515)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2003.1352487","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

受试者的运动导致在不同时间获得的透射和发射PET扫描之间的不对齐。这会导致衰减校正不准确,并导致PET图像出现错误。我们提出了一种新的方法,寻求衰减校正发射正弦图的总计数在所有投影角度之间的方差最小。该方法的基础是,如果透射和发射扫描对齐,则不同投影角度的总计数(衰减校正后)应该相等(即零方差)。因此,通过调整透射和发射图像之间的空间变换参数,将这种方差作为代价函数最小化,从而使两次扫描对齐。通过计算机模拟数据和物理幻影PET研究证明了该方法在二维情况下的有效性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Alignment of transmission and emission scans of PET to correct for subject movement using a sinogram consistency criterion
Subject movement leads to misalignment between transmission and emission PET scans acquired at different times. This leads to inaccurate attenuation correction and causes errors in PET images. We propose a new method, which seeks to minimize the variance of the total counts of attenuation corrected emission sinogram among all projection angles. The underlying basis of the method is that if the transmission and emission scans are aligned, the total counts (after attenuation correction) at different projection angles should be equal (i.e., with zero variance). Therefore, by adjusting the spatial transformation parameters between the transmission and emission images to minimize such a variance as a cost function, the two scans would be aligned. The effectiveness of this method is demonstrated in two-dimensional situations with computer simulated data and with a physical phantom PET study.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信